Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL120030Recalls & reported problems
Shockwave Medical, Inc. · The IVL Catheter is a proprietary lithotripsy device delivered through the peripheral arterial system of the lower extremities to the site of an otherwise difficult to treat calcified stenosis. Energizing the lithotripsy device will generate pulsatile mechanical energy within the target treatment site, disrupting calcium within the lesion and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. The IVL Catheter is comprised of an integrated balloon with an array of lithotripsy emitters for the localized delivery of pulsatile mechanical energy. The system consists of an IVL Catheter, an IVL Connector Cable and an IVL Generator. The IVL Catheter is available in four (4) sizes: 8.0 x 30mm, 9.0 x 30mm, 10.0 x 30mm, and 12.0 x 30mm. The L6 Peripheral IVL Catheter is compatible with a 7 or 8F sheath and has a working length of 110cm. The Shockwave L6 Peripheral IVL Catheter shaft contains an inflation lumen, a guidewire lumen, and the lithotripsy emitters. The inflation lumen is used for inflation and deflation of the balloon with 50/50 saline/contrast medium. The guidewire lumen enables the use of a 0.018” guidewire to facilitate advancement of the catheter to and through the target stenosis. The system is designed as ‘Over-the-wire’ (OTW) with 110cm shaft working length, so an exchange length (300cm) guidewire is indicated. The emitters are positioned along the length of the balloon working length for delivery of pulsatile mechanical energy. The balloon is located near the distal tip of the catheter. Two radiopaque marker bands within the balloon denote the length of the balloon to aid in positioning of the balloon during treatment. The balloon is designed to provide an expandable segment of known length and diameter at a specific pressure. The proximal hub has three ports: one for inflation/deflation of the balloon, one for guidewire lumen, and one for connection to the IVL Connector Cable. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 3
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00195451000416), received between Oct 2, 2024 and Sep 8, 2026 — classified as 3 deaths, 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Detachment of Device or Device Component2
- Material Split, Cut or Torn2
- Failure to Cycle1
- Inflation Problem1
- Leak/Splash1
- Stretched1
Patient problems
- Hemorrhage/Blood Loss/Bleeding5
- Low Blood Pressure/ Hypotension3
- Perforation of Vessels3
- Cardiac Arrest1
- No Clinical Signs, Symptoms or Conditions1
- Rupture1
- Shock1
- Vascular Dissection1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 8, 2026Detachment of Device or Device Component; Material Split, Cut or Torn
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50679647 ↗DeathMay 22, 2026Adverse Event Without Identified Device or Use Problem
THIS EVENT WAS REPORTED TO SHOCKWAVE VIA THE MEDICINES AND HEALTHCARE PRODUCTS REGULATORY AGENCY (MHRA) YELLOW CARD INCIDENT REPORT (MHRA REFERENCE#: 2026/005/008/501/009) INVOLVING TWO SHOCKWAVE L6 PERIPHERAL IVL CATHETERS USED FOR TREATMENT OF A PATIENT'S HEAVILY CALCIFIED ILIAC ARTERIES. THE FIRST BALLOON WILL BE REPORTED UNDER (B)(6) AND THE SECOND UNDER (B)(6) (MDR 3015053858-2026-00044). IT WAS REPORTED THAT DURING A HYBRID PROCEDURE CONSISTING OF COMMON FEMORAL ENDARTERECTOMY WITH PATCHPLASTY AND ENDOVASCULAR ANGIOPLASTY WITH LITHOTRIPSY, KISSING IVL BALLOONS WERE PLACED IN THE ILIAC ARTERIES. THE FIRST SHOCKWAVE L6 12.0MM PERIPHERAL IVL CATHETER SUCCESSFULLY DELIVERED TEN IVL TREATMENT CYCLES (THREE HUNDRED (300) PULSES) TO THE LEFT ILIAC ARTERY WITHOUT INCIDENT. THE RIGHT ILIAC ARTERY WAS PRE-DILATATED WITH A BALLOON, FOLLOWED BY A SECOND SHOCKWAVE L6 10.0MM PERIPHERAL IVL CATHETER WHICH WAS INFLATED AT 2ATM AND DELIVERED ONE IVL TREATMENT CYCLE (THIRTY (30) PULSES) TO THE RIGHT ILIAC ARTERY. THE PATIENT THEN EXPERIENCED A SUDDEN DROP IN BLOOD PRESSURE DUE TO AN RUPTURE IN THE RIGHT ILIAC ARTERY DURING THE SECOND CYCLE. FOLLOWING THE RUPTURE OF THE RIGHT ILIAC ARTERY, THE
Patient problems: Hemorrhage/Blood Loss/Bleeding; Low Blood Pressure/ Hypotension; Perforation of Vessels
MAUDE report 25261110 ↗DeathMay 22, 2026Adverse Event Without Identified Device or Use Problem
THIS EVENT WAS REPORTED TO SHOCKWAVE VIA THE MEDICINES AND HEALTHCARE PRODUCTS REGULATORY AGENCY (MHRA) YELLOW CARD INCIDENT REPORT (MHRA REFERENCE# 2026/005/008/501/009) INVOLVING TWO SHOCKWAVE L6 PERIPHERAL IVL CATHETERS USED FOR TREATMENT OF A PATIENT'S HEAVILY CALCIFIED ILIAC ARTERIES. THE FIRST BALLOON WILL BE REPORTED UNDER (B)(6) AND THE SECOND UNDER (B)(6), (3015053858-2026-00045). IT WAS REPORTED THAT DURING A HYBRID PROCEDURE CONSISTING OF COMMON FEMORAL ENDARTERECTOMY WITH PATCHPLASTY AND ENDOVASCULAR ANGIOPLASTY WITH LITHOTRIPSY, KISSING IVL BALLOONS WERE PLACED IN THE ILIAC ARTERIES. THE FIRST SHOCKWAVE L6 12.0MM PERIPHERAL IVL CATHETER SUCCESSFULLY DELIVERED TEN IVL TREATMENT CYCLES (THREE HUNDRED (300) PULSES) TO THE LEFT ILIAC ARTERY WITHOUT INCIDENT. THE RIGHT ILIAC ARTERY WAS PRE-DILATATED WITH A BALLOON, FOLLOWED BY A SECOND SHOCKWAVE L6 10.0MM PERIPHERAL IVL CATHETER WHICH WAS INFLATED AT 2ATM AND DELIVERED ONE IVL TREATMENT CYCLE (THIRTY (30) PULSES) TO THE RIGHT ILIAC ARTERY. THE PATIENT THEN EXPERIENCED A SUDDEN DROP IN BLOOD PRESSURE DUE TO AN RUPTURE IN THE RIGHT ILIAC ARTERY DURING THE SECOND CYCLE. FOLLOWING THE RUPTURE OF THE RIGHT ILIAC ARTERY, THE PATI
Patient problems: Hemorrhage/Blood Loss/Bleeding; Low Blood Pressure/ Hypotension; Perforation of Vessels
MAUDE report 25260973 ↗DeathApr 3, 2026Failure to Cycle
A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO AID IN PREPARATION FOR REMOVAL OF HIGHLY CALCIFIED 21-YEAR-OLD LEADS. THE IVL CATHETER WAS PREPPED USING A 65/35 CONTRAST MIXTURE. ALL CONNECTIONS BETWEEN THE CATHETER, CONNECTOR CABLE, AND GENERATOR WERE CONFIRMED SECURE. HOWEVER, THERE WAS AN ISSUE FULLY PURGING THE IVL DEVICE. AIR CONTINUED TO RETURN UNTIL THE INDEFLATOR AND STOPCOCK WERE CHANGED. AN ERROR 88 WAS ENCOUNTERED WITH THE USE OF THE IVL CATHETER, BUT IT IS UNKNOWN WHEN THE ERROR APPEARED. THUS, A SECOND IVL CATHETER WAS USED TO COMPLETE IVL TREATMENT. SEVERAL HOURS LATER, THE PATIENT EXPERIENCED DESTABILIZATION DURING AND AFTER LEAD REMOVAL ATTEMPTS. THE PATIENT PASSED AWAY 6 HOURS AFTER THE PROCEDURE WAS COMPLETED. THE PHYSICIAN DID NOT ATTRIBUTE ANY PATIENT SEQUALAE TO THE IVL DEVICES, AND BELIEVES THAT THE PATIENT'S DEATH RESULTED FROM A TEAR IN THE SUPERIOR VENA CAVA (SVC) AFTER THE ATTEMPTED LEAD IMPLANT, WITH THE SVC APPEARING TO HAVE BEEN DAMAGED DURING LEAD REMOVAL. THERE WAS NO FORMAL AUTOPSY PERFORMED, ALTHOUGH THE CHEST WAS OPENED AND THE TEAR WAS IDENTIFIED DURING THAT INSPECTION.
Patient problems: Vascular Dissection
MAUDE report 24776267 ↗InjuryFeb 26, 2026Inflation Problem; Leak/Splash; Stretched; Material Split, Cut or Torn
IT WAS REPORTED THAT A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED DURING A PROCEDURE TO TREAT A LESION IN THE ILIAC ARTERY. A 75-YEAR-OLD MALE PATIENT WAS HOSPITALIZED FOR A THROMBOENDARTERECTOMY OF THE FEMORAL BIFURCATION WITH USE OF A SHOCKWAVE IVL BALLOON. DURING THE PROCEDURE, THE PATIENT EXPERIENCED A SUDDEN DROP IN BLOOD PRESSURE DUE TO A RUPTURE IN THE ILIAC ARTERY. SHOCKWAVE IVL THERAPY WAS TEMPORARILY STOPPED TO MANAGE THE HEMORRHAGIC SHOCK. AFTER THE PATIENT MAINTAINED A SATISFACTORY MEAN ARTERIAL PRESSURE, SHOCKWAVE IVL THERAPY RESUMED. THE ORIGINAL IVL DEVICE WAS SUBSEQUENTLY REINSERTED FOLLOWING THE PATIENT'S HEART ATTACK DURING THE PROCEDURE. ACCESS WAS VIA AN IPSI-LATERAL RIGHT FEMORAL PUNCTURE USING AN 18FR INTRODUCER SHEATH AND A 0.018" GUIDEWIRE WITH A STANDARD ENDOVASCULAR ANEURYSM REPAIR (EVAR) SETUP; NO TAVR EQUIPMENT WAS USED. DURING REINFLATION OF THE IVL BALLOON, THE BALLOON DID NOT INFLATE AND A MAJOR LEAK OCCURRED AT THE JUNCTION BETWEEN THE DELIVERY CATHETER AND THE LATERAL INFLATION PORT. THREE IVL PULSE CYCLES WERE DELIVERED. WHEN THE PHYSICIAN ATTEMPTED TO REMOVE THE IVL BALLOON, SIGNIFICANT ELONGATION OF THE DELIVERY CAT
Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding; Low Blood Pressure/ Hypotension; Perforation of Vessels; Shock
MAUDE report 24460778 ↗MalfunctionNov 4, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO TREAT A LESION IN THE ILIAC ARTERIES. THE IVL CATHETER DELIVERED 300 PULSES. IT WAS REPORTED THAT THE BALLOON PORTION OF THE CATHETER DETACHED FROM THE SHAFT UPON REMOVAL. THE BALLOON REMAINED ON THE 0.018" GUIDEWIRE, ALLOWING REMOVAL OF BOTH THE CATHETER AND BALLOON TOGETHER. A STENT WAS PLACED AS PLANNED, AND NO ADDITIONAL INTERVENTIONS WERE NECESSARY. THE FACILITY REPORTED THAT THE DETACHMENT OCCURRED WHILE THE GUIDEWIRE WAS STILL INSIDE THE PATIENT; HOWEVER, THERE WAS NO ADVERSE EVENT TO THE PATIENT, AND NO FRAGMENTS REMAINED IN THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20604323 ↗InjuryOct 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO TREAT A LESION IN THE ILIAC ARTERIES. THE VESSEL WAS PRE-DILATED WITH A 3.5MM X 100 MM STERLING BALLOON FOLLOWED BY A 4.0 STERLING BALLOON. DURING IVL TREATMENT, THE 12MM IVL CATHETER WAS PLACED IN THE PORTION OF THE VESSEL THAT MEASURED 7-8MM, CAUSING THE ARTERY TO RUPTURE. ABDOMINAL HEMORRHAGE AND LOSS OF BLOOD WAS ALSO NOTED. THE PATIENT REQUIRED CPR (CARDIOPULMONARY RESUSCITATION) AND A BLOOD TRANSFUSION. THE PHYSICIAN USED COVERED STENTS TO REPAIR THE RUPTURE. THE PATIENT WAS DOING FINE FOLLOWING THE PROCEDURE. WHILE THERE WAS NO REPORT OF IVL CATHETER MALFUNCTION, IT WAS REPORTED THAT THE PHYSICIAN WAS AWARE THAT THE 12MM IVL CATHETER WAS 40-50% OVERSIZED FOR THE PATIENT'S VESSEL SEGMENT AND THAT THE USE WAS OFF-LABEL.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Rupture
MAUDE report 20355904 ↗
Other Percutaneous Catheter, Ultrasound devices
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL7060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound8 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL080030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound7 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL100030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 deaths6 reports
- SEISMIQ™ FG-010978Bolt Medical, Inc. · Percutaneous Catheter, Ultrasound3 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL5060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL6560Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- SEISMIQ™ FG-006917Bolt Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL7060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page