Dexcom G6 Transmitter 9445-02Recalls & reported problems
Dexcom, Inc. · G6 Transmitter · Class 2
All models of Dexcom G6 (13) · Dexcom G6 Transmitter (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
The 229,491 reports for this Dexcom G6 transmitter mostly describe no device output or lost wireless signal, and early battery end-of-life warnings. Most list no impact on the patient; 109 injuries and 3 deaths were reported. Many recent reports note that no product was returned, so a cause could not be determined. No recalls name this transmitter.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 229,489
- Deaths reported
- 3
- Injuries reported
- 109
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 229,489 adverse event reports that list this exact device (DI 00386270000385), received between Apr 2, 2018 and Sep 22, 2026 — classified as 3 deaths, 109 injuries, 229,377 malfunctions.
The most frequently reported specific device problem is “No Device Output” (117,073 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- No Device Output117,073
- Wireless Communication Problem91,976
- Premature End-of-Life Indicator10,146
- Appropriate Device Problem Term/Code Not Available4,318
- Device Displays Incorrect Message2,840
- Connection Problem2,511
- Inappropriate or Unexpected Reset193
- Imprecision163
Patient problems
- No Consequences Or Impact To Patient149,850
- No Clinical Signs, Symptoms or Conditions83,556
- Hypoglycemia33
- Loss of consciousness20
- No Known Impact Or Consequence To Patient20
- Insufficient Information13
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available12
- Elevated Ketones/ Diabetic Ketoacidosis11
Recent adverse event reports
The 25 most recent of 229,489 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50862281 ↗MalfunctionSep 17, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50810370 ↗MalfunctionSep 10, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50724578 ↗MalfunctionAug 13, 2026No Device Output
It was reported that a transmitter battery issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50234354 ↗MalfunctionAug 11, 2026No Device Output
It was reported that a transmitter battery issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50208906 ↗MalfunctionJul 29, 2026Wireless Communication Problem
It was reported that a signal loss occurred frequently between (b)(6) 2026. Product was provided for investigation. An external visual inspection was performed and passed. Voltage Test was performed and passed. Nordic Bluetooth Pairing Test was performed and passed. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour frequently within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50069821 ↗MalfunctionJul 27, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50021678 ↗MalfunctionJul 27, 2026Wireless Communication Problem
It was reported that a signal loss occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50013787 ↗MalfunctionJul 1, 2026Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. IT WAS INDICATED THAT THE PATIENT USED AN EXPIRED PRODUCT, WHICH IS MISUSE OF THE DEVICE. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO COMPLETE A DATA TRANSFER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25685788 ↗MalfunctionJun 29, 2026Wireless Communication Problem
IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25661547 ↗MalfunctionJun 23, 2026Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25598155 ↗MalfunctionJun 23, 2026No Device Output
IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25597783 ↗MalfunctionJun 15, 2026Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. IT WAS INDICATED THAT THE PATIENT WAS USING AN UNSUPPORTED OPERATING SYSTEM, WHICH IS MISUSE OF THE DEVICE. IT WAS INDICATED THAT THE PATIENT USED AN EXPIRED PRODUCT, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25504657 ↗MalfunctionJun 10, 2026Wireless Communication Problem
IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25462402 ↗MalfunctionJun 8, 2026No Device Output
IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25431193 ↗MalfunctionJun 1, 2026Low Readings
IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE UPPER BUTTOCKS, WHICH IS OFF-LABEL USAGE OF THE DEVICE, ON (B)(6) 2026. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED VIA DATA. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. THE DATA INVESTIGATION FOUND A DIFFERENCE IN VALUES THAT FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Hyperglycemia
MAUDE report 25351981 ↗MalfunctionMay 14, 2026Wireless Communication Problem
IT WAS REPORTED THAT"SIGNAL LOSS" OCCURRED. IT WAS INDICATED THAT THE PATIENT USED AN EXPIRED PRODUCT, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR INVESTIGATION. THE ALLEGATION AND THE PROBABLE CAUSE CANNOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25167111 ↗MalfunctionMay 12, 2026Wireless Communication Problem
IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. IT WAS INDICATED THAT THE PATIENT USED AN EXPIRED PRODUCT, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25151309 ↗MalfunctionMay 5, 2026Wireless Communication Problem
IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25089000 ↗MalfunctionMay 1, 2026Wireless Communication Problem
IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25062397 ↗MalfunctionApr 25, 2026Wireless Communication Problem
SUBSEQUENT TO THE INITIAL MDR, A CORRECTION IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24991288 ↗MalfunctionApr 2, 2026Wireless Communication Problem
IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24765863 ↗MalfunctionMar 13, 2026No Device Output
IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24598502 ↗MalfunctionMar 13, 2026Wireless Communication Problem
IT WAS REPORTED THAT A PAIRING ISSUE OCCURRED. IT WAS INDICATED THAT THE OPERATING SYSTEM FOR THE SMART DEVICE WAS NOT A SUPPORTED SYSTEM, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. A PAIRING ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24594677 ↗MalfunctionMar 4, 2026No Device Output
IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24505620 ↗
FDA clearance and classification
Granted through the De Novo pathway (DEN170088) on Mar 27, 2018. Submitted by Dexcom, Inc. as “Dexcom G6 Continuous Glucose Monitoring System”. The FDA review took 109 days. FDA record ↗
The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗
FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Common questions about the Dexcom G6 Transmitter 9445-02
Has the Dexcom G6 Transmitter 9445-02 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00386270000385), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Dexcom G6 Transmitter 9445-02?
The FDA's MAUDE database lists 229,489 adverse event reports for this device, received between Apr 2, 2018 and Sep 22, 2026. In those reports, the most common specific device problems are “No Device Output” (117,073), “Wireless Communication Problem” (91,976) and “Premature End-of-Life Indicator” (10,146); the most common patient problems are “Hypoglycemia” (33), “Loss of consciousness” (20) and “Elevated Ketones/ Diabetic Ketoacidosis” (11). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Dexcom G6 Transmitter 9445-02?
3 reports are classified as deaths, 109 as injuries and 229,377 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Dexcom G6 Transmitter 9445-02 FDA approved?
It was granted through the De Novo pathway (DEN170088) on Mar 27, 2018.
Who makes the Dexcom G6 Transmitter 9445-02, and is it still sold?
The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G6 Transmitter, model 9445-02). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices
- Dexcom G6 Transmitter 9445-24Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated240,475 reports
- Dexcom G6 Sensor Applicator Assembly 9500-45Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall4 deaths114,621 reports
- US G6 Transmitter (Retail, Receiver Optional) STT-OE-002Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated76,498 reports
- Dexcom G7 GSS (161) 9500-161Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated70,146 reports
- Dexcom G7 iOS App SW12300Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 recalls1 death54,855 reports
- Dexcom G7 GSS (161) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall5 deaths47,762 reports
- G6 Medicare Transmitter STT-OM-001Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated38,212 reports
- Dexcom G7 GSS (160) 9500-160Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated33,025 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page