Skip to content

SEISMIQ™ FG-006917Recalls & reported problems

Bolt Medical, Inc. · Intravascular Lithotripsy Catheter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850056723053), received between Jul 16, 2026 and Sep 22, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Nonstandard Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Nonstandard Device2
  • Connection Problem1
  • Fluid/Blood Leak1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Fluid/Blood Leak; Nonstandard Device

    Reportable based on device analysis completed on(b)(6)2026. It was reported that the balloon leaked. A 6.0X60 145CM SEISMIQ device was selected for use in an Intravascular Lithotripsy (IVL) procedure. During the procedure, a balloon leak was identified. The procedure was completed with the same device. There were no patient complications as a result of this event and the patient fully recovered.However, device analysis revealed a balloon tear/rupture along with a device fragment.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857639 ↗
  • MalfunctionJul 16, 2026Nonstandard Device; Connection Problem; Output Problem

    IT WAS REPORTED THAT THE CATHETER DID NOT DISPLAY INFLATION PRESSURE ON THE CONSOLE DESPITE THE BALLOON BEING INFLATED. A SEISMIQ IVL CATHETER WAS SELECTED FOR USE TO TREAT AN 80% STENOSED LESION WITH SEVERE CONCENTRIC CALCIFICATION AND MILD VESSEL TORTUOSITY IT WAS REPORTED THAT, AFTER THE BALLOON WAS DEFLATED USING THE INFLATOR AND SUBSEQUENTLY REINFLATED TO 4.0 ATMOSPHERES, THE PRESSURE DID NOT REGISTER ON THE CONSOLE SCREEN; HOWEVER, THE INFLATOR INDICATED 4.0 ATMOSPHERES, AND THE BALLOON WAS OBSERVED TO BE INFLATED. AN ALERT MESSAGE STATING "REMOVE AND REINSERT CATHETER CONNECTOR" WAS DISPLAYED ON THE CONSOLE. TROUBLESHOOTING WAS ATTEMPTED BY UNPLUGGING AND REINSERTING THE CATHETER AND BY DEFLATING AND REINFLATING THE BALLOON; HOWEVER, THE ISSUE WAS NOT RESOLVED. THE DEVICE WAS EXCHANGED, AND THE PROCEDURE WAS COMPLETED WITHOUT PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25847626 ↗

Other Percutaneous Catheter, Ultrasound devices

All Percutaneous Catheter, Ultrasound devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page