G6 Medicare Transmitter STT-OM-001Recalls & reported problems
Dexcom, Inc. · G6 Medicare Transmitter Kit, Dexcom · Class 2
Quick summary
This is a Dexcom G6 Medicare transmitter. Its reports mostly describe lost wireless signal and no output from the device. There are 38,236 reports, with no deaths and 71 injuries. No recalls name this device. Reports dropped to 2,595 in the last 12 months, from 6,797 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 38,339
- Deaths reported
- 0
- Injuries reported
- 70
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 38,339 adverse event reports that list this exact device (DI 00386270001788), received between Mar 11, 2022 and Sep 30, 2026 — classified as 70 injuries, 38,268 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (27,397 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Wireless Communication Problem27,397
- No Device Output10,275
- Connection Problem270
- Appropriate Device Problem Term/Code Not Available164
- Premature End-of-Life Indicator147
- Low Readings39
- Alarm Not Visible34
- High Readings22
Patient problems
- No Clinical Signs, Symptoms or Conditions38,200
- Hyperglycemia68
- Hypoglycemia55
- Loss of consciousness22
- Elevated Ketones/ Diabetic Ketoacidosis12
- Vomiting12
- Fatigue10
- Malaise10
Recent adverse event reports
The 25 most recent of 38,339 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954336 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952500 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951423 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951103 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over an hour was found in connection to the reported allegation. The probable cause of the signal loss was determined to be the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950698 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950236 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. It was indicated that the patient used an expired product, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over an hour was found in connection to the reported allegation. The probable cause of the signal loss was determined to be the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949867 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that "Signal Loss" occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948793 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a signal loss occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947652 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945805 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942120 ↗MalfunctionSep 29, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938118 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937861 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937588 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that signal loss occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. It was indicated that the patient used an expired product, which is misuse of the device. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936548 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that"Signal Loss" occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934932 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that signal loss occurred. It was indicated that the patient used an expired product, which is misuse of the device. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934577 ↗MalfunctionSep 29, 2026No Device Output
It was reported that signal loss occurred. Performance data was reviewed. Signal loss was not found within the investigation window. The allegation was not confirmed. However, a transmitter failure was found in connection to the reported allegation. The probable cause of the transmitter failure could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927210 ↗MalfunctionSep 28, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923863 ↗MalfunctionSep 28, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923315 ↗MalfunctionSep 28, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50922113 ↗MalfunctionSep 28, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917431 ↗MalfunctionSep 28, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50913515 ↗MalfunctionSep 25, 2026No Device Output
It was reported that a transmitter battery issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50903970 ↗MalfunctionSep 25, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50901828 ↗
Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices
- Dexcom G6 Transmitter 9445-24Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated240,475 reports
- Dexcom G6 Transmitter 9445-02Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 deaths229,493 reports
- Dexcom G6 Sensor Applicator Assembly 9500-45Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall4 deaths114,678 reports
- US G6 Transmitter (Retail, Receiver Optional) STT-OE-002Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated76,977 reports
- Dexcom G7 GSS (161) 9500-161Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated70,146 reports
- Dexcom G7 iOS App SW12300Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 recalls1 death55,219 reports
- Dexcom G7 GSS (161) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall5 deaths48,180 reports
- FreeStyle Libre 3 Plus 78768Abbott Diabetes Care Inc · Integrated Continuous Glucose Monitoring system, factory calibrated1 death33,035 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.