Dimension Vista® Flex® reagent cartridge DIGXN 10488927Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · LOCI Digoxin Flex® reagent cartridge, 200 tests · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00842768028335), received between Dec 11, 2017 and Dec 11, 2017 — classified as 15 malfunctions.
The most frequently reported specific device problem is “High Test Results” (15 reports).
2 other recalls cover devices cleared under the same FDA submission (K093441); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results15
Patient problems
- No Known Impact Or Consequence To Patient15
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. A NEW SAMPLE WAS TESTED EARLIER AND REPEATED ON AN ALTERNATE VISTA INSTRUMENT AND HIGH RESULTS WERE OBTAINED. A NEW SAMPLE WAS DRAWN AND TESTED AGAIN ON AN ALTERNATE VISTA INSTRUMENT AND A HIGH RESULT WAS OBTAINED. A NEW SAMPLE WAS DRAWN AND TESTED AT AN ALTERNATE FACILITY USING AN ALTERNATE METHODOLOGY AND A LOWER RESULT WAS OBTAINED. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7104090 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS REPEATED ON THE SAME INSTRUMENT AND A HIGH RESULT WAS OBTAINED. A NEW DRAW WAS TESTED ON THE SAME INSTRUMENT AND AN ALTERNATE VISTA INSTRUMENT AND HIGH RESULTS WERE OBTAINED. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7104083 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7104081 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7104077 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103749 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS REPEATED TWICE ON THE SAME INSTRUMENT AND HIGH RESULTS WERE OBTAINED. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103748 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS TESTED TWICE EARLIER ON AN ALTERNATE VISTA INSTRUMENT AND HIGH RESULTS WERE OBTAINED. THE SAME SAMPLE WAS REPEATED THREE TIMES LATER ON THE SAME VISTA INSTRUMENT AND HIGH RESULTS WERE OBTAINED. A 1:3 AND 1:2 MANUAL DILUTION WAS PERFORMED ON THE LAST TWO REPEATS AND HIGH RESULTS WERE OBTAINED ON THE SAME SAMPLE. THE CLINICIAN STARTED TO QUESTION THE RESULT SINCE IT HAD INCREASED WHEN NO DIGOXIN WAS GIVEN TO THE PATIENT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103746 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103745 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103744 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS REPEATED EARLIER ON AN ALTERNATE VISTA INSTRUMENT AND A HIGH RESULT WAS OBTAINED. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103743 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103741 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS REPEATED ON THE SAME INSTRUMENT AND A HIGH RESULT WAS OBTAINED. A NEW SAMPLE WAS DRAWN AND REPEATED ON THE SAME INSTRUMENT AND ON AN ALTERNATE VISTA AND A HIGH RESULT WAS OBTAINED. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103740 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT OR THE TREATMENT CANCELLATION. THERE WAS NO HARM TO THE PATIENT RELATED TO THE USE OF THE DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103739 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103736 ↗MalfunctionDec 11, 2017High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED DIGOXIN (DIGXN) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS REPEATED ON THE SAME INSTRUMENT AND A HIGH RESULT WAS OBTAINED. THERE IS NO KNOWN REPORT OF PATIENT ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY ELEVATED DIGXN RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7103735 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K093441. One submission often covers many models, so check the product and lot details in each record.
- Z-2607-2017 · Mar 21, 2017 · TerminatedDimension Vista Digoxin (DIGXN) Flex reagent cartridge
- Z-1151-2013 · Feb 21, 2013 · TerminatedSiemens Dimension Vista(R) System Vista Drug 4 Calibrator. The device is used as an in vitro diagnostic product.
Other ENZYME IMMUNOASSAY, DIGOXIN devices
- Digoxin 20737836322Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN5 reports
- ONLINE TDM Digoxin 08057613190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN4 reports
- Atellica CH Dgn 11097526Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN2 reports
- Elecsys Digoxin 07027214190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
- ARCHITECT 01P3227Abbott GmbH · ENZYME IMMUNOASSAY, DIGOXIN1 report
- Access Digoxin 33710Beckman Coulter, Inc. · Enzyme Immunoassay, Digoxin1 report
- Syva® Emit® 2000 Digoxin Assay 10445315Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, DIGOXIN1 report
- ARCHITECT 1P32-25Abbott Laboratories · ENZYME IMMUNOASSAY, DIGOXIN1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page