Convenience kits containing select SKUs of 10mL… recallZ-2581-2026 · Medline Industries, LP · initiated Feb 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
FDA root cause category: No Marketing Application
What the company is doing
This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.
Recall details
- Recall number
- Z-2581-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 27, 2026
- Posted by the FDA
- Jun 25, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 113,843 kits
- Product code
- OEQ
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
Product description
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAM PACK-LF DYNJ21574J JUDKINS PACK DYNJ51126
Code information (models, lots, serial numbers)
DYNJ21574J UDI-DI 10889942153497 Lots 21ALA586 21CLA398 21ELA408 21FLA144 21FLA788 21HLA420 21ILA431 DYNJ51126 UDI-DI 10889942122820 Lots 21ADA917
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ51126Medline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ21574JMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99062).
- Z-2582-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIO CATH COMBINED PACK DYNJ43609S ANGIO PACK DYNJ67987D DYNJ80317A ANGIO SPECIAL PROCEDURE-LF D…
- Z-2583-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. BARIATRIC DYNJ905153J DYNJ905153K DYNJ905153L GYNE LAPAROTOMY PACK-LF DYNJ49593A LAPAROSCOPY …
- Z-2584-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. DR G'S BREAST PACK DYNJ51124B LIPO SUPPLEMENT PACK-LF DYNJ0341174D PLASTIC PACK DYNJ66746…
- Z-2585-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. AAA DYNJ905651B ARTERIOGRAM TRAY 00-402001S AV SHUNT PACK-LF DYNJ0842516F BASIC IR PACK…
- Z-2586-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAPHY PACK-LF DYNJ0854485S DEEP BRAIN STIMULATION PACK DYNJ0842793G NEURO ACCESSOR…
- Z-2587-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CRANIOTOMY PACK-LF DYNJ0101292F DYNJ0101292G HYBRID PACK DYNJ65925A LAMINECTOMY DYNJ9051…
- Z-2588-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. MAJOR BASIN SET DYNJ85723
- Z-2590-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CATARACT FULL BODY DYNJ85743 MAJOR VASCULAR PACK DYNJ81610C MINIMALLY INVASIVE PACK-LF DY…
- Z-2591-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIO PACK DYNJ30565C DYNJ32555B ANGIO PACK-LF DYNJ0774765U ANGIO TRAY DYNJ46153A ANGI…
- Z-2592-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIO DRAPE PACK (ADLUG)642-LF DYNJ47665D ANGIO PACK DYNJ30565D ANGIOGRAPHY PACK-LF DYNJ0…
- Z-2593-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ABDOMINAL VASCULAR-LF DYNJ905291B DYNJ905291C ARTERIOGRAM DYNJ28082G CVICU HEART BAG DYK…
- Z-2594-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CENTRAL LINE TRAY DYNJ41089 FISTULAGRAM PACK DYNJ62827A FISTULOGRAM PACK DYNJ47899A INV…
- Z-2595-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. PICC LINEPACK (PCLUI)642-LF DYNJ47717A
- Z-2596-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. FTMC C-SECTION TRAY-LF DYNJ54603B
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page