Skip to content

Battery Pack portion of the LifeVest device. The LifeVest… recallZ-0061-2008 · Zoll Lifecor Corporation · initiated Sep 14, 2007

Terminated All recalls posted in October 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zoll Lifecor Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Possible malfunction of the battery pack. A significant electrostatic discharge (ESD) can cause the battery pack to malfunction. The malfunction causes the battery pack to discharge more rapidly and prevents the pack from recharging.

FDA root cause category: Device Design

What the company is doing

Patients currently using the LifeVest should continue to do so while the product removal is underway. Advisory notices will be sent via first class mail 9/14/2007 to all active patients using the LifeVest device. Product replacements began on 8/31/07 by shipping out revised battery packs for new patient use. Shipments of new battery packs will continue based on production capacity and material availability. Any patients expected to be using the device longer than 5 months will be provided with 2 replacement Battery Packs. They will be instructed to return their existing packs in the same shipping box.

Recall details

Recall number
Z-0061-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 14, 2007
Posted by the FDA
Oct 18, 2007
Recalling firm
Zoll Lifecor Corporation
Quantity
2006 packs
Product code
MVK
FDA submissions
P010030
Distribution
The product was shipped nationwide to patients. The product was also shipped to distributors in Germany, United Kingdom, Italy, Spain, France, and Denmark.

Product description

Battery Pack portion of the LifeVest device. The LifeVest is a wearable cardioverter defibrillator worn by a patient at risk for sudden cardiac arrest (SCA).

Code information (models, lots, serial numbers)

Product number 10A0894-A01. Serial numbers 71000101 through 71003176.

Other devices under the same FDA submission

12 devices cleared under P010030 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page