Dexcom G5 Mobile Receiver Kit (Medicare) STK-MC-001Recalls & reported problems
Dexcom, Inc. · G5 Mobile Receiver Kit (Medicare) · Class 3
All models of Dexcom G5 (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Dexcom G5 mobile receiver shows glucose readings and alerts from a sensor. Its 494 reports most often describe alarms that did not work, wireless connection losses and unexpected resets. Most patients had no consequences; the reports include 2 injuries and no deaths. One closed recall from July 2019 covers receivers that could get stuck on the startup screen, leaving patients without glucose readings or alerts.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 494
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 494 adverse event reports that list this exact device (DI 00386270000606), received between May 10, 2017 and Jul 6, 2018 — classified as 2 injuries, 492 malfunctions.
The most frequently reported specific device problem is “Defective Alarm” (177 reports).
1 FDA recall record name this device's identifier.
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 12, 2019Class IITerminated
It has been reported that use of the mobile receiver with software version SW10617 rev 4.0.1.048 have reported: 1. the receiver becoming stuck on initialization screen when powering on. This will cause patients not to be able to receive glucose values or alerts; 2. Reinitialization of the receiver without user interaction. If this occurs, the receiver will either reboot and operate normally or require the user to press the Select button on the receiver in order to resume normal operation. In this situation, the user is provided an audio and vibratory alert every 5 minutes. Until the select button is pressed, patients will not receive glucose values or alerts.
Z-0929-2020 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Defective Alarm177
- Wireless Communication Problem107
- Inappropriate or Unexpected Reset62
- Device Inoperable56
- Device Displays Incorrect Message51
- Display or Visual Feedback Problem10
- Device Operates Differently Than Expected9
- No Audible Alarm8
Patient problems
- No Consequences Or Impact To Patient485
- Hypoglycemia8
- Loss of consciousness3
- Confusion/ Disorientation2
- No Known Impact Or Consequence To Patient1
- Sweating1
Recent adverse event reports
The 25 most recent of 494 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 6, 2018Unexpected Shutdown; Unexpected Shutdown
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER CEASING TO FUNCTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7667182 ↗MalfunctionJun 4, 2018Low Audible Alarm; Low Audible Alarm
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER HAD LOW AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF LOW AUDIO OUTPUT COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7567778 ↗MalfunctionMay 10, 2018Imprecision; Imprecision
DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, THERE WAS AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER. THE SENSOR WAS INSERTED AT THE ABDOMEN ON (B)(6) 2018. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE REPORTED INACCURACY COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7505423 ↗MalfunctionApr 24, 2018Device Inoperable; Device Inoperable
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THERE WAS AN UNEXPECTED RECEIVER SHUT-DOWN. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. CONFIRMATION OF THE PROBLEM AND ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7456890 ↗MalfunctionMar 13, 2018Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, PATIENT EXPERIENCED A LOSS OF CONNECTION. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. CONFIRMATION OF THE PROBLEM AND ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7334490 ↗MalfunctionFeb 6, 2018Imprecision; Use of Device Problem; Imprecision; Use of Device Problem
THE COMPLAINT DEVICE WAS RECEIVED FOR EVALUATION AND PASSED EXTERNAL VISUAL INSPECTION. THE RECEIVER SUCCESSFULLY BOOTED AND CHARGED. THE RECEIVER DOWNLOAD LOG WAS REVIEWED AND THERE WERE NO ISSUES RELATED TO THE CUSTOMER COMPLAINT. ALL ESTIMATED CGM VALUES SHOWN ON THE RECEIVER ARE WITHIN THE ACCURATE VALUE RANGE. THE DEVICE PASSED GLOBAL RECEIVER FUNCTIONAL TESTING. THE REPORTED ALLEGATION WAS NOT FOUND AND THE CUSTOMER COMPLAINT WAS NOT CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7246558 ↗MalfunctionJan 19, 2018Device Inoperable; Device Inoperable
DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER CEASING TO FUNCTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7208245 ↗MalfunctionJan 17, 2018Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017, PATIENT EXPERIENCED A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND RECEIVER. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO DATA OR PRODUCT WAS PROVIDED FOR INVESTIGATION. CONFIRMATION OF THE COMPLAINT WAS UNDETERMINED. THE ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7201860 ↗MalfunctionJan 17, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR CALL TECH SUPPORT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. THE RECEIVER WAS VISUALLY INSPECTED AND "CALL TECH SUPPORT" ERROR WAS OBSERVED ON THE SCREEN. THE DATA LOG COULD NOT BE RETRIEVED AND FUNCTIONAL TESTING COULD NOT BE PERFORMED BECAUSE OF THE "CALL TECH SUPPORT" ERROR. CUSTOMER COMPLAINT CONFIRMED BECAUSE OF THE OCCURRENCE OF THE "CALL TECH SUPPORT" ERROR. THE ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7200272 ↗MalfunctionJan 15, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBAT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7190077 ↗MalfunctionJan 15, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBAT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7189901 ↗MalfunctionJan 12, 2018Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND RECEIVER OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A LOSS OF CONNECTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7188667 ↗MalfunctionJan 12, 2018Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND RECEIVER OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A LOSS OF CONNECTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7188656 ↗MalfunctionJan 12, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBAT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7188147 ↗MalfunctionJan 12, 2018Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017, PATIENT EXPERIENCED A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND RECEIVER. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A LOSS OF CONNCECTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7187006 ↗MalfunctionJan 12, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWBAT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7185922 ↗MalfunctionJan 12, 2018Defective Alarm; Defective Alarm
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECTS WERE FOUND. THE RECEIVER WILL BOOT AND CHARGE. THE RECEIVER DOWNLOAD WAS RETRIEVED AND PASSED. THE RECEIVER PASSED ALL RELEVANT TESTS ON FUNCTIONAL TEST STATION. PASSED 'TRY IT','FIXED LOW' TEST AUDIO TONE WAS HEARD (WITHOUT INTERMITTENCE) AT NORMAL VOLUME WHEN TESTED WITH RECEIVER COMMUNICATION TOOL : 7.0.0.75(LOW,MEDIUM,HIGH).OPENED RECEIVER,PASSED INTERNAL VISUAL INSPECTION. SPEAKER MEASURED 7.5 OHMS.(WITHIN TOLERANCE)PASSED 'WIGGLE' TEST. THE REPORTED EVENT OF AN INTERMITTENT AUDIO OUTPUT WAS NOT CONFIRMED. THE ROOT CAUSE COULD NOT BE DETERMINED
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7185708 ↗MalfunctionJan 11, 2018Defective Alarm; Defective Alarm
THE COMPLAINT DEVICE WAS RECEIVED FOR EVALUATION AND PASSED EXTERNAL VISUAL INSPECTION. THE RECEIVER SUCCESSFULLY BOOTED AND CHARGED. THE RECEIVER DOWNLOAD LOG WAS REVIEWED AND THERE WERE NO ISSUES RELATED TO THE CUSTOMER COMPLAINT. AN AUDIO/VIBRATION TEST WAS PERFORMED AND THE DEVICE PASSED. THE SPEAKER RESISTANCE WAS MEASURED AND RESISTANCE WAS WITHIN SPECIFICATION. THE REPORTED ALLEGATION WAS NOT FOUND. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7184580 ↗MalfunctionJan 11, 2018Defective Alarm; Defective Alarm
DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017 THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7184571 ↗MalfunctionJan 11, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR HWRFU. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE RECEIVER DATA LOG WAS DOWNLOADED AND REVIEWED AND HARDWARE ERRORS WERE OBSERVED. THE REPORTED EVENT OF AN ERROR ICON DISPLAY WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7184315 ↗MalfunctionJan 11, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR CALL TECH SUPPORT. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE RECEIVER DATA LOG WAS DOWNLOADED AND REVIEWED AND HARDWARE ERRORS WERE OBSERVED. THE REPORTED EVENT OF AN ERROR ICON DISPLAY WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7184312 ↗MalfunctionJan 11, 2018Wireless Communication Problem; Wireless Communication Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) /2017, PATIENT EXPERIENCED A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND RECEIVER. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A LOSS OF CONNECTION COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7184172 ↗MalfunctionJan 11, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DEXCOM WAS MADE AWARE ON (B)(4) 2017, THAT ON (B)(6) 2017, THE RECEIVER DISPLAYED ERR CALL TECH SUPPORT . NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS RETURNED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7184096 ↗MalfunctionJan 11, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. THE RECEIVER WILL NOT CHARGE AND BOOT.(CALL TECH SUPPORT ALARM ON DISPLAY). THE RECEIVER LOG WAS REVIEWED AND ERRORS WERE FOUND IN THE LOG..THE REPORTED EVENT OF AN ERROR ICON DISPLAY WAS CONFIRMED DURING LOG REVIEW. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7184091 ↗MalfunctionJan 11, 2018Defective Alarm; Defective Alarm
DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017 THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A ROOT CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7183997 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
Other Sensor, Glucose, Invasive, Non-Adjunctive devices
- Dexcom G5 Mobile Transmitter Kit (Medicare) STT-MC-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive18,422 reports
- Dexcom Receiver MT22949Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive7,702 reports
- Dexcom G5 Mobile Replacement Receiver Kit (Medicare) STR-MC-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive4,734 reports
- Dexcom G5 Mobile Android App SW10940Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive3,843 reports
- Dexcom Receiver Kit (Medicare) STK-MD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive963 reports
- Dexcom Receiver Kit STK-SD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive145 reports
- Dexcom Replacement Receiver Kit STR-SD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive33 reports
- Dexcom Replacement Receiver Kit (Medicare) STR-MD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page