Software version 3.3 for Rapid Programmer devices, Model… recallZ-3091-2011 · Advanced Neuromodulation Systems · initiated Jun 2, 2009
Terminated All recalls posted in August 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Advanced Neuromodulation Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product programmed with unapproved software was distributed.
FDA root cause category: Software change control
What the company is doing
St. Jude Medical Neuromodulation contacted all affected customers by telephone between June 2, 2009 and June 5, 2009. Customers were given information about the product, problem and action to be taken. All affected devices were returned to the firm as instructed.
Recall details
- Recall number
- Z-3091-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jun 2, 2009
- Posted by the FDA
- Aug 30, 2011
- Recalling firm
- Advanced Neuromodulation Systems
- Quantity
- 14 units
- Product code
- LGW
- FDA submissions
- P010032
- Distribution
- Distributed in New Hampshire and Puerto Rico.
Product description
Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting implantable pulse generators of the Eon, Genesis, and Renew family.
Code information (models, lots, serial numbers)
Serial numbers: 101565, 101544, 101555, 101548, 101561, 101551, 101563, 101546, 101556, 101553, 101554, 101558, 101545, and 101566
Other devices under the same FDA submission
110 devices cleared under P010032 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
- Penta™ 3228St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11 deaths5,201 reports
- Eon Mini™ 3788St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death4,666 reports
- Prodigy MRI™ 3772St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths3,162 reports
- Octrode™ 3189St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF8 deaths2,238 reports
- Protégé MRI™ 3771St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,997 reports
- Eterna™ 32400St. Jude Medical, Inc. · Stimulator, spinal-cord, totally implanted for pain relief1 recall5 deaths1,547 reports
- Octrode® 3186Advanced Neuromodulation Systems, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths1,455 reports
- Swift-Lock™ 1192St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,400 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page