SEISMIQ™ FG-006917Recalls & reported problems
Bolt Medical, Inc. · Intravascular Lithotripsy Catheter · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850056723053), received between Jul 16, 2026 and Sep 22, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Nonstandard Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Nonstandard Device2
- Connection Problem1
- Fluid/Blood Leak1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Fluid/Blood Leak; Nonstandard Device
Reportable based on device analysis completed on(b)(6)2026. It was reported that the balloon leaked. A 6.0X60 145CM SEISMIQ device was selected for use in an Intravascular Lithotripsy (IVL) procedure. During the procedure, a balloon leak was identified. The procedure was completed with the same device. There were no patient complications as a result of this event and the patient fully recovered.However, device analysis revealed a balloon tear/rupture along with a device fragment.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50857639 ↗MalfunctionJul 16, 2026Nonstandard Device; Connection Problem; Output Problem
IT WAS REPORTED THAT THE CATHETER DID NOT DISPLAY INFLATION PRESSURE ON THE CONSOLE DESPITE THE BALLOON BEING INFLATED. A SEISMIQ IVL CATHETER WAS SELECTED FOR USE TO TREAT AN 80% STENOSED LESION WITH SEVERE CONCENTRIC CALCIFICATION AND MILD VESSEL TORTUOSITY IT WAS REPORTED THAT, AFTER THE BALLOON WAS DEFLATED USING THE INFLATOR AND SUBSEQUENTLY REINFLATED TO 4.0 ATMOSPHERES, THE PRESSURE DID NOT REGISTER ON THE CONSOLE SCREEN; HOWEVER, THE INFLATOR INDICATED 4.0 ATMOSPHERES, AND THE BALLOON WAS OBSERVED TO BE INFLATED. AN ALERT MESSAGE STATING "REMOVE AND REINSERT CATHETER CONNECTOR" WAS DISPLAYED ON THE CONSOLE. TROUBLESHOOTING WAS ATTEMPTED BY UNPLUGGING AND REINSERTING THE CATHETER AND BY DEFLATING AND REINFLATING THE BALLOON; HOWEVER, THE ISSUE WAS NOT RESOLVED. THE DEVICE WAS EXCHANGED, AND THE PROCEDURE WAS COMPLETED WITHOUT PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25847626 ↗
Other Percutaneous Catheter, Ultrasound devices
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL7060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound8 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL080030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound7 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL120030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound3 deaths7 reports
- Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL100030Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 deaths6 reports
- SEISMIQ™ FG-010978Bolt Medical, Inc. · Percutaneous Catheter, Ultrasound3 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL5060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL6560Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
- Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL7060Shockwave Medical, Inc. · Percutaneous Catheter, Ultrasound2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page