Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL100030Recalls & reported problems
Shockwave Medical, Inc. · The IVL Catheter is a proprietary lithotripsy device delivered through the peripheral arterial system of the lower extremities to the site of an otherwise difficult to treat calcified stenosis. Energizing the lithotripsy device will generate pulsatile mechanical energy within the target treatment site, disrupting calcium within the lesion and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. The IVL Catheter is comprised of an integrated balloon with an array of lithotripsy emitters for the localized delivery of pulsatile mechanical energy. The system consists of an IVL Catheter, an IVL Connector Cable and an IVL Generator. The IVL Catheter is available in four (4) sizes: 8.0 x 30mm, 9.0 x 30mm, 10.0 x 30mm, and 12.0 x 30mm. The L6 Peripheral IVL Catheter is compatible with a 7 or 8F sheath and has a working length of 110cm. The Shockwave L6 Peripheral IVL Catheter shaft contains an inflation lumen, a guidewire lumen, and the lithotripsy emitters. The inflation lumen is used for inflation and deflation of the balloon with 50/50 saline/contrast medium. The guidewire lumen enables the use of a 0.018” guidewire to facilitate advancement of the catheter to and through the target stenosis. The system is designed as ‘Over-the-wire’ (OTW) with 110cm shaft working length, so an exchange length (300cm) guidewire is indicated. The emitters are positioned along the length of the balloon working length for delivery of pulsatile mechanical energy. The balloon is located near the distal tip of the catheter. Two radiopaque marker bands within the balloon denote the length of the balloon to aid in positioning of the balloon during treatment. The balloon is designed to provide an expandable segment of known length and diameter at a specific pressure. The proximal hub has three ports: one for inflation/deflation of the balloon, one for guidewire lumen, and one for connection to the IVL Connector Cable. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 2
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00195451000409), received between Jun 6, 2023 and Aug 13, 2026 — classified as 2 deaths, 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Detachment of Device or Device Component2
- Difficult to Remove1
- Improper or Incorrect Procedure or Method1
- Material Split, Cut or Torn1
- Off-Label Use1
Patient problems
- Perforation of Vessels2
- Hematoma1
- No Clinical Signs, Symptoms or Conditions1
- Perforation1
- Retroperitoneal Hemorrhage1
- Vascular Dissection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50244968 ↗InjuryJun 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO TREAT A LESION IN THE ILIAC ARTERIES. 300 IVL PULSES WERE DELIVERED. WHEN THE BALLOON WAS DEFLATED, IT WAS DISCOVERED THAT THE ILIAC ARTERY HAD A PERFORATION. STENTING AND AN ILIAC ARTERY GRAFT SURGERY HAD TO BE PERFORMED TO TREAT THE PERFORATION. NO ADDITIONAL PATIENT HARM WAS REPORTED.
Patient problems: Perforation of Vessels
MAUDE report 22190914 ↗InjuryApr 25, 2025Difficult to Remove; Improper or Incorrect Procedure or Method; Difficult to Remove; Improper or Incorrect Procedure or Method
A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO TREAT A CALCIFIED LESION IN THE ILIAC ARTERIES. THE PHYSICIAN SELECTED A 7FR SHEATH FOR THE IVL CATHETER WHICH IS SMALLER THAN THE RECOMMENDED SHEATH SIZE OF 8FR FOR A 10.0MM CATHETER PER THE DEVICE INSTRUCTIONS FOR USE. DURING THE PROCEDURE, 300 PULSES WERE SUCCESSFULLY DELIVERED. UPON REMOVAL OF THE IVL CATHETER, THE CATHETER BECAME STUCK IN THE 7FR SHEATH AND WAS NOT ABLE TO BE REMOVED. THE PHYSICIAN THEN USED A 20CC SYRINGE TO PULL NEGATIVE AND ATTEMPTED TO PULL OUT THE IVL CATHETER BUT ULTIMATELY HAD TO PULL THE ENTIRE SHEATH OUT. THE BALLOON DID NOT SEPARATE FROM THE CATHETER[?]THERE WAS NO KINKING OR FRAGMENTATION OF THE DEVICE, AND THE BALLOON REMAINED INTACT. WHEN THE PHYSICIAN PUT THE SHEATH BACK IN, THEY NOTICED THEY HAD DAMAGED THE TIP OF THE SHEATH. THEREFORE, A NEW SHEATH WAS OPENED BUT COULD NOT BE SUCCESSFULLY NAVIGATED BACK THROUGH THE ACCESS SITE. AN ANGIOGRAM REVEALED A PERFORATION OF THE EXTERNAL ILIAC ARTERY. SUBSEQUENTLY, BRACHIAL ACCESS WAS OBTAINED, AND A VIABAHN STENT WAS DEPLOYED TO SEAL THE PERFORATION, FOLLOWED BY POST-DILATION BALLOONING. A POST-DILATION BALLOON WAS USED AFTE
Patient problems: Perforation of Vessels
MAUDE report 21914537 ↗InjuryMar 20, 2024Detachment of Device or Device Component; Material Split, Cut or Torn; Detachment of Device or Device Component; Material Split, Cut or Torn
A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO TREAT A LESION IN THE RIGHT AND LEFT COMMON ILIAC ARTERIES. 280 PULSES WERE SUCCESSFULLY DELIVERED BEFORE A BALLOON LOSS OF PRESSURE OCCURRED. UPON REMOVAL, THE BALLOON BECAME DETACHED FROM THE IVL CATHETER. THE PROXIMAL END CAME OUT OVER THE SHEATH WHILE THE DISTAL TIP OF THE BALLOON REMAINED IN THE BODY AND WAS STUCK ON THE SHEATH. SNARING WAS PERFORMED TO REMOVE THE DETACHED PORTION OF THE IVL CATHETER. A DISSECTION/INJURY TO THE LEFT CFA WAS OBSERVED AFTER MULTIPLE SHEATH EXCHANGES AND SNARING OF THE DISTAL PORTION OF THE IVL CATHETER. THE PATIENT WAS SENT FROM THE CATHETER LAB TO THE OPERATING ROOM (OR) FOR A SURGICAL COMMON FETAL ARTERY (CFA) REPAIR. THE PATIENT HAD A SUCCESSFUL REPAIR OF THEIR CFA AND HAS SINCE BEEN DISCHARGED.
Patient problems: Vascular Dissection
MAUDE report 18947209 ↗DeathJul 10, 2023Off-Label Use; Adverse Event Without Identified Device or Use Problem; Off-Label Use; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A MITRAL VALVULOPLASTY PROCEDURE TO TREAT MITRAL STENOSIS USING THREE SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY CATHETERS. THE PROCEDURE WAS PERFORMED OFF-LABEL AND THE PHYSICIAN WAS AWARE OF IT. THE IVLS WERE INTRODUCED INTO THE PATIENT VIA TRANSSEPTAL PUNCTURE. DURING THE PROCEDURE AND AFTER 60 PULSES WERE DELIVERED, THE PATIENT'S BLOOD PRESSURE DROPPED. THE PHYSICIAN IMMEDIATELY REALIZED THAT THE 0.018" STEEL CORE WIRE PERFORATED THE LEFT VENTRICLE (LV). THE PATIENT WAS TRANSFERRED TO THE OPERATING ROOM (OR) FOR OPEN-HEART SURGERY TO FIX THE PERFORATION AND THEN MOVED TO THE INTENSIVE CARE UNIT (ICU). WHILE IN THE ICU, THE PATIENT STARTED TO BLEED AGAIN AND THEN SUBSEQUENTLY EXPIRED. ALL THREE IVL DEVICES FUNCTIONED AS INTENDED AND NO MALFUNCTION WAS REPORTED.
Patient problems: Perforation
MAUDE report 17290260 ↗DeathJun 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A PROCEDURE WHERE TWO SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETERS WERE USED AS PART OF AN OVERALL ENDOVASCULAR TREATMENT PLAN. THE CATHETERS WERE USED DURING THE CASE TO TREAT SEVERELY CALCIFIED LESIONS IN THE RIGHT AND LEFT COMMON ILIAC ARTERIES. THE IVL BALLOONS FUNCTIONED PROPERLY DURING THE CASE, DELIVERING 60 PULSES EACH AT APPROXIMATELY 4 ATMOSPHERES. THE BALLOONS WERE USED TO PRE-TREAT EACH ARTERY BEFORE DEPLOYING STENT GRAFT IN EACH ARTERY. REPORTEDLY, POST PROCEDURE, THE PATIENT EXPERIENCED RETROPERITONEAL BLEED. THE PATIENT WAS BROUGHT BACK IN THE CATH LAB HOWEVER, THE PATIENT EXPIRED.
Patient problems: Hematoma; Retroperitoneal Hemorrhage
MAUDE report 17074891 ↗
Other Percutaneous Catheter, Ultrasound devices
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL7060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound8 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL080030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound7 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL120030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound3 deaths7 reports
- SEISMIQ™ FG-010978Bolt Medical, Inc. · Percutaneous Catheter, Ultrasound3 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL5060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL6560Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- SEISMIQ™ FG-006917Bolt Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL7060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026