AVANOS MCKA2-17-100-4Recalls & reported problems

Avanos Medical, Inc. · MCKA2-17-100-4 AVNS,MULTI-RF,RFQKIT,,17X100,1 · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00193493478859), received between Mar 11, 2024 and Mar 11, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Sparking” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Sparking2

Patient problems

  • Skin Discoloration2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2024Sparking; Sparking

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED ONE INCIDENCE, WHICH WAS ASSOCIATED WITH TWO SEPARATE UNITS, INVOLVING ONE PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 2026095-2024-00013 FOR THE SECOND REPORT IT WAS REPORTED, THE PHYSICIAN STARTED THE RADIOFREQUENCY PROCEDURE, AND THE PROBE BEGAN TO SPARK. THE PHYSICIAN STOPPED THE PROCEDURE, AND THE TWO PROBES IN USE WERE REMOVED. THE PHYSICIAN OPENED A NEW PROBE KIT AND WAS ABLE TO COMPLETE THE PROCEDURE. NO SERIOUS INJURY OCCURRED. THE PATIENT WAS REPORTED TO HAVE A SMALL MARK LOCATED WHERE THE PROBE WAS INSERTED AND HAD FORMED A SPARK. DURING THE PROCEDURE, THE PROBE WAS HELD WITH A CLAMP WHICH WAS BELIEVED TO CAUSE A BREAK IN THE INSULATION AND SPARK. PER ADDITIONAL INFORMATION RECEIVED ON 16FEB2024, THE GROUNDING PAD WAS LOCATED ON THE PATIENT¿S POSTERIOR UPPER THIGH.

    Patient problems: Skin Discoloration

    MAUDE report 18873526 ↗
  • MalfunctionMar 11, 2024Sparking; Sparking

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED ONE INCIDENCE, WHICH WAS ASSOCIATED WITH TWO SEPARATE UNITS, INVOLVING ONE PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 2026095-2024-00012 FOR THE FIRST REPORT. IT WAS REPORTED, THE PHYSICIAN STARTED THE RADIOFREQUENCY PROCEDURE, AND THE PROBE BEGAN TO SPARK. THE PHYSICIAN STOPPED THE PROCEDURE, AND THE TWO PROBES IN USE WERE REMOVED. THE PHYSICIAN OPENED A NEW PROBE KIT AND WAS ABLE TO COMPLETE THE PROCEDURE. NO SERIOUS INJURY OCCURRED. THE PATIENT WAS REPORTED TO HAVE A SMALL MARK LOCATED WHERE THE PROBE WAS INSERTED AND HAD FORMED A SPARK. DURING THE PROCEDURE, THE PROBE WAS HELD WITH A CLAMP WHICH WAS BELIEVED TO CAUSE A BREAK IN THE INSULATION AND SPARK. PER ADDITIONAL INFORMATION RECEIVED ON 16FEB2024, THE GROUNDING PAD WAS LOCATED ON THE PATIENT¿S POSTERIOR UPPER THIGH.

    Patient problems: Skin Discoloration

    MAUDE report 18873523 ↗

Other Probe, radiofrequency lesion devices