Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL6560Recalls & reported problems
Shockwave Medical, Inc. · The Shockwave M5 Intravascular Lithotripsy (IVL) Catheter consists of an array of miniaturized lithotripsy emitters, located inside a balloon, that generate acoustic pressure waves within the target treatment site, disrupting calcium within the lesion and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. Two radiopaque marker bands inside the balloon aid in positioning. The proximal hub has three ports: one for inflation/deflation of the balloon with 50/50 saline/contrast medium, one for the ‘Over-the-wire’ (OTW) guidewire lumen, and one for connection to the IVL Connector Cable (and IVL Generator). · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M732M5IVL65600), received between Apr 3, 2020 and Mar 10, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Difficult to Remove1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2021Difficult to Remove; Difficult to Remove
A SHOCKWAVE INTRAVASCULAR LITHOTRIPSY (IVL) M5 6.5X60 DEVICE WAS USED FOR A SHORT (3MM), STRONGLY CALCIFIED CFA LESION (PROFUNDA). THE LESION WAS DIFFICULT TO CROSS WITH THE WIRE, BUT WAS ULTIMATELY ACHIEVED VIA RETROGRADE. THE VESSEL WAS PRE-DILATATED FIRST WITH 4.0MM BALLOON, FOLLOWED BY TREATMENT WITH THE IVL 6.5MM CATHETER WITH 4 CYCLES WITH A GOOD RESULT. THE PHYSICIAN EXPERIENCED RESISTANCE IN PULLING BACK THE IVL CATHETER TO THE TIP OF THE INTRODUCER SHEATH; THE PHYSICIAN ATTEMPTED TO PULL A LITTLE BIT HARDER WITHOUT SUCCESS SO THE SHEATH AND IVL CATHETER WERE REMOVED TOGETHER. FOLLOWING SUCCESSFUL REMOVAL OF THE IVL DEVICE, A NEW CROSSOVER SHEATH WAS ADVANCED, DURING WHICH A GUIDEWIRE SEPARATION WAS OBSERVED INSIDE THE PATIENT. NEW RETROGRADE ACCESS WAS OBTAINED, AND THE BROKEN WIRE WAS SUCCESSFULLY SNARED AND REMOVED FROM THE PATIENT. THE PHYSICIAN BELIEVES THAT THE GUIDEWIRE DAMAGE WAS INDIRECTLY CONNECTED WITH THE REMOVAL OF THE IVL CATHETER DUE TO THE FACT THAT THE FORCE REQUIRED TO PULL ON THE CATHETER WAS STRONGER THAN USUAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11457880 ↗MalfunctionApr 3, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
THERE IS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THE REPORT. THE SPECIFIC LESION TREATED IS UNKNOWN, BUT IT HAS BEEN CONFIRMED THAT THE PATIENT WAS TREATED WITH SHOCKWAVE M5 PERIPHERAL CATHETER. AT POST-PROCEDURE, IT WAS CONFIRMED THAT THE DISTAL PORTION OF THE CATHETER HAD TORN OFF. THE PROCEDURE INVOLVED A SHOCKWAVE DEVICE WAS USED ALONG WITH A 7F TERUMO SHEATH. CATHETER WAS COMPLETELY REMOVED WITH THE SHEATH. THE IVL RESULT WAS GOOD. ACCORDING TO THE REPORT, THERE WAS NO DEATH, SERIOUS INJURIES OR HARM ON THE PATIENT. IT WAS ALSO REPORTED THAT THERE WAS NO ADDITIONAL INTERVENTION REQUIRED. ADDITIONAL HOSPITALIZATION WAS NOT REQUIRED. THE SCENARIO OF THE GIVEN CASE POTENTIALLY COULD HAVE LED TO DEVICE EMBOLISM INTO THE PATIENT.
Patient problems: No Information
MAUDE report 9924705 ↗
Other Percutaneous Catheter, Ultrasound devices
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL7060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound8 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL080030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound7 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL120030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound3 deaths7 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL100030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 deaths6 reports
- SEISMIQ™ FG-010978Bolt Medical, Inc. · Percutaneous Catheter, Ultrasound3 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL5060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- SEISMIQ™ FG-006917Bolt Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL7060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page