Level 1 recallZ-2759-2026 · ICU Medical Inc. · initiated Jun 24, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ICU Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.

FDA root cause category: Process control

What the company is doing

An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 6/24/26 was sent to customers. Actions for Users: When using the device, all instructions, including warnings and cautions contained in the Instructions for Use must be followed with heightened awareness. Please complete the following actions below. 1. Locate potentially affected devices from the Lots listed in Table 1 above. 2. Destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 3. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure that this communication is delivered there. 4. Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification. Required Actions for Distributors: When using the device, all instructions, including warnings and cautions contained in the Instructions for Use must be followed with heightened awareness. Please complete the following actions below. 1. Locate potentially affected devices from the Lots listed in Table 1 above. 2. Destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 3. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure that this communication is delivered there. 4. Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification. 5. DISTRIBUTORS: Immediately forward this notice to all users to whom you distributed these products. IC

Recall details

Recall number
Z-2759-2026
Status
Open, Classified
Initiated by the company
Jun 24, 2026
Posted by the FDA
Jul 17, 2026
Recalling firm
ICU Medical Inc.
Quantity
39,750 units
Product code
LGZ
FDA submissions
K911383
Distribution
Worldwide - US Nationwide distribution.

Product description

Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70

Code information (models, lots, serial numbers)

UDI-DI: 30695085407007; Lot Numbers: 4293568, 4293569, 4297510, 4297509, 4326087, 4337724, 4337725, 4346349

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

9 devices cleared under K911383 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.