Allura Xper FD recallZ-2928-2026 · Philips Medical Systems Nederland B.V. · initiated Jun 29, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firmware issue potentially leading to loss of geometry movements.
FDA root cause category: Software design
What the company is doing
Philips notified consignees on about 06/29/2026 via certified mailed letter titled URGENT Medical Device Correction. Consignees were provided instructions to properly identify affected systems and table configurations, instructions for restarting the geometry when the issue occurs, to ensure the notification is distributed to all relevant users and retained with device documentation until corrective actions are completed. Consignees were also instructed to post the notification letter near affected devices for ease of reviewing, notify recipients if affected units were transferred, and to complete and return the provided response form. A Local Philips representative will contact consignees to schedule a visit for the implementation of the software update once available. Philips will address the identified issues by implementing a software update (R8.2.102.1) in all affected systems (FCO72200674). Philips currently expects R8.2.102.1 to be available by Q1 2027, subject to regulatory clearance.
Recall details
- Recall number
- Z-2928-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 29, 2026
- Posted by the FDA
- Aug 10, 2026
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 4,252 units
- Product code
- OWB
- FDA submissions
- K133292, K141979, K161563, K162859
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Lao, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Uzbekistan, Vietnam, Yemen.
Product description
Allura Xper FD. Model Numbers: 722026, 722027, 722028, 722029, 722035, 722038, 722039, 722058. Interventional fluoroscopic x-ray system.
Code information (models, lots, serial numbers)
Show the full code information (26k characters)
Model Numbers: 722026, 722027, 722028, 722029, 722035, 722038, 722039, 722058. Software Version: R8.2. UDI Numbers: (01)00884838054202(21)1756, (01)00884838054202(21)1569, (01)00884838059115(21)100, (01)00884838059115(21)239, (01)00884838054202(21)520, (01)00884838054202(21)1956, (01)00884838054189(21)683, (01)00884838054196(21)351, (01)00884838054202(21)129, (01)00884838059115(21)305, (01)00884838054196(21)332, (01)00884838054202(21)1233, (01)00884838054202(21)2206, (01)00884838054202(21)1690, (01)00884838054202(21)928, (01)00884838054189(21)155, (01)00884838054202(21)317, (01)00884838054202(21)1774, (01)00884838054202(21)2502, (01)00884838054202(21)869, (01)00884838054202(21)668, (01)00884838054189(21)618, (01)00884838054189(21)1051, (01)00884838054189(21)1052, (01)00884838054202(21)406, (01)00884838054202(21)540, (01)00884838054202(21)548, (01)00884838054202(21)1017, (01)00884838054202(21)590, (01)00884838054189(21)654, (01)00884838054189(21)786, (01)00884838054196(21)216, (01)00884838054196(21)230, (01)00884838054202(21)596, (01)00884838054202(21)1957, (01)00884838054202(21)110, (01)00884838054202(21)2261, (01)00884838054189(21)1132, (01)00884838054226(21)115, (01)00884838059115(21)259, (01)00884838054202(21)173, (01)00884838054202(21)866, (01)00884838054202(21)1310, (01)00884838059115(21)203, (01)00884838054202(21)1043, (01)00884838054189(21)458, (01)00884838054202(21)466, (01)00884838054226(21)42, (01)00884838054202(21)1077, (01)00884838059115(21)373, (01)00884838054202(21)877, (01)00884838054202(21)896, (01)00884838054202(21)987, (01)00884838054202(21)995, (01)00884838054202(21)1343, (01)00884838054202(21)1439, (01)00884838059115(21)111, (01)00884838054202(21)895, (01)00884838054202(21)265, (01)00884838054202(21)305, (01)00884838054189(21)1100, (01)00884838054189(21)770, (01)00884838054196(21)1, (01)00884838054202(21)1154, (01)00884838054196(21)71, (01)00884838054202(21)1000, (01)00884838054202(21)1909, (01)00884838054202(21)121, (01)00884838054189(21)436, (01)00884838054202(21)1716, (01)00884838054202(21)1983, (01)00884838054202(21)784, (01)00884838054202(21)2061, (01)00884838054202(21)729, (01)00884838054189(21)1338, (01)00884838054202(21)1393, (01)00884838054189(21)1068, (01)00884838054189(21)233, (01)00884838054202(21)525, (01)00884838054202(21)1948, (01)00884838054202(21)1656, (01)00884838054189(21)78, (01)00884838054202(21)209, (01)00884838054202(21)825, (01)00884838054202(21)1461, (01)00884838054202(21)408, (01)00884838059115(21)38, (01)00884838054202(21)570, (01)00884838054202(21)218, (01)00884838054202(21)734, (01)00884838054202(21)1050, (01)00884838054202(21)1751, (01)00884838054202(21)454, (01)00884838054202(21)1428, (01)00884838054219(21)57, (01)00884838054189(21)432, (01)00884838054202(21)666, (01)00884838054219(21)7, (01)00884838054189(21)224, (01)00884838054202(21)488, (01)00884838054226(21)108, (01)00884838054202(21)1094, (01)00884838054202(21)169, (01)00884838059115(21)23, (01)00884838054202(21)2173, (01)00884838054202(21)1214, (01)00884838054202(21)52, (01)00884838054202(21)66, (01)00884838054202(21)493, (01)00884838054202(21)994, (01)00884838054202(21)373, (01)00884838059115(21)290, (01)00884838054202(21)1213, (01)00884838054226(21)173, (01)00884838054202(21)339, (01)00884838054202(21)234, (01)00884838054189(21)93, (01)00884838054189(21)796, (01)00884838054189(21)1072, (01)00884838054202(21)1836, (01)00884838054202(21)364, (01)00884838054202(21)982, (01)00884838054202(21)634, (01)00884838059115(21)124, (01)00884838054202(21)190, (01)00884838054202(21)818, (01)00884838054202(21)1394, (01)00884838054202(21)1396, (01)00884838054202(21)1646, (01)00884838054202(21)386, (01)00884838054226(21)119, (01)00884838054226(21)126, (01)00884838054202(21)1650, (01)00884838054189(21)543, (01)00884838054202(21)2275, (01)00884838054189(21)1248, (01)00884838054202(21)1058, (01)00884838054189(21)561, (01)00884838059115(21)274, (01)00884838054202(21)1353, (01)00884838054202(21)1466, (01)00884838054196(21)319, (01)00884838054202(21)71, (01)00884838059115(21)340, (01)00884838059115(21)366, (01)00884838054189(21)1129, (01)00884838054202(21)910, (01)00884838054189(21)781, (01)00884838054202(21)751, (01)00884838054202(21)1340, (01)00884838054202(21)1978, (01)00884838054189(21)344, (01)00884838054189(21)374, (01)00884838054189(21)634, (01)00884838054202(21)254, (01)00884838054202(21)326, (01)00884838054202(21)166, (01)00884838054202(21)431, (01)00884838054189(21)151, (01)00884838054189(21)908, (01)00884838054202(21)11, (01)00884838054202(21)318, (01)00884838054202(21)1211, (01)00884838054219(21)88, (01)00884838054196(21)43, (01)00884838054196(21)101, (01)00884838054202(21)543, (01)00884838054202(21)1651, (01)00884838054189(21)130, (01)00884838054196(21)29, (01)00884838054196(21)50, (01)00884838054202(21)131, (01)00884838054202(21)133, (01)00884838054202(21)138, (01)00884838054202(21)547, (01)00884838054202(21)550, (01)00884838054219(21)56, (01)00884838054202(21)1326, (01)00884838054202(21)2710, (01)00884838054189(21)370, (01)00884838054202(21)267, (01)00884838054202(21)759, (01)00884838054189(21)235, (01)00884838054189(21)190, (01)00884838054202(21)292, (01)00884838054202(21)821, (01)00884838054189(21)259, (01)00884838054202(21)1232, (01)00884838054189(21)323, (01)00884838054189(21)324, (01)00884838054189(21)969, (01)00884838054202(21)925, (01)00884838054189(21)98, (01)00884838054189(21)99, (01)00884838054189(21)100, (01)00884838054202(21)127, (01)00884838054202(21)132, (01)00884838054202(21)141, (01)00884838054202(21)145, (01)00884838054202(21)180, (01)00884838054202(21)160, (01)00884838054202(21)468, (01)00884838054202(21)961, (01)00884838054189(21)142, (01)00884838054202(21)1736, (01)00884838059115(21)162, (01)00884838054202(21)1619, (01)00884838054189(21)1246, (01)00884838054202(21)2933, (01)00884838054189(21)601, (01)00884838054196(21)112, (01)00884838054202(21)1567, (01)00884838054202(21)289, (01)00884838054202(21)282, (01)00884838054202(21)1130, (01)00884838059115(21)186, (01)00884838054202(21)1100, (01)00884838054202(21)1570, (01)00884838054202(21)529, (01)00884838054202(21)1385, (01)00884838059115(21)4, (01)00884838054189(21)877, (01)00884838054196(21)240, (01)00884838054196(21)241, (01)00884838054202(21)1267, (01)00884838059115(21)196, (01)00884838059115(21)405, (01)00884838054202(21)464, (01)00884838054202(21)2103, (01)00884838054202(21)1395, (01)00884838054202(21)1408, (01)00884838054202(21)280, (01)00884838054202(21)741, (01)00884838054202(21)211, (01)00884838054202(21)261, (01)00884838054202(21)1269, (01)00884838054202(21)1837, (01)00884838054219(21)95, (01)00884838059115(21)156, (01)00884838054202(21)601, (01)00884838054202(21)854, (01)00884838054202(21)1051, (01)00884838054202(21)1988, (01)00884838054226(21)60, (01)00884838059115(21)146, (01)00884838054202(21)744, (01)00884838054202(21)956, (01)00884838054189(21)204, (01)00884838054202(21)2020, (01)00884838054219(21)126, (01)00884838054202(21)2601, (01)00884838054202(21)1261, (01)00884838054202(21)1835, (01)00884838054202(21)559, (01)00884838054202(21)967, (01)00884838054189(21)843, (01)00884838054189(21)1038, (01)00884838054189(21)1231, (01)00884838054196(21)175, (01)00884838054196(21)298, (01)00884838054189(21)114, (01)00884838054202(21)1483, (01)00884838054219(21)94, (01)00884838054202(21)2846, (01)00884838059115(21)285, (01)00884838054202(21)526, (01)00884838054202(21)2745, (01)00884838054189(21)1287, (01)00884838054202(21)2249, (01)00884838054189(21)878, (01)00884838054202(21)1103, (01)00884838054202(21)1311, (01)00884838054202(21)880, (01)00884838054202(21)946, (01)00884838054202(21)949, (01)00884838054189(21)465, (01)00884838054202(21)518, (01)00884838054202(21)519, (01)00884838054202(21)2645, (01)00884838054202(21)731, (01)00884838054202(21)678, (01)00884838054202(21)1116, (01)00884838054202(21)1880, (01)00884838054202(21)911, (01)00884838054202(21)118, (01)00884838054202(21)1705, (01)00884838054202(21)2192, (01)00884838054202(21)142, (01)00884838054226(21)37, (01)00884838054226(21)38, (01)00884838054202(21)376, (01)00884838054189(21)558, (01)00884838054202(21)67, (01)00884838059115(21)277, (01)00884838054202(21)273, (01)00884838054202(21)1171, (01)00884838054202(21)1498, (01)00884838054202(21)1156, (01)00884838054202(21)1279, (01)00884838054189(21)391, (01)00884838054189(21)389, (01)00884838054202(21)522, (01)00884838054226(21)99, (01)00884838054202(21)564, (01)00884838059115(21)65, (01)00884838054189(21)633, (01)00884838054202(21)158, (01)00884838054202(21)1708, (01)00884838054202(21)1710, (01)00884838054202(21)558, (01)00884838054202(21)521, (01)00884838054202(21)664, (01)00884838054226(21)130, (01)00884838054202(21)906, (01)00884838054202(21)1155, (01)00884838054202(21)1085, (01)00884838054202(21)811, (01)00884838054196(21)280, (01)00884838054202(21)462, (01)00884838054202(21)677, (01)00884838054202(21)316, (01)00884838054202(21)1946, (01)00884838054202(21)184, (01)00884838054202(21)185, (01)00884838054226(21)48, (01)00884838054189(21)1050, (01)00884838054202(21)1834, (01)00884838054202(21)675, (01)00884838054226(21)41, (01)00884838054226(21)157, (01)00884838054271(21)3, (01)00884838054202(21)1652, (01)00884838054189(21)909, (01)00884838054189(21)1000, (01)00884838054189(21)600, (01)00884838054202(21)1011, (01)00884838054202(21)203, (01)00884838054202(21)204, (01)00884838059115(21)168, (01)00884838054202(21)1508, (01)00884838054189(21)841, (01)00884838054196(21)107, (01)00884838054202(21)346, (01)00884838054202(21)2615, (01)00884838054202(21)1890, (01)00884838054202(21)2293, (01)00884838054202(21)618, (01)00884838054189(21)185, (01)00884838054202(21)262, (01)00884838054202(21)580, (01)00884838054196(21)183, (01)00884838054196(21)222, (01)00884838054202(21)654, (01)00884838059115(21)56, (01)00884838054202(21)1149, (01)00884838054202(21)1484, (01)00884838054202(21)1995, (01)00884838054202(21)1397, (01)00884838054202(21)1456, (01)00884838054189(21)394, (01)00884838054189(21)713, (01)00884838059115(21)178, (01)00884838054219(21)30, (01)00884838054189(21)731, (01)00884838054202(21)274, (01)00884838054202(21)1288, (01)00884838054202(21)2906, (01)00884838054196(21)41, (01)00884838054202(21)402, (01)00884838054202(21)1556, (01)00884838054202(21)1201, (01)00884838054202(21)1129, (01)00884838054226(21)141, (01)00884838054202(21)976, (01)00884838054189(21)522, (01)00884838059115(21)138, (01)00884838054202(21)479, (01)00884838054202(21)2348, (01)00884838054202(21)2641, (01)00884838054189(21)75, (01)00884838054202(21)447, (01)00884838054202(21)805, (01)00884838054202(21)922, (01)00884838054189(21)646, (01)00884838054196(21)202, (01)00884838054202(21)973, (01)00884838054202(21)1068, (01)00884838054202(21)1069, (01)00884838054226(21)137, (01)00884838054202(21)371, (01)00884838054202(21)23040202, (01)00884838054202(21)167, (01)00884838054189(21)728, (01)00884838054202(21)485, (01)00884838054219(21)31, (01)00884838054202(21)1095, (01)00884838054202(21)1870, (01)00884838054202(21)2253, (01)00884838054189(21)223, (01)00884838054202(21)892, (01)00884838059115(21)212, (01)00884838054202(21)130, (01)00884838054202(21)2295, (01)00884838054202(21)2071, (01)00884838054202(21)1327, (01)00884838054202(21)903, (01)00884838054202(21)1099, (01)00884838059115(21)135, (01)00884838054189(21)765, (01)00884838054226(21)49, (01)00884838054196(21)164, (01)00884838054196(21)153, (01)00884838054202(21)1407, (01)00884838054202(21)620, (01)00884838054202(21)176, (01)00884838054202(21)257, (01)00884838054202(21)1920, (01)00884838054202(21)205, (01)00884838054202(21)2707, (01)00884838054202(21)2709, (01)00884838054202(21)2174, (01)00884838054202(21)819, (01)00884838054202(21)151, (01)00884838054202(21)539, (01)00884838054202(21)1275, (01)00884838059115(21)257, (01)00884838054202(21)656, (01)00884838054202(21)361, (01)00884838054202(21)721, (01)00884838054202(21)1142, (01)00884838054219(21)21, (01)00884838054189(21)192, (01)00884838054202(21)779, (01)00884838054202(21)1111, (01)00884838054202(21)463, (01)00884838054202(21)1122, (01)00884838054189(21)959, (01)00884838054202(21)481, (01)00884838054202(21)806, (01)00884838054219(21)97, (01)00884838054202(21)310, (01)00884838054202(21)491, (01)00884838054202(21)2151, (01)00884838059115(21)337, (01)00884838054189(21)825, (01)00884838054196(21)335, (01)00884838054189(21)942, (01)00884838054202(21)1352, (01)00884838054189(21)1192, (01)00884838054202(21)888, (01)00884838054202(21)697, (01)00884838054202(21)1755, (01)00884838054189(21)990, (01)00884838054202(21)1855, (01)00884838054202(21)445, (01)00884838054202(21)80, (01)00884838054189(21)381, (01)00884838054202(21)840, (01)00884838054202(21)842, (01)00884838054202(21)2310, (01)00884838054202(21)2506, (01)00884838054226(21)63, (01)00884838054202(21)655, (01)00884838054202(21)691, (01)00884838054202(21)379, (01)00884838054189(21)774, (01)00884838054202(21)553, (01)00884838054202(21)377, (01)00884838054202(21)632, (01)00884838054226(21)35, (01)00884838054202(21)638, (01)00884838059115(21)265, (01)00884838054202(21)2007, (01)00884838054202(21)2008, (01)00884838054202(21)968, (01)00884838054202(21)781, (01)00884838054189(21)368, (01)00884838054202(21)154, (01)00884838054202(21)774, (01)00884838054189(21)245, (01)00884838054202(21)1850, (01)00884838054202(21)1522, (01)00884838054196(21)288, (01)00884838054189(21)484, (01)00884838054189(21)485, (01)00884838054189(21)648, (01)00884838054196(21)342, (01)00884838059115(21)109, (01)00884838054189(21)966, (01)00884838054189(21)1268, (01)00884838054202(21)451, (01)00884838054196(21)134, (01)00884838054196(21)156, (01)00884838054189(21)109, (01)00884838054189(21)110, (01)00884838054196(21)265, (01)00884838054196(21)306, (01)00884838054196(21)307, (01)00884838054202(21)140, (01)00884838054202(21)1413, (01)00884838054202(21)1448, (01)00884838054202(21)1538, (01)00884838054189(21)116, (01)00884838054189(21)604, (01)00884838054189(21)627, (01)00884838059115(21)134, (01)00884838054189(21)319, (01)00884838054196(21)152, (01)00884838054196(21)106, (01)00884838054196(21)169, (01)00884838059115(21)71, (01)00884838054202(21)1930, (01)00884838059115(21)238, (01)00884838054202(21)743, (01)00884838054202(21)340, (01)00884838054202(21)535, (01)00884838054202(21)938, (01)00884838054202(21)31, (01)00884838054202(21)206, (01)00884838059115(21)190, (01)00884838054189(21)112, (01)00884838054196(21)42, (01)00884838054202(21)251, (01)00884838054219(21)22, (01)00884838054189(21)716, (01)00884838054196(21)203, (01)00884838054219(21)91, (01)00884838054189(21)396, (01)00884838054226(21)167, (01)00884838054202(21)990, (01)00884838054202(21)1081, (01)00884838054202(21)808, (01)00884838054202(21)1633, (01)00884838054202(21)1634, (01)00884838059115(21)219, (01)00884838054202(21)315, (01)00884838054226(21)71, (01)00884838054202(21)1387, (01)00884838054202(21)1630, (01)00884838054202(21)143, (01)00884838054189(21)531, (01)00884838054196(21)234, (01)00884838054202(21)586, (01)00884838054189(21)1182, (01)00884838054202(21)989, (01)00884838054202(21)1824, (01)00884838054189(21)156, (01)00884838054189(21)253, (01)00884838054202(21)236, (01)00884838054202(21)237, (01)00884838054202(21)238, (01)00884838054202(21)356, (01)00884838054202(21)1481, (01)00884838054226(21)96, (01)00884838054202(21)717, (01)00884838054189(21)965, (01)00884838054226(21)55, (01)00884838054202(21)887, (01)00884838054189(21)444, (01)00884838054189(21)1189, (01)00884838054202(21)1202, (01)00884838054202(21)541, (01)00884838054202(21)565, (01)00884838054202(21)416, (01)00884838054202(21)336, (01)00884838059115(21)197, (01)00884838054202(21)303, (01)00884838054202(21)476, (01)00884838054202(21)2407, (01)00884838054189(21)220, (01)00884838054189(21)258, (01)00884838054202(21)189, (01)00884838054202(21)2635, (01)00884838054189(21)1126, (01)00884838054202(21)1809, (01)00884838054202(21)538, (01)00884838054202(21)904, (01)00884838054189(21)1210, (01)00884838054202(21)785, (01)00884838054189(21)1239, (01)00884838054189(21)102, (01)00884838054189(21)1215, (01)00884838054202(21)2083, (01)00884838054202(21)350, (01)00884838054202(21)473, (01)00884838054189(21)696, (01)00884838054202(21)329, (01)00884838054202(21)549, (01)00884838054202(21)874, (01)00884838054189(21)11, (01)00884838054189(21)297, (01)00884838054189(21)560, (01)00884838054189(21)562, (01)00884838054202(21)669, (01)00884838054202(21)1049, (01)00884838054202(21)684, (01)00884838054202(21)1333, (01)00884838054202(21)582, (01)00884838054202(21)224, (01)00884838059115(21)325, (01)00884838054202(21)1264, (01)00884838054202(21)281, (01)00884838054202(21)475, (01)00884838054202(21)981, (01)00884838054202(21)1447, (01)00884838054189(21)832, (01)00884838054202(21)696, (01)00884838054202(21)1188, (01)00884838054202(21)1189, (01)00884838054202(21)966, (01)00884838054189(21)335, (01)00884838054189(21)664, (01)00884838054202(21)333, (01)00884838054189(21)845, (01)00884838054202(21)1875, (01)00884838054202(21)128, (01)00884838054189(21)778, (01)00884838054202(21)1405, (01)00884838054202(21)1482, (01)00884838054189(21)215, (01)00884838054202(21)930, (01)00884838054202(21)932, (01)00884838054202(21)498, (01)00884838054202(21)1001, (01)00884838054189(21)935, (01)00884838054202(21)647, (01)00884838054202(21)674, (01)00884838054202(21)864, (01)00884838054202(21)1041, (01)00884838054189(21)782, (01)00884838054202(21)1054, (01)00884838054202(21)698, (01)00884838054202(21)1032, (01)00884838054219(21)110, (01)00884838054196(21)209, (01)00884838054196(21)267, (01)00884838054219(21)63, (01)00884838054219(21)100, (01)00884838054202(21)2252, (01)00884838054189(21)614, (01)00884838054189(21)222, (01)00884838054202(21)260, (01)00884838054202(21)630, (01)00884838054196(21)284, (01)00884838054189(21)205, (01)00884838054202(21)275, (01)00884838054189(21)115, (01)00884838054189(21)284, (01)00884838054202(21)419, (01)00884838054189(21)694, (01)00884838054202(21)917, (01)00884838054202(21)1266, (01)00884838054202(21)1472, (01)00884838054202(21)2070, (01)00884838054219(21)60, (01)00884838054202(21)971, (01)00884838059115(21)24, (01)00884838054202(21)1014, (01)00884838059115(21)123, (01)00884838054196(21)263, (01)00884838054202(21)1200, (01)00884838054202(21)1374, (01)00884838054189(21)74, (01)00884838054202(21)1499, (01)00884838054202(21)1222, (01)00884838054189(21)838, (01)00884838054189(21)244, (01)00884838054202(21)426, (01)00884838054202(21)459, (01)00884838054189(21)973, (01)00884838054202(21)592, (01)00884838054196(21)127, (01)00884838054202(21)1514, (01)00884838054219(21)55, (01)00884838054226(21)70, (01)00884838054202(21)531, (01)00884838054202(21)456, (01)00884838054202(21)571, (01)00884838054189(21)985, (01)00884838054202(21)867, (01)00884838054202(21)891, (01)00884838054202(21)2110, (01)00884838059115(21)319, (01)00884838054189(21)101, (01)00884838054202(21)625, (01)00884838054202(21)452, (01)00884838054202(21)453, (01)00884838059115(21)378, (01)00884838054202(21)672, (01)00884838054202(21)136, (01)00884838054202(21)137, (01)00884838054226(21)39, (01)00884838054189(21)855, (01)00884838054202(21)530, (01)00884838054189(21)287, (01)00884838059115(21)61, (01)00884838054202(21)2200, (01)00884838054189(21)378, (01)00884838054202(21)48, (01)00884838054202(21)1783, (01)00884838054202(21)1350, (01)00884838054202(21)1723, (01)00884838054202(21)1454, (01)00884838054189(21)612, (01)00884838054202(21)2922, (01)00884838059115(21)205, (01)00884838054202(21)324, (01)00884838054189(21)725, (01)00884838054219(21)62, (01)00884838054202(21)1098, (01)00884838054189(21)914, (01)00884838059115(21)149, (01)00884838054202(21)256, (01)00884838059115(21)264, (01)00884838054202(21)983, (01)00884838059115(21)167, (01)00884838054189(21)805, (01)00884838054202(21)1856, (01)00884838059115(21)365, (01)00884838059115(21)199, (01)00884838054202(21)978, (01)00884838054226(21)62, (01)00884838054202(21)1132, (01)00884838054202(21)1187, (01)00884838054202(21)363, (01)00884838054202(21)1889, (01)00884838054202(21)1076, (01)00884838054202(21)952, (01)00884838054189(21)622, (01)00884838054202(21)499, (01)00884838054196(21)116, (01)00884838054196(21)174, (01)00884838059115(21)278, (01)00884838054202(21)1441, (01)00884838054202(21)210, (01)00884838054202(21)720, (01)00884838054202(21)556, (01)00884838054196(21)32, (01)00884838054202(21)2843, (01)00884838054202(21)560, (01)00884838054189(21)234, (01)00884838054196(21)39, (01)00884838054196(21)49, (01)00884838054196(21)56, (01)00884838054202(21)194, (01)00884838054202(21)195, (01)00884838054202(21)197, (01)00884838054202(21)227, (01)00884838054202(21)229, (01)00884838054196(21)120, (01)00884838054219(21)129, (01)00884838054202(21)627, (01)00884838059115(21)215, (01)00884838054202(21)1457, (01)00884838054202(21)139, (01)00884838054202(21)552, (01)00884838054189(21)1293, (01)00884838054202(21)2284, (01)00884838054189(21)121, (01)00884838054202(21)217, (01)00884838054202(21)228, (01)00884838054226(21)46, (01)00884838054196(21)123, (01)00884838054196(21)133, (01)00884838054196(21)305, (01)00884838054202(21)122, (01)00884838054202(21)757, (01)00884838054202(21)1424, (01)00884838054202(21)2647, (01)00884838054196(21)246, (01)00884838054202(21)893, (01)00884838059115(21)288, (01)00884838054202(21)879, (01)00884838054226(21)118, (01)00884838054202(21)924, (01)00884838054202(21)934, (01)00884838054202(21)1854, (01)00884838054202(21)313, (01)00884838054189(21)544, (01)00884838054202(21)1004, (01)00884838054202(21)528, (01)00884838054189(21)79, (01)00884838054189(21)559, (01)00884838054189(21)1205, (01)00884838054202(21)159, (01)00884838054202(21)1373, (01)00884838054202(21)1846, (01)00884838054202(21)962, (01)00884838054202(21)1984, (01)00884838054202(21)2422, (01)00884838054202(21)494, (01)00884838054202(21)1949, (01)00884838059115(21)254, (01)00884838054202(21)918, (01)00884838054202(21)998, (01)00884838054189(21)617, (01)00884838054226(21)67, (01)00884838054202(21)1463, (01)00884838054202(21)1191, (01)00884838054202(21)1212, (01)00884838054202(21)1370, (01)00884838054202(21)1002, (01)00884838054189(21)1101, (01)00884838054202(21)490, (01)00884838054202(21)2834, (01)00884838054189(21)738, (01)00884838054202(21)382, (01)00884838054202(21)2760, (01)00884838054189(21)957, (01)00884838054202(21)1600, (01)00884838054202(21)1244, (01)00884838054202(21)1030, (01)00884838054202(21)207, (01)00884838054202(21)208, (01)00884838054202(21)1869, (01)00884838059115(21)94, (01)00884838054189(21)400, (01)00884838054189(21)431, (01)00884838054202(21)2196, (01)00884838054202(21)1286, (01)00884838059115(21)304, (01)00884838054202(21)636, (01)00884838054202(21)901, (01)00884838054202(21)1446, (01)00884838054202(21)1828, (01)00884838054196(21)255, (01)00884838054202(21)1349, (01)00884838054202(21)1351, (01)00884838054202(21)561, (01)00884838054189(21)602, (01)00884838054202(21)680, (01)00884838054202(21)1074, (01)00884838054202(21)1677, (01)00884838059115(21)279, (01)00884838054202(21)908, (01)00884838054202(21)1629, (01)00884838054189(21)201, (01)00884838054202(21)420, (01)00884838054202(21)1, (01)00884838054202(21)272, (01)00884838054202(21)449, (01)00884838054202(21)516, (01)00884838054202(21)533, (01)00884838054202(21)991, (01)00884838054202(21)422, (01)00884838054226(21)78, (01)00884838059115(21)171, (01)00884838054196(21)40, (01)00884838059115(21)82, (01)00884838059115(21)85, (01)00884838054189(21)1340, (01)00884838054202(21)1034, (01)00884838054202(21)889, (01)00884838054202(21)1822, (01)00884838054202(21)306, (01)00884838054202(21)900, (01)00884838054196(21)178, (01)00884838054202(21)914, (01)00884838054226(21)120, (01)00884838059115(21)202, (01)00884838054202(21)671, (01)00884838059115(21)99, (01)00884838054202(21)882, (01)00884838054202(21)372, (01)00884838054202(21)695, (01)00884838054202(21)945, (01)00884838054202(21)125, (01)00884838054202(21)368, (01)00884838054189(21)122, (01)00884838054202(21)1136, (01)00884838054202(21)1639, (01)00884838054189(21)653, (01)00884838054189(21)663, (01)00884838054202(21)215, (01)00884838054202(21)279, (01)00884838054196(21)104, (01)00884838054202(21)347, (01)00884838054202(21)909, (01)00884838054189(21)688, (01)00884838054202(21)631, (01)00884838054202(21)637, (01)00884838059115(21)182, (01)00884838054189(21)563, (01)00884838054219(21)71, (01)00884838054202(21)161, (01)00884838054202(21)534, (01)00884838054202(21)242, (01)00884838054202(21)284, (01)00884838059115(21)22, (01)00884838054202(21)375, (01)00884838054226(21)7, (01)00884838059115(21)200, (01)00884838054189(21)704, (01)00884838054202(21)365, (01)00884838054202(21)579, (01)00884838054202(21)1820, (01)00884838054202(21)809, (01)00884838054202(21)1331, (01)00884838054202(21)1566, (01)00884838054189(21)462, (01)00884838054202(21)1598, (01)00884838054202(21)407, (01)00884838054202(21)489, (01)00884838054202(21)1826, (01)00884838054189(21)191, (01)00884838054202(21)1368, (01)00884838054189(21)931, (01)00884838054202(21)155, (01)00884838054189(21)967, (01)00884838054202(21)1649, (01)00884838054226(21)106, (01)00884838054202(21)496, (01)00884838054202(21)1549, (01)00884838054202(21)1568, (01)00884838054202(21)1845, (01)00884838054202(21)1980, (01)00884838054202(21)710, (01)00884838054202(21)793, (01)00884838054202(21)213, (01)00884838054202(21)692, (01)00884838054202(21)929, (01)00884838054202(21)861, (01)00884838059115(21)129, (01)00884838054202(21)1035, (01)00884838059115(21)59, (01)00884838054202(21)1024, (01)00884838054202(21)1256, (01)00884838054189(21)265, (01)00884838054202(21)222, (01)00884838054202(21)394, (01)00884838054202(21)428, (01)00884838054202(21)1022, (01)00884838054202(21)343, (01)00884838054202(21)424, (01)00884838054202(21)435, (01)00884838054202(21)437, (01)00884838054219(21)36, (01)00884838054202(21)1811, (01)00884838059115(21)172, (01)00884838054189(21)826, (01)00884838054196(21)266, (01)00884838054202(21)1168, (01)00884838054202(21)940, (01)00884838054202(21)1657.
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls2 deaths3,984 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD20/15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths776 reports
- Allura Xper FD Allura Xper FD10/10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System19 recalls603 reports
- Allura Xper FD Allura Xper FD20/20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System18 recalls250 reports
- Allura Xper FD Allura Xper FD20/10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System19 recalls172 reports
- Allura Xper FD Allura Xper FD20/20 OR TablePhilips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System7 recalls13 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.