ARCHITECT 1P32-25Recalls & reported problems
Abbott Laboratories · ARCHITECT iDigoxin Reagent Kit · Class 2
All models of ARCHITECT (51)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380740001766), received between Sep 20, 2018 and Sep 20, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- High Test Results1
Patient problems
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 20, 2018High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED DIGOXIN (LN 1P32) RESULT FOR ONE PATIENT ON THE ARCHITECT I2000SR ANALYZER. THE FOLLOWING DATA WAS PROVIDED FROM (B)(6): SID (B)(6), INITIAL 3.5 NMOL/L. THE DOCTOR TOOK THE PATIENT OFF DIGOXIN FOR 2 WEEKS TO MONITOR, YET THE PATIENT CONTINUED TO HAVE CRITICALLY HIGH LEVELS OF DIGOXIN. ON (B)(6): RESULT: 3.5 NMOL/L; AT ANOTHER HOSPITAL (ON SIEMENS ANALYZER) SAMPLE RESULTED AS 0.2 NMOL/L. ON (B)(6): RESULT: 3.2 NMOL/L; AT ANOTHER HOSPITAL (UNKNOWN METHOD) THE SAMPLE RESULTED AS 0.3 NMOL/L. NO OTHER PATIENT DETAILS WERE PROVIDED.
Patient problems: No Code Available
MAUDE report 7895367 ↗
Other ENZYME IMMUNOASSAY, DIGOXIN devices
- Dimension Vista® Flex® reagent cartridge DIGXN 10488927Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN15 reports
- Digoxin 20737836322Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN5 reports
- ONLINE TDM Digoxin 08057613190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN4 reports
- Atellica CH Dgn 11097526Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN2 reports
- Access Digoxin 33710Beckman Coulter, Inc. · Enzyme Immunoassay, Digoxin1 report
- ARCHITECT 01P3227Abbott GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
- Syva® Emit® 2000 Digoxin Assay 10445315Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN1 report
- Elecsys Digoxin 07027214190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page