Dexcom G7 iOS App SW12300Recalls & reported problems
Dexcom, Inc. · G7 iOS App · Class 2
All models of Dexcom G7 (32)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is the Dexcom G7 iOS app. Reports mostly describe the app shutting down unexpectedly, alarms that do not work and wireless connection problems. There are 54,718 reports, including 1 death and 171 injuries. Three recalls are open, one involving glucose readings that pile up in the background and then replay. Reports rose to 46,116 in the last 12 months, from 8,195 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54,855
- Deaths reported
- 1
- Injuries reported
- 156
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 54,855 adverse event reports that list this exact device (DI 00386270001863), received between Dec 30, 2022 and Sep 30, 2026 — classified as 1 death, 156 injuries, 54,694 malfunctions.
The most frequently reported specific device problem is “Protective Measures Problem” (39,306 reports).
3 FDA recall records name this device's identifier, 3 still open.
3 other recalls cover devices cleared under the same FDA submission (K213919); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 14, 2026Class IIOpen, Classified
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
Z-2163-2026 · Root cause: Software design · Recall details · FDA record ↗
- Jul 24, 2025Class IOpen, Classified
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.
Z-2447-2025 · Root cause: Software design · Recall details · FDA record ↗
- May 5, 2025Class IIOpen, Classified
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts will not trigger, there is potential for the missed detection of a hyperglycemic or hypoglycemic event or a treatment decision made based off incorrect data. There is no impact to transmitter communication with concurrently connected displays. A concurrently connected receiver, Apple Watch, and/or insulin pump will continue to receive EGVs directly from the transmitter when the app is in the problematic state. There is no impact to Automated Insulin Dosing (AID) therapy as the AID system continues to receive EGVs directly from the transmitter.
Z-1906-2025 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Protective Measures Problem39,306
- Unintended Application Program Shut Down7,141
- Operating System Version or Upgrade Problem1,885
- Wireless Communication Problem1,371
- Appropriate Device Problem Term/Code Not Available1,336
- Defective Alarm1,262
- No Audible Alarm1,203
- Display or Visual Feedback Problem1,011
Patient problems
- No Clinical Signs, Symptoms or Conditions54,693
- Hypoglycemia101
- Loss of consciousness66
- Confusion/ Disorientation29
- Convulsion/Seizure21
- Hypoglycemic Shock19
- Dizziness18
- Diaphoresis14
Recent adverse event reports
The 25 most recent of 54,855 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Unintended Application Program Shut Down
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954759 ↗MalfunctionSep 30, 2026Unintended Application Program Shut Down
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954639 ↗MalfunctionSep 30, 2026Unintended Application Program Shut Down
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954623 ↗MalfunctionSep 30, 2026Unintended Application Program Shut Down
It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that an app crash occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951161 ↗MalfunctionSep 30, 2026Unintended Application Program Shut Down
It was reported that an alert/notification settings occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. The allegation was undetermined via data. However, it was found that an app crash occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951152 ↗MalfunctionSep 30, 2026No Audible Alarm
It was reported that an alert/notification settings issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951151 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that a pairing issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. A pairing issue was not found within the investigation window. The allegation was not confirmed. However, signal loss over an hour was found in connection to the reported allegation. It was determined that the signal loss was related to the mobile application. The probable cause of the signal loss was determined to be the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949821 ↗MalfunctionSep 30, 2026Defective Alarm
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949789 ↗MalfunctionSep 30, 2026No Audible Alarm
It was reported that an alert/notification settings issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948470 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that signal loss occurred. It was determined that loss of connection was related to the Mobile Application. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948398 ↗MalfunctionSep 30, 2026Unintended Application Program Shut Down
It was reported that an alert/notification settings occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. The allegation was undetermined via data. However, it was found that an app crash occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948142 ↗MalfunctionSep 30, 2026Display or Visual Feedback Problem
It was reported that a display issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Data was provided for investigation. Confirmation of the complaint was confirmed via data. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947714 ↗MalfunctionSep 30, 2026Display or Visual Feedback Problem
It was reported that a display issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Data was provided for investigation. Confirmation of the complaint was confirmed via data. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947648 ↗MalfunctionSep 30, 2026Display or Visual Feedback Problem
It was reported that an alert/notification settings occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. The allegation was undetermined via data. However, it was found that EGVs not updating issue occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947339 ↗MalfunctionSep 30, 2026Unintended Application Program Shut Down
It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that an app crash occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947321 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that an alert/notification settings issue occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. An alert issue was not found within the investigation window. The allegation was undetermined. However, signal loss over 1 hour was found in in connection to the reported allegation. The probable cause of the signal loss was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947319 ↗MalfunctionSep 30, 2026Display or Visual Feedback Problem
It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that EGVs not updating issue occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947309 ↗MalfunctionSep 30, 2026Wireless Communication Problem
It was reported that signal loss occurred. It was determined that loss of connection was related to the Mobile Application. A review of the share logs was performed and signal loss was found within the investigation window. The allegation was confirmed. The probable cause was the transmitter and app were unable to complete a data transfer. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943810 ↗MalfunctionSep 30, 2026Display or Visual Feedback Problem
It was reported that a display issue occurred. Data was provided for investigation. Confirmation of the complaint was confirmed via data. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943322 ↗MalfunctionSep 30, 2026Wireless Communication Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941491 ↗MalfunctionSep 29, 2026Unintended Application Program Shut Down
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50940302 ↗MalfunctionSep 29, 2026Operating System Version or Upgrade Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50939760 ↗MalfunctionSep 29, 2026Operating System Version or Upgrade Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50939734 ↗MalfunctionSep 29, 2026No Audible Alarm
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938510 ↗MalfunctionSep 29, 2026Display or Visual Feedback Problem
It was reported that an alert/notification settings occurred. It was indicated that the patient was using an unsupported operating system, which is misuse of the device. Performance data was reviewed. The allegation was undetermined via data. However, it was found that EGVs not updating issue occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938473 ↗
FDA clearance and classification
Cleared through 510(k) K213919 on Dec 7, 2022 (decision: Substantially Equivalent). Submitted by Dexcom, Inc. as “Dexcom G7 Continuous Glucose Monitoring System”. The FDA review took 357 days. FDA record ↗
The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗
FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.
Recalls under the same FDA submission
These recalls involve devices cleared under K213919. One submission often covers many models, so check the product and lot details in each record.
- Z-2446-2025 · Jul 24, 2025 · Open, ClassifiedBrand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product
- Z-2448-2025 · Jul 24, 2025 · Open, ClassifiedBrand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Produc
- Z-1944-2025 · May 12, 2025 · Open, ClassifiedDexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013
Common questions about the Dexcom G7 iOS App SW12300
Has the Dexcom G7 iOS App SW12300 been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-2163-2026 (Class II), began on Apr 14, 2026: A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Dexcom G7 iOS App SW12300?
The FDA's MAUDE database lists 54,855 adverse event reports for this device, received between Dec 30, 2022 and Sep 30, 2026. In those reports, the most common specific device problems are “Protective Measures Problem” (39,306), “Unintended Application Program Shut Down” (7,141) and “Operating System Version or Upgrade Problem” (1,885); the most common patient problems are “Hypoglycemia” (101), “Loss of consciousness” (66) and “Confusion/ Disorientation” (29). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Dexcom G7 iOS App SW12300?
1 report is classified as a death, 156 as injuries and 54,694 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Dexcom G7 iOS App SW12300 FDA approved?
It was cleared through 510(k) K213919 on Dec 7, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Dexcom G7 iOS App SW12300, and is it still sold?
The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G7 iOS App, model SW12300). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices
- Dexcom G6 Transmitter 9445-24Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated240,475 reports
- Dexcom G6 Transmitter 9445-02Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 deaths229,489 reports
- Dexcom G6 Sensor Applicator Assembly 9500-45Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall4 deaths114,621 reports
- US G6 Transmitter (Retail, Receiver Optional) STT-OE-002Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated76,498 reports
- Dexcom G7 GSS (161) 9500-161Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated70,146 reports
- Dexcom G7 GSS (161) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall5 deaths47,762 reports
- G6 Medicare Transmitter STT-OM-001Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated38,212 reports
- Dexcom G7 GSS (160) 9500-160Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated33,025 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page