Convenience kits containing select SKUs of 10mL… recallZ-2593-2026 · Medline Industries, LP · initiated Feb 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
FDA root cause category: No Marketing Application
What the company is doing
This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.
Recall details
- Recall number
- Z-2593-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 27, 2026
- Posted by the FDA
- Jun 25, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 113, 843 kits
- Product code
- OEZ
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
Product description
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ABDOMINAL VASCULAR-LF DYNJ905291B DYNJ905291C ARTERIOGRAM DYNJ28082G CVICU HEART BAG DYKM2130A ENDOVASCULAR PACK DYNJ0678934J EP PACK DYNJ23456J GENERAL ENDO PACK-LF DYNJ0842873K HYBRID PACK DYNJ65925B DYNJ65925C NON VASCULAR PACK DYNJ68037 PACEMAKER DYNJ901681D PACEMAKER PACK DYNJ51853C DYNJ51853D DYNJ60606A PACER PACK DYNJ40952A RADIOLOGY VASCULAR ACCESS TRAY DYNJVB1303A SM- INTERVENTIONAL DRAPE PK-LF DYNJ0780143K VASCULAR ACCESS PACK-LF DYNJ0101291D VASCULAR DSC DYNJ69172 DYNJ69172B VASCULAR HYBRID DYNJ57543F DYNJ57543I DYNJ57543J VASCULAR PACK DYNJ61702A
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
DYNJ905291B UDI-DI 10193489950373 Lots 21GBO405 21JBX734 21LBO216 DYNJ905291C UDI-DI 10195327062545 Lots 22BBB434 22CBT553 22CBT556 22FBE950 22GBZ363 22LBQ583 23ABE635 23CBI463 DYNJ28082G UDI-DI 10195327050559 Lots 22BBL073 22DBT712 22FBU975 22IBD336 22LBH042 23BBG319 23CBW711 23EBF014 DYKM2130A UDI-DI 10195327630904 Lot 24ELA817 DYNJ0678934J UDI-DI 10195327637378 Lot 24FMI941 DYNJ23456J UDI-DI 10193489449839 Lot 21BMB927 DYNJ0842873K UDI-DI 10195327286613 Lots 22KDB463 23ADA210 23CDA339 23CDB773 23EDB190 23FDB457 23HDA476 23IDC327 23JDB182 23JDC057 24ADB081 24BDB286 24CDB934 24GDA042 24HDA590 24HDC238 DYNJ65925B UDI-DI 10193489925326 Lots 21FLA630 21GLA415 21HLA270 21JLA166 21KLA318 22CLA057 22CLA817 22DLA206 22ELA964 DYNJ65925C UDI-DI 10195327201371 Lots 23CLB075 23ELA853 23HLA442 23ILA022 DYNJ68037 UDI-DI 10193489473889 Lots 21ABT991 21HBL486 DYNJ901681D UDI-DI 10193489308099 Lot 24EMH538 DYNJ51853C UDI-DI 10193489269000 Lots 21CBB992 21DBA936 21FBE790 21HBN274 21KBA699 21KBP057 DYNJ51853D UDI-DI 10195327101336 Lots 22EBB104 22HBF054 22IBE507 22JBJ069 22KBS971 22LBT245 22NBG258 22OBI766 23ABI851 DYNJ60606A UDI-DI 10193489693386 Lot 24BBA469 DYNJ40952A UDI-DI 10193489941906 Lots 21GBR992 21JBX464 22BBS266 22CBP195 22EBS044 DYNJVB1303A UDI-DI 10193489928358 Lots 23ELB088 23FLA355 DYNJ0780143K UDI-DI 10195327477240 Lot 24AMI797 DYNJ0101291D UDI-DI 10193489381931 Lots 21CDC066 21EDB518 21GDC932 21HDC814 22BDA495 22DDB731 22DDC457 22EDB642 22GDA964 22JDA001 22JDC181 DYNJ69172 UDI-DI 10193489899603 Lots 21GMA692 21IMH791 DYNJ69172B UDI-DI 10195327487027 Lot 23KMC637 DYNJ57543F UDI-DI 10195327279028 Lot 23AMF931 DYNJ57543I UDI-DI 10195327359362 Lot 23CMD266 DYNJ57543J UDI-DI 10195327416027 Lots 23GMD945 23GMG266 DYNJ61702A UDI-DI 10193489470321 Lots 21CBN901 21FBM858 21GBE705
Devices named in this recall
22 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ901681DMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ905291CMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ57543FMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ0842873KMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ65925CMedline Industries, Inc. · Cardiovascular procedure kit1 recall1 report
- Medline Industries, Inc. DYNJ51853CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ0101291DMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61702AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ68037Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ60606AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ69172Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ65925BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJVB1303AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ40952AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ28082GMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ51853DMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ57543IMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ57543JMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ0780143KMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ69172BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYKM2130AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ0678934JMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99062).
- Z-2581-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAM PACK-LF DYNJ21574J JUDKINS PACK DYNJ51126
- Z-2582-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIO CATH COMBINED PACK DYNJ43609S ANGIO PACK DYNJ67987D DYNJ80317A ANGIO SPECIAL PROCEDURE-LF D…
- Z-2583-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. BARIATRIC DYNJ905153J DYNJ905153K DYNJ905153L GYNE LAPAROTOMY PACK-LF DYNJ49593A LAPAROSCOPY …
- Z-2584-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. DR G'S BREAST PACK DYNJ51124B LIPO SUPPLEMENT PACK-LF DYNJ0341174D PLASTIC PACK DYNJ66746…
- Z-2585-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. AAA DYNJ905651B ARTERIOGRAM TRAY 00-402001S AV SHUNT PACK-LF DYNJ0842516F BASIC IR PACK…
- Z-2586-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAPHY PACK-LF DYNJ0854485S DEEP BRAIN STIMULATION PACK DYNJ0842793G NEURO ACCESSOR…
- Z-2587-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CRANIOTOMY PACK-LF DYNJ0101292F DYNJ0101292G HYBRID PACK DYNJ65925A LAMINECTOMY DYNJ9051…
- Z-2588-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. MAJOR BASIN SET DYNJ85723
- Z-2590-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CATARACT FULL BODY DYNJ85743 MAJOR VASCULAR PACK DYNJ81610C MINIMALLY INVASIVE PACK-LF DY…
- Z-2591-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIO PACK DYNJ30565C DYNJ32555B ANGIO PACK-LF DYNJ0774765U ANGIO TRAY DYNJ46153A ANGI…
- Z-2592-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIO DRAPE PACK (ADLUG)642-LF DYNJ47665D ANGIO PACK DYNJ30565D ANGIOGRAPHY PACK-LF DYNJ0…
- Z-2594-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CENTRAL LINE TRAY DYNJ41089 FISTULAGRAM PACK DYNJ62827A FISTULOGRAM PACK DYNJ47899A INV…
- Z-2595-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. PICC LINEPACK (PCLUI)642-LF DYNJ47717A
- Z-2596-2026 · Open, ClassifiedConvenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. FTMC C-SECTION TRAY-LF DYNJ54603B
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page