Elecsys Digoxin 07027214190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04015630940400), received between Jan 20, 2023 and Jan 20, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- High Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 20, 2023High Test Results; High Test Results
THE INITIAL REPORTER QUESTIONED A HIGH RESULT FOR 1 PATIENT TESTED FOR ELECSYS DIGOXIN (DIGOXIN) ON A COBAS E 801 ANALYTICAL UNIT. ON (B)(6) 2022 THE RESULT FROM THE ABBOTT METHOD WAS 1.87 UG/L. ON (B)(6) 2022 THE RESULT FROM THE E801 MODULE WAS 3.22 UG/L. THIS RESULT DID NOT CORRESPOND TO THE PATIENT¿S CLINICAL CONDITION. THE RESULT FROM THE E801 MODULE WAS REPORTED OUTSIDE OF THE LABORATORY. THE CUSTOMER SUSPECTS AN INTERFERENCE AFFECTING THE ROCHE RESULT AS THE PATIENT HAS RENAL FAILURE; THIS IS MENTIONED IN THE PRODUCT LABELING FOR THE ASSAY. THE E801 MODULE SERIAL NUMBER WAS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16213761 ↗
Other ENZYME IMMUNOASSAY, DIGOXIN devices
- Dimension Vista® Flex® reagent cartridge DIGXN 10488927Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN15 reports
- Digoxin 20737836322Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN5 reports
- ONLINE TDM Digoxin 08057613190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN4 reports
- Atellica CH Dgn 11097526Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN2 reports
- Access Digoxin 33710Beckman Coulter, Inc. · Enzyme Immunoassay, Digoxin1 report
- ARCHITECT 01P3227Abbott GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
- Syva® Emit® 2000 Digoxin Assay 10445315Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN1 report
- ARCHITECT 1P32-25Abbott Laboratories · ENZYME IMMUNOASSAY, DIGOXIN1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page