Skip to content

i-STAT HL7 Interface recallZ-0865-2010 · Abbott Point of Care Inc. · initiated Sep 29, 2009

Terminated All recalls posted in February 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abbott Point of Care has determined that it is possible under certain circumstances that some test records could have their status update messages mis-matched with another record, resulting in the incorrect patient name, order number of comment being included in the CDS test record only.

FDA root cause category: Software design

What the company is doing

Urgent Recall Notification letters were sent by Fed Ex to all US customers on September 29, 2009 and worldwide on October 1, 2009. Questions should be directed to Abbott Point of Care Technical Support at 1-800-366-8010, Option 1.

Recall details

Recall number
Z-0865-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 29, 2009
Posted by the FDA
Feb 26, 2010
Recalling firm
Abbott Point of Care Inc.
Quantity
36
Product code
JQP
FDA submissions
K001387
Distribution
Nationwide and Foreign: Germany, the United Kingdom, Switzerland and the Netherlands.

Product description

i-STAT HL7 Interface The HL7 Interface product is an accessory to the analyzer

Code information (models, lots, serial numbers)

i-STAT Catalog number: 111610 Abbott list number (US and Canada) 08K4501

Other devices under the same FDA submission

7 devices cleared under K001387 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.