This Dexcom G5 Mobile transmitter kit has 18,422 adverse event reports in the FDA database, including 19 injuries and no deaths. Reports mostly describe wireless communication problems, transmitters that give no output or show failure errors, and low battery alerts, and most say the patient was not affected. No recalls name this device. The most recent report is from June 2021.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 18,422
- Deaths reported
- 0
- Injuries reported
- 19
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18,422 adverse event reports that list this exact device (DI 00386270000620), received between May 26, 2017 and Jun 9, 2021 — classified as 19 injuries, 18,403 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (8,409 reports).
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Wireless Communication Problem8,409
- No Device Output6,289
- Device Displays Incorrect Message1,838
- Premature End-of-Life Indicator1,400
- Appropriate Device Problem Term/Code Not Available139
- Imprecision109
- Connection Problem102
- Inappropriate or Unexpected Reset69
Patient problems
- No Consequences Or Impact To Patient18,391
- No Clinical Signs, Symptoms or Conditions33
- Hypoglycemia9
- Loss of consciousness7
- Confusion/ Disorientation2
- Elevated Ketones/ Diabetic Ketoacidosis2
- Fatigue2
- Hyperglycemia2
Recent adverse event reports
The 25 most recent of 18,422 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT THE TRANSMITTER UNPAIRED ITSELF. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11968917 ↗MalfunctionMay 4, 2021No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11770026 ↗MalfunctionApr 5, 2021No Device Output; No Device Output
IT WAS REPORTED THAT A TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11615956 ↗MalfunctionFeb 24, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT A PAIRING FAILURE OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE CHECK WAS PERFORMED AND FAILED. CLOUD/SHARE DATA REVIEW: TRANSMITTER, PAIRING FAILURE WAS OBSERVED IN THE CLOUD DATA. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A FAILED TRANSMITTER ERROR. IN ADDITION, TRANSMITTER FAILED ERROR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE REPORTED EVENT OF A PAIRING FAILURE IS REPORTABLE BASED ON THE FINDING OF A TRANSMITTER FAILED ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11376795 ↗MalfunctionFeb 18, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT A LOW TRANSMITTER BATTERY ALERT OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0 VDC. A REVIEW OF THE AVAILABLE SHARE LOGS WAS PERFORMED AND A LOW TRANSMITTER BATTERY ALERT WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE LOW TRANSMITTER BATTERY VOLTAGE. IN ADDITION, A FAILED TRANSMITTER ERROR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE REPORTED ISSUE OF A LOW TRANSMITTER BATTERY ALERT IS REPORTABLE BASED ON THE FINDING OF A FAILED TRANSMITTER ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11344283 ↗MalfunctionFeb 11, 2021No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11314953 ↗MalfunctionFeb 11, 2021No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11314781 ↗MalfunctionFeb 9, 2021Premature End-of-Life Indicator; Premature End-of-Life Indicator
IT WAS REPORTED THAT A LOW TRANSMITTER BATTERY ALERT OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0 VDC. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND A LOW TRANSMITTER BATTERY ALERT WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE LOW TRANSMITTER BATTERY VOLTAGE. IN ADDITION, A FAILED TRANSMITTER ERROR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE REPORTED ISSUE OF A LOW TRANSMITTER BATTERY ALERT IS REPORTABLE BASED ON THE FINDING OF A FAILED TRANSMITTER ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11305370 ↗MalfunctionFeb 4, 2021Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11281085 ↗MalfunctionJan 29, 2021No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11255047 ↗MalfunctionJan 14, 2021No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11170304 ↗MalfunctionJan 8, 2021No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11143047 ↗MalfunctionDec 21, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11047380 ↗MalfunctionDec 18, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11037774 ↗MalfunctionDec 10, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A "PAIR NEW TRANSMITTER" MESSAGE OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED AS A "PAIR NEW TRANSMITTER" MESSAGE WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A TRANSMITTER FAILED ERROR. THE REPORTED EVENT OF A "PAIR NEW TRANSMITTER" MESSAGE IS REPORTABLE BASED ON THE FINDING OF A TRANSMITTER FAILED ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10983952 ↗MalfunctionDec 9, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10978914 ↗MalfunctionNov 26, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10908867 ↗MalfunctionNov 18, 2020No Device Output; No Device Output
SUBSEQUENT TO THE INITIAL MDR, A CORRECTION IS REQUIRED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10860311 ↗MalfunctionNov 11, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10823494 ↗MalfunctionNov 6, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0V. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10804709 ↗MalfunctionOct 23, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO ESTABLISH A CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10729968 ↗MalfunctionOct 23, 2020No Device Output; No Device Output
THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0V. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10727424 ↗MalfunctionOct 19, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10702076 ↗MalfunctionOct 15, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. TRANSMITTER VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0 VOLTAGE. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10686404 ↗MalfunctionOct 3, 2020No Device Output; No Device Output
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10624266 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
Other Sensor, Glucose, Invasive, Non-Adjunctive devices
- Dexcom Receiver MT22949Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive7,702 reports
- Dexcom G5 Mobile Replacement Receiver Kit (Medicare) STR-MC-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive4,734 reports
- Dexcom G5 Mobile Android App SW10940Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive3,843 reports
- Dexcom Receiver Kit (Medicare) STK-MD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive963 reports
- Dexcom G5 Mobile Receiver Kit (Medicare) STK-MC-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive1 recall494 reports
- Dexcom Receiver Kit STK-SD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive145 reports
- Dexcom Replacement Receiver Kit STR-SD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive33 reports
- Dexcom Replacement Receiver Kit (Medicare) STR-MD-001Dexcom, Inc. · Sensor, Glucose, Invasive, Non-Adjunctive27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page