ARCHITECT 2K91-33Recalls & reported problems
Abbott GmbH · ARCHITECT CA19-9XR Reagent Kit · Class 1
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740137304), received between Jun 26, 2018 and Jun 13, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Low Test Results2
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 13, 2024Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED ARCHITECT CA 19-9 RESULTS FOR ONE PATIENT DIAGNOSED WITH PANCREATIC CANCER. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2024, SID (B)(6) WAS SENT FROM AN OUTPATIENT FACILITY. THE ARCHITECT CA 19-9 RESULT WAS <2 U/ML, REPEATED <2 U/ML. ON (B)(6) 2024, THE SAME PATIENT WITH DIFFERENT SID (SID (B)(6) WAS SENT FROM AN OUTPATIENT FACILITY. THE ARCHITECT CA 19-9 RESULT WAS 40,243 U/ML. THE CUSTOMER RETESTED THE SAMPLE FROM (B)(6) 2024, (SID (B)(6) AND THE RETEST WAS 30,360 U/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19527385 ↗MalfunctionDec 27, 2019Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DECREASED ARCHITECT CA 19-9XR RESULT FOR 1 PATIENT WHO IS DIAGNOSED WITH PANCREATIC CANCER. THE FOLLOWING DATA WAS PROVIDED (REFERENCE RANGE: 0 TO 37 U/L): SID (B)(6) INITIAL RESULT = >1200 U/L, REPEAT WITH 1:10 AUTODILUTION = 166 U/L. SAMPLE WAS REPEAT TWICE WITHOUT A DILUTION = >1200 U/L FOR BOTH RUNS, REPEATED TWICE WITH 1:10 AUTODILUTION = 1522 U/L AND 1390 U/L. PREVIOUS CA19-9 RESULT FROM NOVEMBER WAS HIGH. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9526086 ↗MalfunctionMay 1, 2019High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CA19-9 RESULTS WHILE USING THE ARCHITECT CA19-9XR ASSAY. THE CUSTOMER PROVIDED THE FOLLOWING RESULTS: ON (B)(6) 2019: SID (B)(6): INITIAL: 251.66 U/ML, RETEST: 3.6 U/ML. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8570272 ↗MalfunctionJun 26, 2018High Test Results; High Test Results
THE FOLLOWING DATA WAS PROVIDED FROM THE ARCHITECT SID (B)(6) INITIAL 1865.3 U/ML. WHICH WAS COMPARED TO A RESULT FROM ANOTHER METHOD: 500 U/ML. THERE IS NO INFORMATION ON THE PATIENTS CLINICAL HISTORY. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7638655 ↗
Other Single (specified) analyte controls (assayed and unassayed) devices
- CONTOUR® NEXT 7314Ascensia Diabetes Care US Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)227 reports
- FastPack® IP SHBG FastPack® IP SHBG 30 ImmunoassayQualigen Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)2 recalls0 reports
- CONTOUR® NEXT 7377Ascensia Diabetes Care US Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)95 reports
- ARCHITECT 7K62-25Abbott Ireland · SINGLE (SPECIFIED) ANALYTE CONTROLS (ASSAYED AND UNASSAYED)94 reports
- ARCHITECT 08K2812Abbott GmbH · Single (specified) analyte controls (assayed and unassayed)1 recall12 reports
- CONTOUR® NEXT 9697Ascensia Diabetes Care US Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)56 reports
- VITROS 6192257Ortho Clinical Diagnostics · Single (specified) analyte controls (assayed and unassayed)1 recall0 reports
- McKesson 5011Mckesson Medical-Surgical Inc. · Single (Specified) Analyte Controls (Assayed And Unassayed)1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.