Skip to content

HeartWare Ventricular Assist System: HeartWare Battery… recallZ-1887-2016 · Heartware, Inc. · initiated Jan 7, 2016

Class I Terminated All recalls posted in July 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Heartware, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Heartware Ventricular Assist System Battery Cell experiencing premature depletion.

FDA root cause category: Software design

What the company is doing

The firm, HeartWare, sent an "URGENT MEDICAL DEVICE RECALL" letter dated January 7, 2016, to all their consignees informing them that of the HeartWare HVAD Battery recall (device exchange). The letter described the product, problem and actions to be taken. The customers were instructed to: 1. Promptly complete the attached Acknowledgment Form and return it to HeartWare no later than 30 days from the date of this letter via email to [email protected] or provide verbal Acknowledgement by calling Customer Service line , 1(844)491-7871; or fax to +1 (844)721-3044.. 2. Immediately identify and quarantine affected batteries in hospital inventory, and apply a RECALLED PRODUCT DO NOT USE sticker (provided with this notice) as illustrated below. Replace quarantined inventory AC Adapters and batteries by requesting a Return Goods Authorization (RGA) through a qualified representative or directly through HeartWare Customer Service (via telephone 1-877-367-4823 or email [email protected]). A qualified representative will update the Controller software for affected Controller units in inventory. 3. Identify, collect and quarantine affected product in patient possession. Arrange for current patients to bring their Batteries to a clinic appointment as soon as possible, (at least within the next three months), with urgency dependent upon individual patient circumstances. During their regularly scheduled appointment, a qualified representative will assist with quarantining and replacing affected product. Apply a RECALLED PRODUCT DO NOT USE sticker as illustrated on affected batteries. Replace affected patient product using hospital inventory, and replace consumed hospital inventory by requesting an RGA as described above. 4. Return affected product to HeartWare. Your HeartWare representative will assist you with returning affected product to HeartWare. 5. Complete and return the attached Completion Form no later than 6 months from the date of this let

Recall details

Recall number
Z-1887-2016
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Jan 7, 2016
Posted by the FDA
Jul 8, 2016
Recalling firm
Heartware, Inc.
Quantity
18, 631
Product code
DSQ
FDA submissions
P100047
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and Washington D.C.

Product description

HeartWare Ventricular Assist System: HeartWare Battery, Lithium Ion, 14.8 V, Rx only. Catalog 1650 (US) and 1650, 1650-DE and A00035 (OUS) Product Usage: Used as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare Ventricular Assist System is designed for in-hospital and out-of-hospital settings, including transportation via fixed-wing aircraft or helicopter.

Code information (models, lots, serial numbers)

Serial numbers ranging BAT000001 to BAT199999.

Other devices under the same FDA submission

33 devices cleared under P100047 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page