Access Digoxin 33710Recalls & reported problems
Beckman Coulter, Inc. · The Access Digoxin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of digoxin level in human serum ising the Access Immunoassay Systems. · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15099590225797), received between Oct 2, 2017 and Oct 2, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 2, 2017Low Test Results; Low Test Results
THE CUSTOMER REPORTED OBTAINING NON-REPRODUCIBLE DIGOXIN (ACCESS DIGOXIN) RESULTS ON THE LABORATORY'S ACCESS 2 IMMUNOASSAY SYSTEM (SERIAL NUMBER (B)(4)) FOR ONE PATIENT. THE PATIENT'S SAMPLE WAS ANALYZED IN TRIPLICATE AND RECOVERED WITH IMPRECISE ACCESS DIGOXIN RESULTS. THESE RESULTS WERE NOT RELEASED FROM THE LABORATORY. THERE WAS NO IMPACT OR CHANGE TO PATIENT CARE OR TREATMENT IN ASSOCIATION WITH THE NON-REPRODUCIBLE ACCESS DIGOXIN RESULTS. QUALITY CONTROL (QC) WAS RECOVERING WITHIN SPECIFICATIONS PRIOR TO THE EVENT. THE CUSTOMER REPORTED THAT QC WAS OUT OF SPECIFICATIONS THE DAY FOLLOWING THIS EVENT. THE PATIENT'S SAMPLE WAS COLLECTED IN A PLASMA SEPARATION TUBE. THE CUSTOMER WAS NOT ABLE TO PROVIDE CENTRIFUGE INFORMATION SUCH AS SPEED, TIME OR TEMPERATURE. NO ISSUES WITH SAMPLE INTEGRITY WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6908255 ↗
Other Enzyme Immunoassay, Digoxin devices
- Dimension Vista® Flex® reagent cartridge DIGXN 10488927Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN15 reports
- Digoxin 20737836322Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN5 reports
- ONLINE TDM Digoxin 08057613190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN4 reports
- Atellica CH Dgn 11097526Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN2 reports
- ARCHITECT 1P32-25Abbott Laboratories · ENZYME IMMUNOASSAY, DIGOXIN1 report
- Syva® Emit® 2000 Digoxin Assay 10445315Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN1 report
- ARCHITECT 01P3227Abbott GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
- Elecsys Digoxin 07027214190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.