Skip to content

Convenience kits containing select SKUs of 10mL… recallZ-2594-2026 · Medline Industries, LP · initiated Feb 27, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

FDA root cause category: No Marketing Application

What the company is doing

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Recall details

Recall number
Z-2594-2026
Status
Open, Classified
Initiated by the company
Feb 27, 2026
Posted by the FDA
Jun 25, 2026
Recalling firm
Medline Industries, LP
Quantity
113, 843 kits
Product code
OFF
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.

Product description

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CENTRAL LINE TRAY DYNJ41089 FISTULAGRAM PACK DYNJ62827A FISTULOGRAM PACK DYNJ47899A INVASIVE LINE INSERTION DYNJ44123B NON VASCULAR PACK DYNJ68037A VASCULAR ACCESS PACK-LF DYNJ0101291F

Code information (models, lots, serial numbers)

DYNJ41089 UDI-DI 10884389891360 Lot 21BBB592 DYNJ62827A UDI-DI 10193489306606 Lot 24BBN289 DYNJ47899A UDI-DI 10889942574681 Lots 21GBI377 21IBL299 21LBE099 21LBU529 22JBV803 DYNJ44123B UDI-DI 10195327381653 Lots 23FBR833 23GBN041 23HBW014 23JBO264 23JBW218 23KBI873 DYNJ68037A UDI-DI 10195327003630 Lots 21JBQ304 21LBQ205 22ABN185 22DBT349 22EBL611 22HBH550 22IBN880 22JBH689 22KBS987 22OBA368 23ABD642 23BBH830 23DBL865 23EBT367 23FBM946 23GBT261 DYNJ0101291F UDI-DI 10195327286644 Lots 22KDB751 23EDA119 23EDB388 23FDA396 23GDA313 23HDA759 23LDB909 24ADB667 24GDA294 24HDB515

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99062).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page