Skip to content

GORE VIABIL Biliary Endoprosthesis recallZ-2915-2026 · W L Gore & Associates, Inc. · initiated May 22, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from W L Gore & Associates, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010).

FDA root cause category: Process control

What the company is doing

On 05/22/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was sent via email to customers informing them that, 10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010). Customers are instructed to: . Identify and return any unused devices within the scope of this recall for replacement. . Complete and sign the enclosed CUSTOMER RESPONSE / DEVICE RECONCILIATION FORM and return to [email protected] within 2 weeks of receipt of this notification. . This notice needs to be shared with those who should be aware within your institution or to any organization where potentially affected devices have been transferred (as appropriate). . Implanting Physicians: If you have implanted one of the potentially mislabeled GORE VIABIL Short Wire Biliary Endoprosthesis devices, please be mindful of the potential health outcomes and plan to take the needed measures that could minimize any potential health hazards associated with this mislabeling. If a listed device has been used, please indicate the used device(s) on the CUSTOMER RESPONSE / DEVICE RECONCILIATION FORM and return to [email protected] For questions or further assistance - contact Julie Skaife, Global Product Specialist & Brand Manager at [email protected].

Recall details

Recall number
Z-2915-2026
Status
Open, Classified
Initiated by the company
May 22, 2026
Posted by the FDA
Aug 7, 2026
Recalling firm
W L Gore & Associates, Inc.
Quantity
13 units
Product code
FGE
FDA submissions
K170740
Distribution
U.S. Nationwide distribution in the states of GA and RI.

Product description

GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.

Code information (models, lots, serial numbers)

Incorrect Catalogue Number/UDI: VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796, VSWVH1006 / 00733132638796; Lot Serial No. 30970948, 30970949, 30970950, 30970951, 30970952, 30970953, 30970954, 30970955, 30970956, 30970957, 30970958, 30970959, 30970960.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

6 devices cleared under K170740 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99143).

Sources and updates