Why it was recalled
The firm has received reports of users encountering difficulties in usage leading to surgical delays or minor damage to a threaded surface of the shell implant.
FDA root cause category: Device Design
What the company is doing
The firm sent Urgent: Device Removal letters dated 3/15/2010. The letter identified the affected product, described the issue, discussed the clinical implications, and provided a list of required actions. The consignees were instructed to perform an inventory of the recalled product and report the amount on hand to Zimmer by completing the Inventory Return Certification Form. After Zimmer received the report, they would send out new product to the consignee. After new product was received the consignees were instructed to return the recalled product to Zimmer. The firm expects to have all product replaced by 4/30/2010. If there are any questions or additional info needed, consignees should contact Zimmer, Inc. at 1-800-613-6131 or 574-372-4463.
Recall details
- Recall number
- Z-0072-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 19, 2010
- Posted by the FDA
- Oct 15, 2010
- Recalling firm
- Zimmer Inc.
- Quantity
- 1125 all products
- Product code
- LXH
- FDA submissions
- K091508
- Distribution
- Worldwide Distribution -- US, Singapore, Australia, Portugal, Germany, Switzerland, Spain, Finland, United Kingdom, Israel, Italy and Sweden.
Product description
Continuum Acetabular Systems Shell Inserter Adapter without Rotational Control, 00-8790-003-10, Zimmer Inc., Warsaw, IN. Used to mate the inserter with the shell implant.
Code information (models, lots, serial numbers)
61255960, 61286739, 61293344, 61297228, 61301146, 61303916, 61321619, 61341496, 61341497, 61351886, 61361912, 61361919, 61369266, 61371666, 61371667, 61387139, 61396637 and 61413131.
Other devices under the same FDA submission
40 devices cleared under K091508 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Continuum® Trilogy® Allofit® Longevity® 00-8751-010-36Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented8 reports
- Continuum® Trabecular Metal™ 00-8757-056-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented8 reports
- Continuum® Trabecular Metal™ 00-8757-048-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented5 reports
- Continuum® Trabecular Metal™ 00-8757-052-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Continuum® Trabecular Metal™ 00-8757-054-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8751-009-32Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8751-011-36Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8751-013-36Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Trilogy® 00-8753-050-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Trilogy® 00-8753-052-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- Continuum® Trabecular Metal™ 00-8757-050-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Continuum® Trabecular Metal™ 00-8757-060-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 55157).
- Z-0071-2011 · TerminatedContinuum Acetabular Systems Shell Inserter Adapter with Rotational Control, 00-8790-003-00, Zimmer Inc., Warsaw, IN. Used to mate the inserter with the shell implant.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page