Skip to content

Arrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr… recallZ-0790-2011 · Arrow International, Inc., Division of Teleflex Medical Inc. · initiated Oct 11, 2010

Terminated All recalls posted in December 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International, Inc., Division of Teleflex Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

SuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.

FDA root cause category: Device Design

What the company is doing

Arrow international initiated a voluntary recall on 12/21/10 by letter requesting users to discontinue use and return all unused 5800 Series IAB with SuperFlex Introducers to Arrow. This field action supercedes the Safety Alert IAB 'Stuck in Sheath' letter issued on 10/8/2010. Questions should be directed towards their local sales representative or the IABP support Line at 1-866-396-2111 or 1-617-389-6400.

Recall details

Recall number
Z-0790-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 11, 2010
Posted by the FDA
Dec 23, 2010
Recalling firm
Arrow International, Inc., Division of Teleflex Medical Inc.
Quantity
5289 units
Product code
DSP
FDA submissions
K000729, K021462
Distribution
Worldwide Distribution -- USA, Canada, Australia, Belgium, Czechoslovakia, Germany, Russia, Spain, Finland, France, Great Britain, Hungary, Italy, Netherlands, Poland, Saudi Arabia, Switzerland, Brazil, Australia, Chile, Colombia, Ecuador, India, Indonesia, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Peru, Philippines, Taiwan, Thailand, and Venezuela.

Product description

Arrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr, 30cc. Model: IAB-05830-U

Code information (models, lots, serial numbers)

Lot Numbers: KF0087756 MF0053016 MF0053017 MF0053018 MF0053019 MF0073778 MF0074136 MF0074137 MF5114292 MF7076257 MF7118004 MF7128505 MF8103263 MF8124064 MF8124178 MF8124307 MF8124408 MF9014573 MF9025207 MF9025208 MF9025213 MF9025226 MF9025227 MF9035382 MF9035448 MF9046246 MF9056839 MF9056840 MF9057065 MF9057066 MF9067735 MF9067736 MF9077955 MF9077968 MF9077969 MF9078203 MF9078343 MF9088501 MF9099356 MF9099357 MF9099358 MF9099359 MF9110054 MS9056710 MS9057152 MS9057153 MS9109664 MS9110295 MS9110368

Other devices under the same FDA submission

13 devices cleared under K000729, K021462 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 56965).