US G6 Transmitter (Retail, Receiver Optional) STT-OE-002Recalls & reported problems
Dexcom, Inc. · G6 Retail Tx-IFU Bundle Kit · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is a Dexcom G6 transmitter sold at retail. Its reports mostly describe lost wireless signal, no output from the device and connection problems. There are 76,635 reports, with no deaths and 111 injuries. No recalls name this device. Reports dropped to 8,227 in the last 12 months, from 17,750 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 76,498
- Deaths reported
- 0
- Injuries reported
- 117
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 76,498 adverse event reports that list this exact device (DI 00386270001771), received between Apr 12, 2022 and Sep 30, 2026 — classified as 117 injuries, 76,381 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (53,719 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Wireless Communication Problem53,719
- No Device Output21,355
- Connection Problem650
- Appropriate Device Problem Term/Code Not Available409
- Premature End-of-Life Indicator186
- Low Readings136
- Alarm Not Visible61
- Unintended Application Program Shut Down52
Patient problems
- No Clinical Signs, Symptoms or Conditions76,182
- Hyperglycemia205
- Hypoglycemia72
- Elevated Ketones/ Diabetic Ketoacidosis26
- Vomiting26
- Loss of consciousness25
- Fatigue14
- Nausea13
Recent adverse event reports
The 25 most recent of 76,498 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50955085 ↗MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954990 ↗MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954764 ↗MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954743 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954292 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952897 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952282 ↗MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952274 ↗InjurySep 30, 2026Appropriate Device Problem Term/Code Not Available
It was reported that an adverse event occurred. The sensor was inserted into the skin on (b)(6)2026. On 09/16/2026, it was reported that the patient experienced loss of consciousness while not being in a sensor session. The available information states that the patient experienced signal loss with their Dexcom G6 sensor on 09/16. No symptoms were reported from that session, but it was reported that due to this the patient had no Dexcom G6 sensors left. An unknown time after this, the patient´s doctor provided a G7 sensor for 15 days. However, the patient´s phone was not compatible with the G7 sensor, and the patient was unable to start a sensor session. The patient then would have experienced loss of consciousness. However, the date reported for the loss of consciousness was (b)(6)2026. No fingerstick was reported and it was unknown if the patient experienced a hypo or hyperglycemic event. It was noted that this is a date before the G6 malfunction would have occurred, but the discrepancy could not be clarified. No further information regarding possible treatment and or hospital visit was documented. At the time of the report the patient´s current condition was unknown. No other det
Patient problems: Loss of consciousness
MAUDE report 50952044 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that "Signal Loss" occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952031 ↗MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952006 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951715 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951709 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951707 ↗MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951703 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951702 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951700 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951696 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951672 ↗MalfunctionSep 30, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951669 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. It was indicated that the patient used an expired product, which is misuse of the device. Performance data was reviewed. The allegation was confirmed due to the finding of a transmitter failed error within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951668 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950764 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over one hour was found in connection to the reported allegation. The probable cause of the signal loss could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950747 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over an hour was found in connection to the reported allegation. The probable cause of the signal loss was determined to be the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950731 ↗MalfunctionSep 30, 2026No Device Output
It was reported that a transmitter battery issue occurred. Data was received but does not reflect full investigation window. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950681 ↗
FDA clearance and classification
The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗
Devices of this type normally need 510(k) clearance before sale.
FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Common questions about the US G6 Transmitter (Retail, Receiver Optional) STT-OE-002
Has the US G6 Transmitter (Retail, Receiver Optional) STT-OE-002 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00386270001771), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the US G6 Transmitter (Retail, Receiver Optional) STT-OE-002?
The FDA's MAUDE database lists 76,498 adverse event reports for this device, received between Apr 12, 2022 and Sep 30, 2026. In those reports, the most common specific device problems are “Wireless Communication Problem” (53,719), “No Device Output” (21,355) and “Connection Problem” (650); the most common patient problems are “Hyperglycemia” (205), “Hypoglycemia” (72) and “Elevated Ketones/ Diabetic Ketoacidosis” (26). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the US G6 Transmitter (Retail, Receiver Optional) STT-OE-002?
None of the reports are classified as deaths, 117 as injuries and 76,381 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the US G6 Transmitter (Retail, Receiver Optional) STT-OE-002 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the US G6 Transmitter (Retail, Receiver Optional) STT-OE-002, and is it still sold?
The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand US G6 Transmitter (Retail, Receiver Optional), model STT-OE-002). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices
- Dexcom G6 Transmitter 9445-24Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated240,475 reports
- Dexcom G6 Transmitter 9445-02Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 deaths229,489 reports
- Dexcom G6 Sensor Applicator Assembly 9500-45Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall4 deaths114,621 reports
- Dexcom G7 GSS (161) 9500-161Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated70,146 reports
- Dexcom G7 iOS App SW12300Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 recalls1 death54,855 reports
- Dexcom G7 GSS (161) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall5 deaths47,762 reports
- G6 Medicare Transmitter STT-OM-001Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated38,212 reports
- Dexcom G7 GSS (160) 9500-160Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated33,025 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.