Skip to content

ARCHITECT 8D06-41Recalls & reported problems

Abbott GmbH · ARCHITECT Syphilis TP Reagent Kit · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740117122), received between Sep 17, 2020 and Jul 11, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 11, 2024False Negative Result; False Negative Result

    THE CUSTOMER OBSERVED FALSE NONREACTIVE ARCHITECT SYPHILIS TP RESULTS FOR A NEWBORN WITH A POSITIVE SYPHILIS MOTHER. THE PATIENT WAS REDRAWN AND THE EXPECTED POSITIVE RESULT WAS OBTAINED. THE FOLLOWING DATA WAS PROVIDED (<1.00 S/CO IS NONREACTIVE, >/=1.00 S/CO IS REACTIVE): INITIAL RESULT = 0.87 S/CO SAMPLE WAS A SHORT SAMPLE. REDRAW SID (B)(6) RESULT = 21.62 S/CO NO IMPACT TO DONOR MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19723886 ↗
  • MalfunctionSep 17, 2020False Negative Result; False Negative Result

    THE CUSTOMER OBTAINED A FALSE NEGATIVE ARCHITECT SYPHILIS TP RESULT. THE SAMPLE GENERATED 0.13 AND RETEST ON SECOND ANALYZER 0.15 S/CO ON ARCHITECT. THE SAMPLE GENERATED POSITIVE RESULTS ON TPPA AND RPR METHODS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10545738 ↗

Other SINGLE (SPECIFIED) ANALYTE CONTROLS (ASSAYED AND UNASSAYED) devices

All SINGLE (SPECIFIED) ANALYTE CONTROLS (ASSAYED AND UNASSAYED) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.