ONLINE TDM Digoxin 08057613190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613336121214), received between Dec 2, 2022 and Oct 4, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Low Test Results1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 4, 2024Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE DIGOXIN RESULTS FROM THE COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 0.00 NG/ML WITH A DATA FLAG. THE SAMPLE WAS REPEATED 10 TIMES AND THE RESULT WAS ALWAYS 0.00 NG/ML WITH A DATA FLAG. THE SAMPLE WAS REPEATED ON A UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM (BECKMAN) INSTRUMENT AND THE RESULT WAS 0.33 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20369679 ↗MalfunctionMar 15, 2024High Test Results; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE DIGOXIN (DIG) RESULTS FROM THE COBAS C 503 ANALYTICAL UNIT THAT SIS NOT MATCH THE CLINICAL PICTURE FOR THE PATIENTS. THE CUSTOMER ALLEGED MANY PATIENT SAMPLES HAD RESULTS AROUND 0.4 AND 0.6 WITHOUT ANY DIGOXIN TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18914795 ↗MalfunctionMar 7, 2024High Test Results; High Test Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE DIGOXIN RESULT FROM THE COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 0.45 G/L FOR A SAMPLE THAT WAS EXPECTED TO HAVE A NEGATIVE RESULT. AS THE RESULTS DID NOT AGREE WITH THE PATIENT'S CLINICAL PICTURE, THE SAMPLE WAS SENT TO ANOTHER LABORATORY WHERE IT WAS TESTED WITH ABBOTT METHODOLOGY AND WITH MASS SPECTROMETRY. NO SPECIFIC DATA WAS PROVIDED, BUT THE RESULTS WERE CONSIDERED TO BE NEGATIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18857499 ↗MalfunctionDec 2, 2022Low Test Results; Low Test Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED FOR DIGOXIN (DIG) ON A COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL RESULT OF THE FIRST SAMPLE TESTED ON THE C503 MODULE WAS < 0.3¿G/L. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY WHERE IT WAS QUESTIONED BY THE DOCTOR AS IT WAS NOT PLAUSIBLE. THE SAMPLE WAS REPEATED ON A COBAS C6000 SYSTEM WITH A RESULT OF < 0.3¿G/L. THE INITIAL RESULT FOR THE SECOND SAMPLE (NEW BLOOD DRAWN ON 28-OCT-2022) TESTED ON THE C503 MODULE WAS < 0.3¿G/L. THE SAMPLE WAS REPEATED ON A COBAS C6000 SYSTEM WITH A RESULT OF < 0.3¿G/L. THE PATIENT WAS SENT TO ANOTHER LABORATORY WHERE A NEW SAMPLE WAS OBTAINED AND THE RESULT FROM A COBAS 8000 SYSTEM WAS 0.76 UG/L. A THIRD SAMPLE WAS OBTAINED ON 03-NOV-2022. A SERUM SAMPLE AND A LITHIUM HEPARIN SAMPLE WERE OBTAINED AT THE SAME TIME. THE SERUM SAMPLE WAS TESTED ON THE C503 MODULE WITH A RESULT OF 0.3 UG/L. THE LITHIUM HEPARIN SAMPLE WAS SENT TO THE OTHER LABORATORY AND THE RESULT FROM A COBAS 8000 SYSTEM WAS 1.0 UG/L. THE HIGH RESULTS FROM THE OTHER LABORATORY (1.0 G/L AND 0.76 G/L) WERE CONSIDERED TO BE CORRECT. THE C503 MODULE SERIAL NUMBER WAS (B)(4). THE SERIAL NUMBERS FOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15910805 ↗
Other ENZYME IMMUNOASSAY, DIGOXIN devices
- Dimension Vista® Flex® reagent cartridge DIGXN 10488927Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN15 reports
- Digoxin 20737836322Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN5 reports
- Atellica CH Dgn 11097526Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN2 reports
- Access Digoxin 33710Beckman Coulter, Inc. · Enzyme Immunoassay, Digoxin1 report
- ARCHITECT 01P3227Abbott GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
- Syva® Emit® 2000 Digoxin Assay 10445315Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN1 report
- ARCHITECT 1P32-25Abbott Laboratories · ENZYME IMMUNOASSAY, DIGOXIN1 report
- Elecsys Digoxin 07027214190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page