Dexcom Receiver Kit (Medicare) STK-MD-001Recalls & reported problems

Dexcom, Inc. · Dexcom Receiver Kit (Medicare) · Class 3

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom Receiver Kit shows glucose readings from a Dexcom sensor. Many of its 963 reports describe wireless connection problems, failure to get readings and unexpected shutdowns, and most note no impact on the patient. The reports mention no deaths and 5 injuries. No recalls name this device, and only 4 reports came in the last 12 months compared with 15 in the previous 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
963
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 963 adverse event reports that list this exact device (DI 00386270000859), received between May 14, 2019 and Apr 22, 2026 — classified as 5 injuries, 958 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (285 reports).

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 5 reportsApr 2024: 5 reportsMay 2024: 3 reportsJun 2024: 5 reportsJul 2024: 3 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem285
  • Unable to Obtain Readings213
  • Unexpected Shutdown189
  • Defective Alarm81
  • Display or Visual Feedback Problem71
  • Inappropriate or Unexpected Reset59
  • No Audible Alarm40
  • Low Audible Alarm11

Patient problems

  • No Consequences Or Impact To Patient587
  • No Clinical Signs, Symptoms or Conditions386
  • Hypoglycemia3
  • Loss of consciousness3
  • Coma1
  • Elevated Ketones/ Diabetic Ketoacidosis1
  • Hypothermia1
  • Nausea1

Recent adverse event reports

The 25 most recent of 963 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956046 ↗
  • MalfunctionDec 1, 2025No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23684077 ↗
  • MalfunctionNov 10, 2025Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAY OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23511718 ↗
  • MalfunctionNov 4, 2025No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23466296 ↗
  • MalfunctionAug 20, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22854887 ↗
  • MalfunctionAug 6, 2025Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    THE COMPLAINT STATES THAT "RECEIVER REINITIALIZATION" WAS REPORTED. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. PROBLEM COULD NOT BE CONFIRMED AND PROBABLE CAUSE CANNOT BE DETERMINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22731745 ↗
  • MalfunctionJul 30, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22672005 ↗
  • MalfunctionJun 13, 2025Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAY OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22202501 ↗
  • MalfunctionJun 5, 2025Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22151838 ↗
  • MalfunctionMay 5, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21963549 ↗
  • MalfunctionApr 2, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21755763 ↗
  • MalfunctionApr 2, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21751947 ↗
  • MalfunctionMar 24, 2025Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET THE CRITERIA OF A REPORTABLE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21681780 ↗
  • MalfunctionMar 20, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21657598 ↗
  • MalfunctionJan 28, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT A DISPLAY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21257915 ↗
  • MalfunctionJan 16, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT A DISPLAY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21172256 ↗
  • MalfunctionJan 10, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21132076 ↗
  • MalfunctionDec 30, 2024Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAY OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21044871 ↗
  • MalfunctionDec 13, 2024Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20929564 ↗
  • MalfunctionOct 24, 2024Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20526137 ↗
  • MalfunctionOct 16, 2024No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20462762 ↗
  • MalfunctionAug 27, 2024Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20081294 ↗
  • MalfunctionAug 14, 2024Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19984323 ↗
  • MalfunctionAug 13, 2024No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19977646 ↗
  • MalfunctionJul 25, 2024Unable to Obtain Readings; Unable to Obtain Readings

    PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19830699 ↗

FDA clearance and classification

Its record lists premarket approval application P120005, decided on Oct 5, 2012 with FDA decision code APWD (not a standard approval code). Check the FDA record for its current status. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: A non-adjunctive invasive glucose sensor is intended to determine glucose levels, and the direction and rate of change of glucose levels in people with diabetes. The device is indicated to replace information obtained from standard blood glucose monitoring devices to make diabetes-related treatment decisions. The device also provides historical glucose information, facilitating long-term therapy adjustments.

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a

Common questions about the Dexcom Receiver Kit (Medicare) STK-MD-001

Has the Dexcom Receiver Kit (Medicare) STK-MD-001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00386270000859), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Dexcom Receiver Kit (Medicare) STK-MD-001?

The FDA's MAUDE database lists 963 adverse event reports for this device, received between May 14, 2019 and Apr 22, 2026. In those reports, the most common specific device problems are “Wireless Communication Problem” (285), “Unable to Obtain Readings” (213) and “Unexpected Shutdown” (189); the most common patient problems are “Hypoglycemia” (3), “Loss of consciousness” (3) and “Coma” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dexcom Receiver Kit (Medicare) STK-MD-001?

None of the reports are classified as deaths, 5 as injuries and 958 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Dexcom Receiver Kit (Medicare) STK-MD-001, and is it still sold?

The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom Receiver Kit (Medicare), model STK-MD-001). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.