GORE VIABIL Biliary Endoprosthesis VSWVH1006Recalls & reported problems
W. L. Gore & Associates, Inc. · VIABIL SHORT WIRE BILIARY ENDO 10mmX6cmX200cm 8.5Fr · Class 2 · Implant
All models of GORE VIABIL Biliary Endoprosthesis (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132638796), received between Aug 23, 2021 and Jan 3, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K170740); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 22, 2026Class IIOpen, Classified
10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010).
Z-2915-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture2
- Unraveled Material2
- Break1
- Difficult to Remove1
- Improper or Incorrect Procedure or Method1
- Material Puncture/Hole1
- Off-Label Use1
Patient problems
- Unspecified Hepatic or Biliary Problem2
- Biliary Cirrhosis1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 3, 2024Fracture; Improper or Incorrect Procedure or Method; Fracture; Improper or Incorrect Procedure or Method
IT WAS REPORTED TO GORE BY CONMED ((B)(6)) THAT DURING AN UNKNOWN PROCEDURE USING THE GORE® VIABIL® SHORT WIRE BILIARY ENDOPROSTHESIS, THE STENT BROKE INTO PIECES AND ALL PIECES WERE EVENTUALLY REMOVED. THE DEVICE IS NOT AVAILABLE FOR RETURN. THERE IS NO INDICATION OF PATIENT IMPACT. IT WAS FURTHER REPORTED THIS CASE WAS A PLANNED REMOVAL. THE DEVICE WAS PLACED IN THE COMMON BILE DUCT ON (B)(6), 2023. THERE WAS NO OUT OF THE ORDINARY MANIPULATION OF THE ENDOSCOPE OR CATHETER DURING THE REMOVAL. A DUODENOSCOPE WAS USED AND THE INITIAL ATTEMPT WAS WITH A SNARE, WHICH IS WHEN THE STENT FRACTURED. MULTIPLE ATTEMPTS WITH THE SNARE AND RAPTOR GRASPER WERE MADE. THE DEVICE WAS DISCARDED AFTER THE CASE.
Patient problems: Unspecified Hepatic or Biliary Problem
MAUDE report 18440598 ↗MalfunctionAug 19, 2022Break; Fracture; Off-Label Use; Material Puncture/Hole; Difficult to Remove; Unraveled Material; Fracture; Off-Label Use; Material Puncture/Hole; Difficult to Remove
IT WAS REPORTED TO GORE BY CONMED (CEN-49927) THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE USING THE GORE® VIABIL® BILIARY ENDOPROSTHESIS, THE DEVICE UNRAVELED UPON REMOVAL ATTEMPT. REPORTEDLY, ALL PIECES WERE REMOVED AND NO PAIN WAS CAUSED TO THE PATIENT. IT WAS REPORTED THE DEVICE WAS DISCARDED AT THE FACILITY AND THIS WAS FOUND DURING THE PROCEDURE. REPORTEDLY, THE PROCEDURE BEING PERFORMED WAS A STENT REMOVAL FOR AN UNKNOWN REASON. IT WAS REPORTED AND CONFIRMED THE STENT UNRAVELED AND THERE WAS BREAKAGE, RESULTING IN A STENT WIRE FRACTURE. REPORTEDLY, THE ANATOMICAL LOCATION OF THE DEVICE WHEN IT WOULD NOT DEPLOY IS UNKNOWN, IT WAS REPORTED THE LENGTH, DIAMETER AND STIFFNESS OF THE GUIDEWIRE IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Hepatic or Biliary Problem
MAUDE report 15259791 ↗InjuryAug 23, 2021Unraveled Material; Unraveled Material
IT WAS REPORTED TO GORE BY CONMED ((B)(4)) THAT DURING AN ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY). THE GORE¿ VIABIL¿ BILIARY ENDOPROSTHESIS DISTAL 4 CM, UNRAVELED ON REMOVAL. REPORTEDLY THE REMAINING PROXIMAL 2 CM STENT IS EMBEDDED INTO THE TISSUE AND WAS NOT ABLE TO BE REMOVED. IT WAS REPORTED THAT THE PROXIMAL 2 CM OF STENT REMAINS IN THE DUCT. REPORTEDLY THERE WAS A 7 HOUR DELAY IN THE PROCEDURE. REPORTEDLY A PLASTIC STENT WAS PLACED TO KEEP THE DUCT PATENT. IT WAS REPORTED THAT A REMOVAL PROCEDURE WAS SCHEDULED FOR (B)(6) 2021. REPORTEDLY THE PATIENT WENT HOME.
Patient problems: Biliary Cirrhosis
MAUDE report 12357120 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170740. One submission often covers many models, so check the product and lot details in each record.
- Z-0158-2027 · Sep 3, 2026 · Open, ClassifiedGeneric Name: GORE VIABIL Short Wire Biliary Endoprosthesis; Model/Catalog Number: VSWVH0808 and VSWVN1006 Product Description: The GORE VIABIL Short Wire Biliary Endoprosthesis
- Z-2914-2026 · May 22, 2026 · Open, ClassifiedGORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1010. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may b
- Z-1179-2024 · Jan 19, 2024 · Open, ClassifiedGORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2) Catalog Number VSWVN1008, 10 mm x 8 cm; no holes
Other Catheter, Biliary, Diagnostic devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page