Skip to content

Convenience kits containing select SKUs of 10mL… recallZ-2590-2026 · Medline Industries, LP · initiated Feb 27, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

FDA root cause category: No Marketing Application

What the company is doing

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Recall details

Recall number
Z-2590-2026
Status
Open, Classified
Initiated by the company
Feb 27, 2026
Posted by the FDA
Jun 25, 2026
Recalling firm
Medline Industries, LP
Quantity
113,843 kits
Product code
LRO
FDA submissions
K213481
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.

Product description

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CATARACT FULL BODY DYNJ85743 MAJOR VASCULAR PACK DYNJ81610C MINIMALLY INVASIVE PACK-LF DYNJ0843063L PEDIATRIC CATH ANGIO PACK DYNJ54755J PERIPHERAL ARTERIOGRAM RAD DYNJ61383C POSTERIOR SPINE PACK-LF DYNJ0843339O UPPER EXTREMITY DYNJ61907A VISCERAL ANGIOGRAM PACK-LF DYNJ0429121L

Code information (models, lots, serial numbers)

DYNJ85743 UDI-DI 10195327542337 Lots 24ABP082 24BBA479 DYNJ81610C UDI-DI 10195327556938 Lot 24BBB405 DYNJ0843063L UDI-DI 10195327567071 Lots 23LDB992 24BDA641 24BDA782 24FDA795 24GDB014 24HDB074 DYNJ54755J UDI-DI 10195327549909 Lots 24ABJ741 24CBI600 24EBA488 24FBM084 24GBN355 24HBH571 DYNJ61383C UDI-DI 10195327593346 Lot 24CLA801 DYNJ0843339O UDI-DI 10195327567057 Lots 24CDB480 24DDA248 24GDB068 DYNJ61907A UDI-DI 10195327542450 Lot 24ABP069 DYNJ0429121L UDI-DI 10195327587697 Lot 24BMH315

Devices named in this recall

8 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

321 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99062).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page