The GMK knee system: designed for cemented use in total… recallZ-0266-2018 · Medacta Usa Inc · initiated Oct 24, 2017
Class II Terminated All recalls posted in December 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medacta Usa Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques.
FDA root cause category: Device Design
What the company is doing
All surgeon customers shall be notified via email by October 24, 2017 of the reminder to follow the surgical technique. To ensure that surgeons receive the notification, we will send the letter via email so that we can track and ensure the message was received. If the email is returned, we would then send a personalized letter directly to the surgeon. A copy of this notification is included with the submission of this report.
Recall details
- Recall number
- Z-0266-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 24, 2017
- Posted by the FDA
- Dec 20, 2017
- Recalling firm
- Medacta Usa Inc
- Quantity
- 12,768 units
- Product code
- JWH
- FDA submissions
- K090988
- Distribution
- nationwide
Product description
The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF
Code information (models, lots, serial numbers)
all serial numbers
Other devices under the same FDA submission
81 devices cleared under K090988 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- GMK Primary Total Knee System 02.07.1204RMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer65 reports
- GMK Primary Total Knee System 02.07.1202RMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer61 reports
- GMK Primary Total Knee System 02.07.1205RMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer61 reports
- GMK Primary Total Knee System 02.07.1204LMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer59 reports
- GMK Primary Total Knee System 02.07.1205LMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer54 reports
- GMK Primary Total Knee System 02.07.1203LMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer53 reports
- GMK Primary Total Knee System 02.07.1203RMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer53 reports
- GMK Primary Total Knee System 02.07.1202LMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer49 reports
- GMK Primary Total Knee System 02.07.0034RPMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer42 reports
- GMK Primary Total Knee System 02.07.0035RPMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer36 reports
- GMK Primary Total Knee System 02.07.0310PSFMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer18 reports
- GMK Primary Total Knee System 02.07.1206RMedacta International SA · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer15 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page