Skip to content

Percutaneous Insertion Tray For Use with: Arrow Ultra 8 IAB… recallZ-0969-2011 · Arrow International, Inc., Division of Teleflex Medical Inc. · initiated Oct 11, 2010

Terminated All recalls posted in December 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International, Inc., Division of Teleflex Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

SuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.

FDA root cause category: Device Design

What the company is doing

Arrow international initiated a voluntary recall on 12/21/10 by letter requesting users to discontinue use and return all unused 5800 Series IAB with SuperFlex Introducers to Arrow. This field action supercedes the Safety Alert IAB 'Stuck in Sheath' letter issued on 10/8/2010. Questions should be directed towards their local sales representative or the IABP support Line at 1-866-396-2111 or 1-617-389-6400.

Recall details

Recall number
Z-0969-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 11, 2010
Posted by the FDA
Dec 23, 2010
Recalling firm
Arrow International, Inc., Division of Teleflex Medical Inc.
Quantity
714
Product code
DSP
FDA submissions
K000729, K021462
Distribution
Worldwide Distribution -- USA, Canada, Australia, Belgium, Czechoslovakia, Germany, Russia, Spain, Finland, France, Great Britain, Hungary, Italy, Netherlands, Poland, Saudi Arabia, Switzerland, Brazil, Australia, Chile, Colombia, Ecuador, India, Indonesia, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Peru, Philippines, Taiwan, Thailand, and Venezuela.

Product description

Percutaneous Insertion Tray For Use with: Arrow Ultra 8 IAB (Intra Aortic Balloon) Catheters Model: IAK-05845

Code information (models, lots, serial numbers)

Lot Numbers: KF0118523 MF8092597 MF8103402 MF8114055 MF9014548 MF9046091 MF9056823 MF9057130 MF9078053 MF9088776 MF9099410 MF9109907

Other devices under the same FDA submission

13 devices cleared under K000729, K021462 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 56965).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page