Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL5060Recalls & reported problems

Shockwave Medical, Inc. · The Shockwave M5 Intravascular Lithotripsy (IVL) Catheter consists of an array of miniaturized lithotripsy emitters, located inside a balloon, that generate acoustic pressure waves within the target treatment site, disrupting calcium within the lesion and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. Two radiopaque marker bands inside the balloon aid in positioning. The proximal hub has three ports: one for inflation/deflation of the balloon with 50/50 saline/contrast medium, one for the ‘Over-the-wire’ (OTW) guidewire lumen, and one for connection to the IVL Connector Cable (and IVL Generator). · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00195451000157), received between Dec 6, 2021 and Dec 26, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Material Separation1
  • Material Split, Cut or Torn1

Patient problems

  • Foreign Body Embolism1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    A SHOCKWAVE M5+ PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO SUCCESSFULLY TREAT A LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). WHILE ATTEMPTING TO REMOVE THE CATHETER FROM THE SHEATH, A DETACHMENT OCCURRED, AND A FRAGMENT OF THE BALLOON FELL INSIDE THE PATIENT. THE SEPARATION OCCURRED WHILE THE BALLOON WAS FULLY DEFLATED. THE PHYSICIAN USED A SPECIALIZED DEVICE TO RETRIEVE AND TAKE OUT THE FRAGMENT. A SECOND IVL CATHETER WAS USED TO COMPLETE THE SURGERY. THERE WAS NO REPORTED PATIENT HARM. THE PHYSICIAN ATTRIBUTED THE SEPARATION TO QUALITY ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21023666 ↗
  • MalfunctionDec 6, 2021Material Separation; Detachment of Device or Device Component; Material Split, Cut or Torn; Material Separation; Detachment of Device or Device Component; Material Split, Cut or Torn

    A SHOCKWAVE M5 PERIPHERAL LITHOTRIPSY (IVL) DEVICE WAS USED TO TREAT A LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). THE SHOCKWAVE M5 PERIPHERAL LITHOTRIPSY (IVL) DEVICE SHOWED A BALLOON LOSS OF PRESSURE AT THE SECOND CYCLE OF PULSES. THE PHYSICIAN THEN STARTED TO PULL THE CATHETER OUT AND, UPON REMOVAL, THE PHYSICIAN ENCOUNTERED RESISTANCE AND THE IVL CATHETER GOT TRAPPED INTO THE INTRODUCER SHEATH CAUSING PART OF THE IVL BALLOON TO BREAK OFF. A PORTION OF THE IVL BALLOON REMAINED IN THE ARTERY OF THE PATIENT, THE PHYSICIAN ELECTED NOT TO TREAT/REMOVE THE DETACHED COMPONENT DUE TO THE CRITICAL CONDITION OF THE PATIENT. AS OF THE DATE OF THIS REPORT, NO PATIENT SEQUELAE HAVE BEEN REPORTED.

    Patient problems: Foreign Body Embolism

    MAUDE report 12937176 ↗

Other Percutaneous Catheter, Ultrasound devices

All Percutaneous Catheter, Ultrasound devices →