NA TCN-15Recalls & reported problems
Boston Scientific Neuromodulation Corporation · Nitinol TC Electrode, 15cm · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00813250013061), received between Jun 5, 2017 and Jun 1, 2021 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Material Discolored” (4 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Apr 29, 2026Open, Classified
Reusable electrodes may not meet expected performance levels.
Z-2403-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- May 30, 2019Terminated
The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
Z-1696-2020 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Discolored4
- Break3
- Material Separation3
- Contamination /Decontamination Problem1
- Defective Device1
- Fluid/Blood Leak1
- Fracture1
- Leak/Splash1
Patient problems
- No Known Impact Or Consequence To Patient8
- No Clinical Signs, Symptoms or Conditions1
- No Patient Involvement1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 1, 2021Defective Device; Defective Device
IT WAS REPORTED THAT THE ELECTRODE WAS BROKEN AT THE HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11916424 ↗MalfunctionAug 8, 2019Fluid/Blood Leak; Fluid/Blood Leak
A REPORT WAS RECEIVED THAT THE FACILITY WAS UNABLE TO CLEAN 3 PROBES DUE TO THE HUB LEAKING AT THE BASE OF THE PROBE. THE FACILITY CLAIMS THAT IT WAS BLOOD LEAKING OUT OF THE HUB.
Patient problems: No Patient Involvement
MAUDE report 8876322 ↗MalfunctionAug 1, 2018Material Discolored; Fracture; Material Discolored; Fracture
DEVICE ANALYSIS WAS PERFORMED ON THE RETURNED ELECTRODE. THE ELECTRODE PASSED THE FUNCTIONAL TEST HOWEVER A VISUAL INSPECTION REVEALED DISCOLORED EPOXY. THE COLOR OF NEWLY CURED EPOXY IS AMBER LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED. THIS FAILURE MECHANISM IS UNRELATED TO THE REPORTED COMPLAINT AND IS DUE TO THERMAL STRESS ASSOCIATED WITH MULTIPLE AUTOCLAVE CYCLES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7740421 ↗MalfunctionOct 2, 2017Leak/Splash; Leak/Splash
THE RETURNED ELECTRODE VISUAL INSPECTION REVEALED TRACES OF POTTING EPOXY ON THE HUB. THE EPOXY LEAKED THROUGH THE SHAFT OPENING IN ITS LIQUID STATE DURING THE MANUFACTURING PROCESS, AND WAS NOT REMOVED AFTER CURING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6908612 ↗MalfunctionSep 11, 2017Break; Material Discolored; Contamination /Decontamination Problem; Break; Material Discolored; Contamination /Decontamination Problem
THE DEVICE ANALYSIS PERFORMED SHOWED THAT THE EPOXY HAS CHIP OUT AND DISCOLORATION. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6854680 ↗MalfunctionAug 25, 2017Break; Break
DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODE REVEALED THAT THE SHAFT WAS BROKEN OFF FROM THE HUB.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6820888 ↗MalfunctionAug 11, 2017Material Discolored; Material Erosion; Material Discolored; Material Erosion
DEVICE ANALYSIS PERFORMED ON ALL THE RETURNED ELECTRODES SHOWED THAT THE EPOXY WAS CHIPPED OUT AND DISCOLORED DUE TO THERMAL STRESS ASSOCIATED WITH AUTOCLAVE CYCLES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6788910 ↗MalfunctionJun 20, 2017Break; Material Discolored; Material Separation; Break; Material Discolored; Material Separation
A REPORT WAS RECEIVED THAT THE PLASTIC ON THE ELECTRODE WAS BREAKING OFF AND COMING APART.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6654512 ↗MalfunctionJun 6, 2017Material Separation; Material Separation
A REPORT WAS RECEIVED THAT AN ELECTRODE WAS NOT REACHING TEMPERATURE. DEVICE ANALYSIS SHOWED THAT THE ELECTRODE SHAFT WAS BROKEN AT THE HUB.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6614592 ↗MalfunctionJun 5, 2017Material Separation; Material Separation
A REPORT WAS RECEIVED THAT AN ELECTRODE WAS NOT COMING UP TO TEMPERATURE. DEVICE ANALYSIS CONFIRMED THROUGH VISUAL INSPECTION THAT THE ELECTRODE STILL WORKED BUT HAD A LOOSE HUB.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6612500 ↗
Other PROBE, RADIOFREQUENCY LESION devices
- NA TCN-10-3MBoston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION2 recalls9 reports
- NA CSK-TC10Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION1 recall1 death58 reports
- NA CLASSICSt. Jude Medical, Inc. · PROBE, RADIOFREQUENCY LESION1 recall40 reports
- Intracept RLV 0055Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall30 reports
- Intracept RLV 0045Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall1 death23 reports
- Intracept RLV0065Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall14 reports
- NA CSK-TC5Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION1 recall3 reports
- Intracept RLV 0075Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page