Dexcom G6 Transmitter 9445-24Recalls & reported problems

Dexcom, Inc. · G6 Nuevo Transmitter · Class 2

All models of Dexcom G6 (13) · Dexcom G6 Transmitter (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

Reports for this Dexcom G6 transmitter almost all describe lost wireless signal (174,479 reports) or no output from the device (62,456), often linked to the transmitter battery. Most reports list no symptoms; 86 describe injuries, and low and high blood sugar are the most listed patient problems. Reporting has nearly stopped, with 1 report in the last 12 months. No recalls name this transmitter.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
240,475
Deaths reported
0
Injuries reported
86
Recalls naming it
0

Summary

The FDA's MAUDE database contains 240,475 adverse event reports that list this exact device (DI 00386270001627), received between Dec 16, 2020 and Jan 28, 2026 — classified as 86 injuries, 240,389 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (174,479 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 5,382 reportsJan 2024: 5,893 reportsFeb 2024: 5,066 reportsMar 2024: 3,464 reportsApr 2024: 2,964 reportsMay 2024: 1,866 reportsJun 2024: 2,014 reportsJul 2024: 1,851 reportsAug 2024: 1,317 reportsSep 2024: 1,293 reportsOct 2024: 1,106 reportsNov 2024: 724 reportsDec 2024: 491 reportsJan 2025: 516 reportsFeb 2025: 242 reportsMar 2025: 239 reportsApr 2025: 115 reportsMay 2025: 172 reportsJun 2025: 184 reportsJul 2025: 192 reportsAug 2025: 157 reportsSep 2025: 94 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265,893
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem174,479
  • No Device Output62,456
  • Connection Problem2,239
  • Appropriate Device Problem Term/Code Not Available524
  • Premature End-of-Life Indicator499
  • Alarm Not Visible85
  • Detachment of Device or Device Component59
  • Imprecision52

Patient problems

  • No Clinical Signs, Symptoms or Conditions240,365
  • Hypoglycemia49
  • Hyperglycemia47
  • Loss of consciousness25
  • Elevated Ketones/ Diabetic Ketoacidosis14
  • Vomiting10
  • Fatigue9
  • Convulsion/Seizure7

Recent adverse event reports

The 25 most recent of 240,475 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2026Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24201055 ↗
  • MalfunctionSep 19, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23101812 ↗
  • MalfunctionSep 19, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TESTING WAS PERFORMED AND FAILED. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23099326 ↗
  • MalfunctionSep 19, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23094820 ↗
  • MalfunctionSep 19, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. A TRANSMITTER BATTERY ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23094818 ↗
  • MalfunctionSep 19, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT SENSOR FAILURE OCCURRED. THE SENSOR WAS INSERTED INTO THE THIGH, WHICH IS OFF-LABEL USAGE OF THE DEVICE. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. A SENSOR FAILURE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, A TRANSMITTER FAILURE WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE TRANSMITTER FAILURE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. ALLEGATION CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23092966 ↗
  • MalfunctionSep 19, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT SENSOR FAILURE OCCURRED. THE SENSOR WAS INSERTED INTO THE THIGH, WHICH IS OFF-LABEL USAGE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. A SENSOR FAILURE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, A TRANSMITTER FAILURE WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE TRANSMITTER FAILURE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23092909 ↗
  • MalfunctionSep 18, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23088671 ↗
  • MalfunctionSep 18, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23086507 ↗
  • MalfunctionSep 18, 2025No Device Output; No Device Output

    AFTER SUBMISSION OF THE INITIAL MDR PRODUCT WAS RECEIVED ON 9/4/2025 IT WAS DETERMINED THIS COMPLAINT IS NOT REPORTABLE PER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23086007 ↗
  • MalfunctionSep 17, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23079953 ↗
  • MalfunctionSep 17, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23077740 ↗
  • MalfunctionSep 17, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23076693 ↗
  • MalfunctionSep 17, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT SENSOR FAILURE OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. A SENSOR FAILURE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, A TRANSMITTER FAILURE WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23076124 ↗
  • MalfunctionSep 17, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A PAIRING ISSUE OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. A PAIRING FAILURE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, A TRANSMITTER FAILURE WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE TRANSMITTER FAILURE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23076098 ↗
  • MalfunctionSep 17, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A PAIRING ISSUE OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. PERFORMANCE DATA WAS REVIEWED. A PAIRING FAILURE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, A TRANSMITTER FAILURE WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE TRANSMITTER FAILURE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23075560 ↗
  • MalfunctionSep 17, 2025No Device Output; Wireless Communication Problem; No Device Output; Wireless Communication Problem

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE WAS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE.PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23075541 ↗
  • MalfunctionSep 17, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23075405 ↗
  • MalfunctionSep 17, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A TRANSMITTER BATTERY ISSUE OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23072872 ↗
  • MalfunctionSep 17, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. DATA WAS RECEIVED BUT DOES NOT REFLECT FULL INVESTIGATION WINDOW. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23072588 ↗
  • MalfunctionSep 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. DATA WAS RECEIVED BUT DOES NOT REFLECT FULL INVESTIGATION WINDOW. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23069039 ↗
  • MalfunctionSep 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23068446 ↗
  • MalfunctionSep 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23067117 ↗
  • MalfunctionSep 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE WAS THE TRANSMITTER AND APP WERE UNABLE TO RESTORE THE CONNECTION. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23066683 ↗
  • MalfunctionSep 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. IT WAS INDICATED THE PATIENT STARTED THEIR TRANSMITTER PAST THE 'USE BY' DATE, WHICH IS MISUSE OF THE DEVICE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23063887 ↗

FDA clearance and classification

The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.

Common questions about the Dexcom G6 Transmitter 9445-24

Has the Dexcom G6 Transmitter 9445-24 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00386270001627), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Dexcom G6 Transmitter 9445-24?

The FDA's MAUDE database lists 240,475 adverse event reports for this device, received between Dec 16, 2020 and Jan 28, 2026. In those reports, the most common specific device problems are “Wireless Communication Problem” (174,479), “No Device Output” (62,456) and “Connection Problem” (2,239); the most common patient problems are “Hypoglycemia” (49), “Hyperglycemia” (47) and “Loss of consciousness” (25). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dexcom G6 Transmitter 9445-24?

None of the reports are classified as deaths, 86 as injuries and 240,389 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Dexcom G6 Transmitter 9445-24 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Dexcom G6 Transmitter 9445-24, and is it still sold?

The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G6 Transmitter, model 9445-24). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →