Skip to content

Arrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr… recallZ-0791-2011 · Arrow International, Inc., Division of Teleflex Medical Inc. · initiated Oct 11, 2010

Terminated All recalls posted in December 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International, Inc., Division of Teleflex Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

SuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.

FDA root cause category: Device Design

What the company is doing

Arrow international initiated a voluntary recall on 12/21/10 by letter requesting users to discontinue use and return all unused 5800 Series IAB with SuperFlex Introducers to Arrow. This field action supercedes the Safety Alert IAB 'Stuck in Sheath' letter issued on 10/8/2010. Questions should be directed towards their local sales representative or the IABP support Line at 1-866-396-2111 or 1-617-389-6400.

Recall details

Recall number
Z-0791-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 11, 2010
Posted by the FDA
Dec 23, 2010
Recalling firm
Arrow International, Inc., Division of Teleflex Medical Inc.
Quantity
13, 391 units
Product code
DSP
FDA submissions
K000729, K021462
Distribution
Worldwide Distribution -- USA, Canada, Australia, Belgium, Czechoslovakia, Germany, Russia, Spain, Finland, France, Great Britain, Hungary, Italy, Netherlands, Poland, Saudi Arabia, Switzerland, Brazil, Australia, Chile, Colombia, Ecuador, India, Indonesia, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Peru, Philippines, Taiwan, Thailand, and Venezuela.

Product description

Arrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr, 40cc. Model: IAB-05840-U

Code information (models, lots, serial numbers)

Lot Numbers: KF0087855 KF0087856 KF0098040 KF0108374 MF0011189 MF0021654 MF0032271 MF0042406 MF0042501 MF0042786 MF0053023 MF0063353 MF0073903 MF0074071 MF7065606 MF7086741 MF7118151 MF7128507 MF8103280 MF8103571 MF8113919 MF8113951 MF8124345 MF9014575 MF9014817 MF9024965 MF9024993 MF9025112 MF9025209 MF9035509 MF9035685 MF9035955 MF9046057 MF9046247 MF9046501 MF9056746 MF9057069 MF9057070 MF9067275 MF9067738 MF9067740 MF9077954 MF9078048 MF9078052 MF9078348 MF9078349 MF9088685 MF9100025 MS0053114 MS9056684 MS9056857 MS9057154 MS9110263 MS9110269

Other devices under the same FDA submission

13 devices cleared under K000729, K021462 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 56965).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page