G6 Medicare Transmitter STT-OM-001Recalls & reported problems
Dexcom, Inc. · G6 Medicare Transmitter Kit, Dexcom · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is a Dexcom G6 Medicare transmitter. Its reports mostly describe lost wireless signal and no output from the device. There are 38,236 reports, with no deaths and 71 injuries. No recalls name this device. Reports dropped to 2,595 in the last 12 months, from 6,797 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 38,212
- Deaths reported
- 0
- Injuries reported
- 70
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 38,212 adverse event reports that list this exact device (DI 00386270001788), received between Mar 11, 2022 and Sep 30, 2026 — classified as 70 injuries, 38,141 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (27,323 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Wireless Communication Problem27,323
- No Device Output10,227
- Connection Problem270
- Appropriate Device Problem Term/Code Not Available164
- Premature End-of-Life Indicator147
- Alarm Not Visible34
- Low Readings33
- High Readings22
Patient problems
- No Clinical Signs, Symptoms or Conditions38,079
- Hyperglycemia62
- Hypoglycemia55
- Loss of consciousness22
- Elevated Ketones/ Diabetic Ketoacidosis12
- Vomiting12
- Fatigue10
- Malaise10
Recent adverse event reports
The 25 most recent of 38,212 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954336 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952500 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951423 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951103 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over an hour was found in connection to the reported allegation. The probable cause of the signal loss was determined to be the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950698 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that signal loss occurred. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950236 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. It was indicated that the patient used an expired product, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over an hour was found in connection to the reported allegation. The probable cause of the signal loss was determined to be the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949867 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that "Signal Loss" occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948793 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a signal loss occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947652 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945805 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942120 ↗MalfunctionSep 29, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938118 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937861 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937588 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that signal loss occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. It was indicated that the patient used an expired product, which is misuse of the device. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936548 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that\"Signal Loss\" occurred. It was indicated that the patient used an expired product, which is misuse of the device. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934932 ↗MalfunctionSep 29, 2026Wireless Communication Problem
It was reported that signal loss occurred. It was indicated that the patient used an expired product, which is misuse of the device. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934577 ↗MalfunctionSep 29, 2026No Device Output
It was reported that signal loss occurred. Performance data was reviewed. Signal loss was not found within the investigation window. The allegation was not confirmed. However, a transmitter failure was found in connection to the reported allegation. The probable cause of the transmitter failure could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927210 ↗MalfunctionSep 28, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923863 ↗MalfunctionSep 28, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923315 ↗MalfunctionSep 28, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50922113 ↗MalfunctionSep 28, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917431 ↗MalfunctionSep 28, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50913515 ↗MalfunctionSep 25, 2026No Device Output
It was reported that a transmitter battery issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50903970 ↗MalfunctionSep 25, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50901828 ↗
FDA clearance and classification
The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗
Devices of this type normally need 510(k) clearance before sale.
FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Common questions about the G6 Medicare Transmitter STT-OM-001
Has the G6 Medicare Transmitter STT-OM-001 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00386270001788), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the G6 Medicare Transmitter STT-OM-001?
The FDA's MAUDE database lists 38,212 adverse event reports for this device, received between Mar 11, 2022 and Sep 30, 2026. In those reports, the most common specific device problems are “Wireless Communication Problem” (27,323), “No Device Output” (10,227) and “Connection Problem” (270); the most common patient problems are “Hyperglycemia” (62), “Hypoglycemia” (55) and “Loss of consciousness” (22). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the G6 Medicare Transmitter STT-OM-001?
None of the reports are classified as deaths, 70 as injuries and 38,141 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the G6 Medicare Transmitter STT-OM-001 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the G6 Medicare Transmitter STT-OM-001, and is it still sold?
The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand G6 Medicare Transmitter, model STT-OM-001). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices
- Dexcom G6 Transmitter 9445-24Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated240,475 reports
- Dexcom G6 Transmitter 9445-02Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 deaths229,489 reports
- Dexcom G6 Sensor Applicator Assembly 9500-45Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall4 deaths114,621 reports
- US G6 Transmitter (Retail, Receiver Optional) STT-OE-002Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated76,498 reports
- Dexcom G7 GSS (161) 9500-161Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated70,146 reports
- Dexcom G7 iOS App SW12300Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 recalls1 death54,855 reports
- Dexcom G7 GSS (161) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall5 deaths47,762 reports
- Dexcom G7 GSS (160) 9500-160Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated33,025 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page