Arrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr… recallZ-0791-2011 · Arrow International, Inc., Division of Teleflex Medical Inc. · initiated Oct 11, 2010
Terminated All recalls posted in December 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, Inc., Division of Teleflex Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
SuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.
FDA root cause category: Device Design
What the company is doing
Arrow international initiated a voluntary recall on 12/21/10 by letter requesting users to discontinue use and return all unused 5800 Series IAB with SuperFlex Introducers to Arrow. This field action supercedes the Safety Alert IAB 'Stuck in Sheath' letter issued on 10/8/2010. Questions should be directed towards their local sales representative or the IABP support Line at 1-866-396-2111 or 1-617-389-6400.
Recall details
- Recall number
- Z-0791-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 11, 2010
- Posted by the FDA
- Dec 23, 2010
- Recalling firm
- Arrow International, Inc., Division of Teleflex Medical Inc.
- Quantity
- 13, 391 units
- Product code
- DSP
- FDA submissions
- K000729, K021462
- Distribution
- Worldwide Distribution -- USA, Canada, Australia, Belgium, Czechoslovakia, Germany, Russia, Spain, Finland, France, Great Britain, Hungary, Italy, Netherlands, Poland, Saudi Arabia, Switzerland, Brazil, Australia, Chile, Colombia, Ecuador, India, Indonesia, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Peru, Philippines, Taiwan, Thailand, and Venezuela.
Product description
Arrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr, 40cc. Model: IAB-05840-U
Code information (models, lots, serial numbers)
Lot Numbers: KF0087855 KF0087856 KF0098040 KF0108374 MF0011189 MF0021654 MF0032271 MF0042406 MF0042501 MF0042786 MF0053023 MF0063353 MF0073903 MF0074071 MF7065606 MF7086741 MF7118151 MF7128507 MF8103280 MF8103571 MF8113919 MF8113951 MF8124345 MF9014575 MF9014817 MF9024965 MF9024993 MF9025112 MF9025209 MF9035509 MF9035685 MF9035955 MF9046057 MF9046247 MF9046501 MF9056746 MF9057069 MF9057070 MF9067275 MF9067738 MF9067740 MF9077954 MF9078048 MF9078052 MF9078348 MF9078349 MF9088685 MF9100025 MS0053114 MS9056684 MS9056857 MS9057154 MS9110263 MS9110269
Other devices under the same FDA submission
13 devices cleared under K000729, K021462 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ARROW IPN000260Teleflex Incorporated · System, balloon, intra-aortic and control4 deaths698 reports
- ARROW IPN000257Teleflex Incorporated · System, balloon, intra-aortic and control3 deaths233 reports
- ARROW IPN915322Teleflex Incorporated · System, balloon, intra-aortic and control1 recall29 reports
- ARROW IPN039793Teleflex Incorporated · System, balloon, intra-aortic and control62 reports
- ARROW IPN915321Teleflex Incorporated · System, balloon, intra-aortic and control1 recall7 reports
- ARROW IPN915320Teleflex Incorporated · System, balloon, intra-aortic and control1 recall0 reports
- ARROW IPN000255Teleflex Incorporated · System, balloon, intra-aortic and control32 reports
- ARROW IPN039788Teleflex Incorporated · System, balloon, intra-aortic and control29 reports
- ARROW IPN039782Teleflex Incorporated · System, balloon, intra-aortic and control23 reports
- ARROW IPN039778Teleflex Incorporated · System, balloon, intra-aortic and control16 reports
- ARROW IPN038075Teleflex Incorporated · System, balloon, intra-aortic and control12 reports
- ARROW IPN038074Teleflex Incorporated · System, balloon, intra-aortic and control7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 56965).
- Z-0790-2011 · TerminatedArrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr, 30cc. Model: IAB-05830-U
- Z-0792-2011 · TerminatedFiberOptix IAB Sheath, Ultra 8 Fr, 30cc. Model: IAB-05830-LWS
- Z-0793-2011 · TerminatedFiberOptix IAB Ultra 8 Fr, 40cc Model: IAB-05840-LWS
- Z-0969-2011 · TerminatedPercutaneous Insertion Tray For Use with: Arrow Ultra 8 IAB (Intra Aortic Balloon) Catheters Model: IAK-05845
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.