NA CSK-TC10Recalls & reported problems

Boston Scientific Neuromodulation Corporation · CSK TC Electrode, 10cm · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Boston Scientific CSK TC10 radiofrequency lesioning probe is a reusable electrode used during pain-relief ablation procedures. Most reports describe the electrode tip breaking off or separating from its hub or connector, sometimes inside the patient, along with heating that does not reach the intended level and leaking. There are 58 reports, including 1 death and 4 injuries; patients mention headache, dizziness, itching and blisters after the procedure. One open recall, started in April 2026, says the reusable electrodes may not meet expected performance levels.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
58
Deaths reported
1
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 00813250010107), received between Jun 6, 2017 and Sep 18, 2024 — classified as 1 death, 4 injuries, 53 malfunctions.

The most frequently reported specific device problem is “Material Separation” (18 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 29, 2026Open, Classified

    Reusable electrodes may not meet expected performance levels.

    Z-2402-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Separation18
  • Break7
  • Insufficient Heating7
  • Adverse Event Without Identified Device or Use Problem6
  • Defective Device4
  • Leak/Splash4
  • Fluid/Blood Leak3
  • Structural Problem3

Patient problems

  • No Known Impact Or Consequence To Patient36
  • No Patient Involvement7
  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient5
  • Discomfort3
  • Dizziness3
  • Headache3
  • Burn(s)1

Recent adverse event reports

The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.

  • InjurySep 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED BLISTERS ON THE LEFT SCAPULA AREA ACCOMPANIED BY ITCHING FOLLOWING THE PROCEDURE. THE PATIENT WAS REFERRED TO HOME HEALTH AND WAS DOING WELL.

    Patient problems: Burn(s); Hypersensitivity/Allergic reaction

    MAUDE report 20244277 ↗
  • MalfunctionAug 12, 2024Structural Problem; Structural Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A RADIOFREQUENCY ABLATION (RFA) PROCEDURE DURING WHICH AN ELECTRODE BECAME DETACHED FROM THE CONNECTOR WHILE INSIDE THE PATIENT. THE OUTCOME OF THE PROCEDURE WAS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19967036 ↗
  • MalfunctionAug 12, 2024Structural Problem; Structural Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A RADIOFREQUENCY ABLATION (RFA) PROCEDURE DURING WHICH AN ELECTRODE BECAME DETACHED FROM THE CONNECTOR WHILE INSIDE THE PATIENT. THE OUTCOME OF THE PROCEDURE WAS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19967031 ↗
  • MalfunctionAug 12, 2024Structural Problem; Adverse Event Without Identified Device or Use Problem; Structural Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A RADIOFREQUENCY ABLATION (RFA) PROCEDURE DURING WHICH AN ELECTRODE BECAME DETACHED FROM THE CONNECTOR WHILE INSIDE THE PATIENT. THE OUTCOME OF THE PROCEDURE WAS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19966545 ↗
  • InjuryJul 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A COMPETITORS DEVICE IMPLANTED PRIOR TO THE RADIOFREQUENCY ABLATION (RFA) PROCEDURE. AFTER THE RFA PROCEDURE WAS COMPLETED SUCCESSFULLY, THE PATIENT OBSERVED THE IMPLANTED DEVICE SETTINGS HAD CHANGED UNEXPECTEDLY WITHOUT INTERVENTION. ADDITIONALLY, A WEEK AFTER THE RFA PROCEDURE, THE PATIENT EXPERIENCED SYMPTOMS OF HEADACHE, DIZZINESS, AND DISCOMFORT. THE PAIN PHYSICIAN ASSUMED THESE CHANGES WERE RELATED TO THE RFA PROCEDURE HOWEVER, THE NEUROSURGEON ASSESSED THAT THE IMPLANTED DEVICE APPEARED TO BE FUNCTIONING NORMALLY. DESPITE THIS, THE PATIENT WAS HOSPITALIZED AND UNDERWENT A REVISION PROCEDURE TO REPLACE THE COMPETITORS DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS POSTOPERATIVELY.

    Patient problems: Discomfort; Dizziness; Headache

    MAUDE report 19822326 ↗
  • InjuryJul 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A COMPETITORS DEVICE IMPLANTED PRIOR TO THE RADIOFREQUENCY ABLATION (RFA) PROCEDURE. AFTER THE RFA PROCEDURE WAS COMPLETED SUCCESSFULLY, THE PATIENT OBSERVED THE IMPLANTED DEVICE SETTINGS HAD CHANGED UNEXPECTEDLY WITHOUT INTERVENTION. ADDITIONALLY, A WEEK AFTER THE RF PROCEDURE, THE PATIENT EXPERIENCED SYMPTOMS OF HEADACHE, DIZZINESS, AND DISCOMFORT. THE PAIN PHYSICIAN ASSUMED THESE CHANGES WERE RELATED TO THE RF PROCEDURE HOWEVER, THE NEUROSURGEON ASSESSED THAT THE IMPLANTED DEVICE APPEARED TO BE FUNCTIONING NORMALLY. DESPITE THIS, THE PATIENT WAS HOSPITALIZED AND UNDERWENT A REVISION PROCEDURE TO REPLACE THE COMPETITORS DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS POSTOPERATIVELY.

    Patient problems: Discomfort; Dizziness; Headache

    MAUDE report 19822323 ↗
  • InjuryJul 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A COMPETITORS DEVICE IMPLANTED PRIOR TO THE RADIOFREQUENCY ABLATION (RFA) PROCEDURE. AFTER THE RFA PROCEDURE WAS COMPLETED SUCCESSFULLY, THE PATIENT OBSERVED THE IMPLANTED DEVICE SETTINGS HAD CHANGED UNEXPECTEDLY WITHOUT INTERVENTION. ADDITIONALLY, A WEEK AFTER THE RFA PROCEDURE, THE PATIENT EXPERIENCED SYMPTOMS OF HEADACHE, DIZZINESS, AND DISCOMFORT. THE PAIN PHYSICIAN ASSUMED THESE CHANGES WERE RELATED TO THE RFA PROCEDURE HOWEVER, THE NEUROSURGEON ASSESSED THAT THE IMPLANTED DEVICE APPEARED TO BE FUNCTIONING NORMALLY. DESPITE THIS, THE PATIENT WAS HOSPITALIZED AND UNDERWENT A REVISION PROCEDURE TO REPLACE THE COMPETITORS DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS POSTOPERATIVELY.

    Patient problems: Discomfort; Dizziness; Headache

    MAUDE report 19822322 ↗
  • MalfunctionFeb 20, 2023Defective Device; Defective Device

    IT WAS REPORTED THAT THE ELECTRODE WAS BROKEN AND WAS SEPARATED FROM THE HUB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16404896 ↗
  • MalfunctionDec 30, 2022Insufficient Heating; Temperature Problem; Insufficient Heating; Temperature Problem

    IT WAS REPORTED THAT THE ELECTRODE WAS BROKEN AT THE HUB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16077228 ↗
  • MalfunctionAug 1, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE FACILITY THAT DURING THE STERILIZATION PROCESS THE CENTRAL STERILIZATION TECHNICIAN NOTICED LIQUID SEEPING OUT OF THE BASE OF THE ELECTRODE WHERE THE PROBE WAS ATTACHED. THIS WAS NOTICED AFTER THE ELECTRODE WAS USED IN ONE PROCEDURE ONLY. THEY SUSPECTED THE LIQUID WAS BLOOD AS THE LIQUID WAS RED IN COLOR AND THEY DID NOT REUSE THE ELECTRODE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8852982 ↗
  • MalfunctionMay 1, 2019Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED BY THE FACILITY THAT DURING THE STERILIZATION PROCESS THE CENTRAL STERILIZATION TECHNICIAN NOTICED LIQUID SEEPING OUT OF THE BASE OF THE ELECTRODE WHERE THE PROBE WAS ATTACHED. THIS WAS NOTICED AFTER THE ELECTRODE WAS USED IN ONE PROCEDURE ONLY. THEY SUSPECTED THE LIQUID WAS BLOOD AS THE LIQUID WAS RED IN COLOR AND THEY DID NOT REUSE THE ELECTRODE.

    Patient problems: No Patient Involvement

    MAUDE report 8571698 ↗
  • MalfunctionMay 1, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE FACILITY THAT DURING THE STERILIZATION PROCESS THE CENTRAL STERILIZATION TECHNICIAN NOTICED LIQUID SEEPING OUT OF THE BASE OF THE ELECTRODE WHERE THE PROBE WAS ATTACHED. THIS WAS NOTICED AFTER THE ELECTRODE WAS USED IN ONE PROCEDURE ONLY. THEY SUSPECTED THE LIQUID WAS BLOOD AS THE LIQUID WAS RED IN COLOR AND THEY DID NOT REUSE THE ELECTRODE.

    Patient problems: No Patient Involvement

    MAUDE report 8571596 ↗
  • MalfunctionMay 1, 2019Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED BY THE FACILITY THAT DURING THE STERILIZATION PROCESS THE CENTRAL STERILIZATION TECHNICIAN NOTICED LIQUID SEEPING OUT OF THE BASE OF THE ELECTRODE WHERE THE PROBE WAS ATTACHED. THIS WAS NOTICED AFTER THE ELECTRODE WAS USED IN ONE PROCEDURE ONLY. THEY SUSPECTED THE LIQUID WAS BLOOD AS THE LIQUID WAS RED IN COLOR AND THEY DID NOT REUSE THE ELECTRODE.

    Patient problems: No Patient Involvement

    MAUDE report 8571591 ↗
  • MalfunctionMay 1, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE FACILITY THAT DURING THE STERILIZATION PROCESS THE CENTRAL STERILIZATION TECHNICIAN NOTICED LIQUID SEEPING OUT OF THE BASE OF THE ELECTRODE WHERE THE PROBE WAS ATTACHED. THIS WAS NOTICED AFTER THE ELECTRODE WAS USED IN ONE PROCEDURE ONLY. THEY SUSPECTED THE LIQUID WAS BLOOD AS THE LIQUID WAS RED IN COLOR AND THEY DID NOT REUSE THE ELECTRODE.

    Patient problems: No Patient Involvement

    MAUDE report 8571578 ↗
  • MalfunctionNov 15, 2018Insufficient Heating; Insufficient Heating

    DEVICE ANALYSIS PERFORMED ON THE RETURNED ELECTRODES FOUND THAT THE SHAFT OF THE ELECTRODE WAS SEPARATED FROM THE HUB. THE DAMAGE WAS LIKELY DUE TO UNINTENDED EXCESSIVE EXTERNAL MECHANICAL FORCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8075128 ↗
  • MalfunctionNov 15, 2018Insufficient Heating; Insufficient Heating

    DEVICE ANALYSIS PERFORMED ON THE RETURNED ELECTRODES FOUND THAT THE SHAFT OF THE ELECTRODE WAS SEPARATED FROM THE HUB. THE DAMAGE WAS LIKELY DUE TO UNINTENDED EXCESSIVE EXTERNAL MECHANICAL FORCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8074563 ↗
  • MalfunctionOct 31, 2018Defective Device; Defective Device

    A REPORT WAS RECEIVED THAT THE ELECTRODE WAS BROKEN. THE SHAFT BROKE OFF THE HUB.

    Patient problems: No Patient Involvement

    MAUDE report 8026727 ↗
  • MalfunctionOct 2, 2018Insufficient Heating; Insufficient Heating

    A REPORT WAS RECEIVED THAT THE RETURNED ELECTRODE REVEALED THAT THE SHAFT HAD BECOME SEPARATED FROM THE HUB DUE TO EXCESSIVE EXTERNAL MECHANICAL FORCE.

    Patient problems: No Patient Involvement

    MAUDE report 7926953 ↗
  • MalfunctionSep 19, 2018Insufficient Heating; Insufficient Heating

    DEVICE ANALYSIS WAS PERFORMED ON THE RETURNED DAMAGED ELECTRODE THAT WAS NOT HEATING. THE COMPLAINT WAS CONFIRMED. THE SHAFT OF THE ELECTRODE WAS FOUND TO BE COMPLETELY SEPARATED FROM THE HUB. THE PROBABLE CAUSE IS UNINTENDED USE ERROR CAUSED OR CONTRIBUTED TO EVENT. THE SEPARATED SHAFT WAS LIKELY DUE TO UNINTENDED EXCESSIVE EXTERNAL MECHANICAL FORCE. THE SHAFT WAS NOT RETURNED FOR ANALYSIS. THE FUNCTIONAL TEST COULD NOT BE PERFORMED DUE TO THE SEPARATED SHAFT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7892118 ↗
  • MalfunctionSep 19, 2018Insufficient Heating; Insufficient Heating

    A REPORT WAS RECEIVED THAT THE RETURNED ELECTRODE REVEALED THE RETURNED ELECTRODE WAS FOUND TO BE COMPLETELY SEPARATED FROM THE HUB.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7889938 ↗
  • MalfunctionSep 19, 2018Insufficient Heating; Insufficient Heating

    A REPORT WAS RECEIVED THAT THE RETURNED ELECTRODE REVEALED THE RETURNED ELECTRODE WAS FOUND TO BE COMPLETELY SEPARATED FROM THE HUB.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7889509 ↗
  • MalfunctionAug 7, 2018Material Separation; Material Separation

    DEVICE ANALYSIS PERFORMED ON THE ELECTRODES REVEALED THAT THE SHAFTS OF ALL THIRTEEN ELECTRODES WERE FOUND TO BE COMPLETELY SEPARATED FROM THE HUB DUE TO EXCESSIVE EXTERNAL MECHANICAL FORCE. THE SHAFTS WERE NOT RETURNED FOR ANALYSIS. THE FUNCTIONAL TESTS COULD NOT BE PERFORMED DUE TO THE SEPARATED SHAFTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7758016 ↗
  • MalfunctionAug 1, 2018Material Separation; Material Separation

    DEVICE ANALYSIS WAS PERFORMED ON THE RETURNED ELECTRODE WHICH REVEALED THE SHAFT OF THE ELECTRODE TO BE COMPLETELY SEPARATED FROM THE HUB DUE TO EXCESSIVE EXTERNAL MECHANICAL FORCE. THE SHAFT WAS NOT RETURNED FOR ANALYSIS. THE FUNCTIONAL TEST COULD NOT BE PERFORMED DUE TO THE SEPARATED SHAFT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7740252 ↗
  • MalfunctionAug 1, 2018Fracture; Fracture

    THE VISUAL INSPECTION ON THE RETURNED ELECTRODE REVEALED THAT THE ELECTRODE SHAFT WAS BROKEN RIGHT AT THE HUB. THE FUNCTIONAL TEST COULD NOT BE PERFORMED DUE TO THE BROKEN ELECTRODE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7739898 ↗
  • MalfunctionAug 1, 2018Material Separation; Material Separation

    THE DEVICE ANALYSIS PERFORMED ON THE TWO RETURNED ELECTRODE REVEALED THAT THE SHAFTS OF THE ELECTRODES WERE FOUND TO BE COMPLETELY SEPARATED FROM THE HUB DUE TO EXCESSIVE EXTERNAL MECHANICAL FORCE. THE FUNCTIONAL TESTS COULD NOT BE PERFORMED DUE TO THE SEPARATED SHAFT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7739793 ↗

Other PROBE, RADIOFREQUENCY LESION devices

All PROBE, RADIOFREQUENCY LESION devices →

Sources and updates