Dexcom Receiver MT22949Recalls & reported problems

Dexcom, Inc. · Dexcom Receiver · Class 3

Quick summary

This Dexcom receiver is a device that shows glucose readings from a Dexcom sensor. Reports mostly describe wireless signal loss, frozen screens, alarms that did not sound, unexpected shutdowns, and readings that could not be obtained. There are 7,702 reports, with no deaths and 61 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
7,702
Deaths reported
0
Injuries reported
61
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7,702 adverse event reports that list this exact device (DI 00386270000323), received between Apr 22, 2017 and Nov 14, 2023 — classified as 61 injuries, 7,641 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (2,472 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem2,472
  • Defective Alarm1,511
  • Unable to Obtain Readings1,300
  • Unexpected Shutdown964
  • Device Displays Incorrect Message314
  • Display or Visual Feedback Problem298
  • No Audible Alarm294
  • Inappropriate or Unexpected Reset274

Patient problems

  • No Consequences Or Impact To Patient7,081
  • No Clinical Signs, Symptoms or Conditions584
  • Hypoglycemia48
  • Loss of consciousness33
  • Seizures9
  • Sweating8
  • Confusion/ Disorientation6
  • Fall3

Recent adverse event reports

The 25 most recent of 7,702 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE RECEIVER DISPLAY WAS FROZEN. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18132850 ↗
  • MalfunctionNov 10, 2023No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT NO AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18114413 ↗
  • MalfunctionAug 21, 2023Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17586594 ↗
  • MalfunctionAug 8, 2023Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17484786 ↗
  • MalfunctionApr 3, 2023Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED FOR THE G6 RECEIVER. HOWEVER, DATA FOR THE G6 IOS APPLICATION WAS PROVIDED FOR EVALUATION. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16670478 ↗
  • MalfunctionMar 30, 2023Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT WAS PERFORMED AND PASSED. NORDIC BLUETOOTH PAIRING TEST WAS PERFORMED AND PASSED. FUNCTIONAL TEST WAS PERFORMED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERRORS RELATED TO THE COMPLAINT. THE ALLEGATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED AS THE ALLEGATION WAS NOT FOUND. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16648542 ↗
  • MalfunctionMar 22, 2023Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED FOR THE G6 RECEIVER. HOWEVER, DATA FOR THE G6 IOS APP WAS PROVIDED FOR EVALUATION. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE THE MOBILE DEVICE LOST POWER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16594422 ↗
  • MalfunctionFeb 27, 2023Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERIOR VISUAL INSPECTION WAS PERFORMED AND PASSED. CHARGE AND BOOT TESTING WAS PERFORMED AND FAILED. INTERNAL VISUAL INSPECTION WAS PERFORMED AND FAILED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERRORS RELATED TO THE COMPLAINT. THE ALLEGATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED AS THE ALLEGATION WAS NOT FOUND. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16449688 ↗
  • MalfunctionFeb 8, 2023Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. DATA WAS RECEIVED FOR EVALUATION. HOWEVER, THE ALLEGED PRODUCT IS NOT PRESENT WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16326419 ↗
  • MalfunctionJan 17, 2023Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16189018 ↗
  • MalfunctionDec 30, 2022Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OVER ONE HOUR OCCURRED FOR THE G6 RECEIVER. HOWEVER, DATA FOR THE G6 ANDROID APPLICATION WAS PROVIDED FOR EVALUATION. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APPLICATION WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16080887 ↗
  • MalfunctionDec 28, 2022Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16065048 ↗
  • MalfunctionDec 1, 2022Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15895986 ↗
  • MalfunctionAug 23, 2022Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. DATA WAS RECEIVED FOR EVALUATION. HOWEVER, THE ALLEGED PRODUCT IS NOT PRESENT WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15280316 ↗
  • MalfunctionFeb 17, 2022Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13553839 ↗
  • MalfunctionFeb 15, 2022Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13534011 ↗
  • MalfunctionFeb 14, 2022Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13524336 ↗
  • MalfunctionFeb 10, 2022No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT THE RECEIVER HAD NO AUDIO OUTPUT. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. MANUAL FUNCTIONAL "TRY IT" TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE AUDIO SPEAKER RESISTANCE WAS MEASURED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13506775 ↗
  • MalfunctionFeb 1, 2022Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13416896 ↗
  • MalfunctionJan 28, 2022Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. DATA WAS RECEIVED FOR EVALUATION. HOWEVER, THE ALLEGED PRODUCT IS NOT PRESENT WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13385885 ↗
  • MalfunctionJan 26, 2022Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13360603 ↗
  • MalfunctionJan 26, 2022Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13360508 ↗
  • MalfunctionJan 26, 2022Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT INTERMITTENT AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13359835 ↗
  • MalfunctionJan 26, 2022Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT NO VIBRATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13359567 ↗
  • MalfunctionJan 26, 2022Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT INTERMITTENT AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13359554 ↗

Other Sensor, Glucose, Invasive, Non-Adjunctive devices

All Sensor, Glucose, Invasive, Non-Adjunctive devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.