Dexcom Replacement Receiver Kit (Medicare) STR-MD-001Recalls & reported problems
Dexcom, Inc. · Replacement Receiver Kit (Medicare) · Class 3
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Dexcom replacement receiver is a handheld display for glucose sensor readings. Reports mostly describe the receiver failing to show readings, alarms not sounding as expected, display problems, and unexpected shutdowns or resets. There are 27 reports naming its identifier, with no deaths and 1 injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 00386270000842), received between Jul 29, 2022 and Jul 2, 2025 — classified as 1 injury, 26 malfunctions.
The most frequently reported specific device problem is “Unable to Obtain Readings” (10 reports).
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Unable to Obtain Readings10
- Defective Alarm4
- No Audible Alarm4
- Display or Visual Feedback Problem3
- Unexpected Shutdown3
- Inappropriate or Unexpected Reset2
- Wireless Communication Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions27
Recent adverse event reports
The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 2, 2025Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A RECEIVER REINITIALIZATION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22382858 ↗MalfunctionJan 30, 2025Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT AN UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21274087 ↗MalfunctionSep 13, 2024No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT AN ALERT/NOTIFICATION SETTINGS ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20213920 ↗MalfunctionJul 25, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT A DISPLAY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19834146 ↗MalfunctionNov 10, 2023Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18115228 ↗MalfunctionAug 14, 2023Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17542765 ↗MalfunctionJul 6, 2023Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17268161 ↗InjuryMay 18, 2023No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT INTERMITTENT AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16952835 ↗MalfunctionMay 15, 2023Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16930093 ↗MalfunctionApr 28, 2023Defective Alarm; Defective Alarm
THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND PASSED. A MANUAL VIBRATION TEST WAS PERFORMED AND PASSED. THE VIBRATOR MOTOR RESISTANCE WAS MEASURED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16832475 ↗MalfunctionApr 28, 2023No Audible Alarm; No Audible Alarm
SUBSEQUENT TO THE INITIAL MDR, A CORRECTION IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16832466 ↗MalfunctionApr 20, 2023Defective Alarm; Defective Alarm
IT WAS REPORTED THAT INTERMITTENT AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16777347 ↗MalfunctionApr 5, 2023Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16684763 ↗MalfunctionMar 14, 2023Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
IT WAS REPORTED THAT THE RECEIVER INITIALIZED WITHOUT A MANUAL RESTART OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16538116 ↗MalfunctionDec 29, 2022Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16074423 ↗MalfunctionNov 22, 2022Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15843683 ↗MalfunctionNov 21, 2022No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT THE RECEIVER HAD NO AUDIO OUTPUT. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15840670 ↗MalfunctionOct 31, 2022Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15700540 ↗MalfunctionOct 25, 2022Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15668021 ↗MalfunctionOct 12, 2022Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15591437 ↗MalfunctionOct 11, 2022Defective Alarm; Defective Alarm
IT WAS REPORTED THAT INTERMITTENT AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15580033 ↗MalfunctionOct 1, 2022Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15525297 ↗MalfunctionSep 27, 2022Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15492059 ↗MalfunctionAug 26, 2022Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE RECEIVER DISPLAY WAS FROZEN. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15303708 ↗MalfunctionAug 24, 2022Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15287282 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
Other Sensor, Glucose, Invasive, Non-Adjunctive devices
- Dexcom G5 Mobile Transmitter Kit (Medicare) STT-MC-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive18,422 reports
- Dexcom Receiver MT22949Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive7,702 reports
- Dexcom G5 Mobile Replacement Receiver Kit (Medicare) STR-MC-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive4,734 reports
- Dexcom G5 Mobile Android App SW10940Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive3,843 reports
- Dexcom Receiver Kit (Medicare) STK-MD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive963 reports
- Dexcom G5 Mobile Receiver Kit (Medicare) STK-MC-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive1 recall494 reports
- Dexcom Receiver Kit STK-SD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive145 reports
- Dexcom Replacement Receiver Kit STR-SD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive33 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page