Medtronic Mosaic Porcine Bioprostheses with Cinch, Heart… recallZ-1225-2009 · Medtronic Inc · initiated Jan 26, 2009
Terminated All recalls posted in May 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic identified several heart valves (Mosaic Porcine Bioprosthesis with Cinch & Hancock II Bioprosthesis with Cinch) that were inadvertently distributed after exposure to temperatures exceeding the 77 F (25 C) maximum storage temperature published on the labeling. The product is being retrieved as they do not meet the handling requirements.
FDA root cause category: Storage
What the company is doing
A Medtronic "Urgent Medical Device Recall Notice" dated January 26, 2009, was sent to Risk Managers of each affected account and to each Implanting Physician. The letter described the problem, product and informed consignees that they will be retrieving the non-implanted vales. If additional information is needed, customers are to contact their local Medtronic sales representative or technical services at 1-877-526-7890.
Recall details
- Recall number
- Z-1225-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 26, 2009
- Posted by the FDA
- May 29, 2009
- Recalling firm
- Medtronic Inc
- Quantity
- 5
- Product code
- DYE
- FDA submissions
- P990064
- Distribution
- Nationwide Distribution -- Including states of CT, FL, MD, MA, MI, NC, OH, OR, PA, SD, TN, TX, WI, and WA.
Product description
Medtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 Medtronic Parkway, Minneapolis, MN 55432-5604, Manufactured at: Santa Ana, Ca. Indicated for the replacement of malfunctioning native or aortic and mitral heart valves.
Code information (models, lots, serial numbers)
Lot Serial Number 21A06D2415, 27A05F7021, 23A06W1410, 25A04V1591, and 29A07F0075.
Other devices under the same FDA submission
101 devices cleared under P990064 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- MOSAIC ULTRA® 305U23Medtronic, Inc. · heart-valve, non-allograft tissue239 reports
- MOSAIC ULTRA® 305U25Medtronic, Inc. · heart-valve, non-allograft tissue1 death234 reports
- MOSAIC® 310C29Medtronic, Inc. · heart-valve, non-allograft tissue1 recall13 deaths133 reports
- MOSAIC® 310C31Medtronic, Inc. · heart-valve, non-allograft tissue1 recall11 deaths126 reports
- MOSAIC® 305C27Medtronic, Inc. · heart-valve, non-allograft tissue170 reports
- MOSAIC ULTRA® 305U223Medtronic, Inc. · heart-valve, non-allograft tissue3 deaths163 reports
- MOSAIC® 305C21Medtronic, Inc. · heart-valve, non-allograft tissue161 reports
- MOSAIC ULTRA® 305U225Medtronic, Inc. · heart-valve, non-allograft tissue3 deaths151 reports
- MOSAIC ULTRA® 305U221Medtronic, Inc. · heart-valve, non-allograft tissue135 reports
- MOSAIC ULTRA® 305U27Medtronic, Inc. · heart-valve, non-allograft tissue1 death124 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 51220).
- Z-1226-2009 · TerminatedHancock II Bioprostheses with Cinch, Heart Valve, Model # T505 & T510; Catalog #s T505C2501, T505C2101, T505U25ID , T505C2301, T505C2301, T505C2501, T505C2101, T510C29ID, T510C25ID…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.