Bio-modular reverse shoulder humeral tray with lock ring… recallZ-2541-2011 · Biomet, Inc. · initiated Apr 22, 2011
Terminated All recalls posted in June 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.
FDA root cause category: Employee error
What the company is doing
The firm, Biomet Orthopedics, sent an "URGENT MEDICAL DEVICE RECALL" letter dated April 22, 2011 to their customers. The letter describes the product, problem, and actions to be taken. The customers were instructed to immediately locate the recalled product; discontinue use; return the product back to Biomet; confirm receipt of the notice by calling 800-348-9500 extension 1368 or 1325, and complete and return the attached form via fax to: 574-372-1683 or mail to: Biomet, 56 East Bell Drive, Warsaw, IN 46580. Customers with questions regarding this recall should call 800-348-9500 extension 1368 or 1325.
Recall details
- Recall number
- Z-2541-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 22, 2011
- Posted by the FDA
- Jun 9, 2011
- Recalling firm
- Biomet, Inc.
- Quantity
- 1,575 units all products
- Product code
- KWS
- FDA submissions
- K080642
- Distribution
- Worldwide distribution: USA (nationwide) including: Puerto Rico and countries including: Australia, Canada, Chile, Columbia, Costa Rica, Korea and Netherlands.
Product description
Bio-modular reverse shoulder humeral tray with lock ring, 44 mm 10 degree angle, sterile, single use, REF 11-113685, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
Code information (models, lots, serial numbers)
M026410, M583200, M583210, M907780, M907790 and M907800.
Other devices under the same FDA submission
46 devices cleared under K080642 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Comprehensive Reverse Shoulder 115310Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented1 death99 reports
- COMPREHENSIVE INSTRUMENTATION 110028879Biomet Orthopedics, LLC · shoulder prosthesis, reverse configuration71 reports
- Comprehensive® Reverse Shoulder XL-115364Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented1 recall7 reports
- Comprehensive® Reverse Shoulder XL-115366Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented1 recall4 reports
- Comprehensive® Reverse Shoulder 180552Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 death51 reports
- Comprehensive® Reverse Shoulder 180553Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 death47 reports
- Comprehensive® Reverse Shoulder 180550Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented2 deaths43 reports
- Comprehensive Reverse Shoulder 115370Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED42 reports
- Comprehensive® Reverse Shoulder 180551Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented42 reports
- Comprehensive® Shoulder System 118001Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED41 reports
- Comprehensive® Reverse Shoulder 180554Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED28 reports
- Comprehensive® Instrumentation 110028879Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED28 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 58571).
- Z-2542-2011 · TerminatedBio-modular reverse shoulder humeral tray with lock ring, 44 mm standard, sterile, single use, REF 11-113680, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder…
- Z-2543-2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm +10 mm, sterile, single use, REF 115348, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Should…
- Z-2544-2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm +5 mm, sterile, single use, REF 115345, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulde…
- Z-2545-2011 · TerminatedComprehensive reverse shoulder humeral tray with locking ring, 44 mm standard, sterile, single use, REF 115340, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Sho…
- Z-2546-2011 · TerminatedCustom bi-angular to reverse shoulder humeral tray 44 mm sterile, REF CP561548, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated fo…
- Z-2547-2011 · TerminatedCustom solar comprehensive reverse humeral tray with lock ring 44 mm +5mm, sterile, REF CP561659, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product…
- Z-2548-2011 · TerminatedCustom biomodular humeral tray 44 mm sterile, REF CP561686, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated for use in patients who…
- Z-2549-2011 · TerminatedCustom comprehensive reverse humeral tray set w/ lock rings 44 mm + 15 mm & + 18 mm sterile, REF CP561705, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder…
- Z-2550-2011 · TerminatedCustom comprehensive reverse shoulder bio-modular base 44 mm humeral tray standard, REF CP561510, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page