Epix Inline Laparoscopic Scissors CB050Recalls & reported problems
Applied Medical Resources Corporation · Laparoscopic Scissors · Class 2
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00607915124519), received between Nov 9, 2016 and Jun 13, 2017 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Torn Material” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Torn Material5
- Failure to Cut1
- Melted1
- Output Problem1
- Overheating of Device1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient2
- Injury1
- Tissue Damage1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 13, 2017Torn Material; Torn Material
PRE-OPERATION - SCISSOR REMOVED FROM PACKAGE AND UPON INSPECTION PRIOR TO CASE, THEY NOTICED A BREAK IN THE INSULATION ON THE DISTAL END OF THE SCISSOR. TYPE OF INTERVENTION - NA. PATIENT STATUS - NA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6639145 ↗MalfunctionMar 6, 2017Failure to Cut; Failure to Cut
PROCEDURE PERFORMED UNKNOWN - "SCISSORS CUTTING DETERIORATED THROUGHOUT CASE WHERE MULTIPLE (5-6) FIBROIDS WHERE REMOVED. DR. FOX CONTINUED TO USE THEM UNTIL THEY FAILED TO CUT AT ALL. IT WAS A LONG CASE AND HE MADE MANY CUTS WITH MINIMAL USE OF THE CAUTERY." PATIENT STATUS - GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6382350 ↗MalfunctionFeb 23, 2017Torn Material; Torn Material
UNKNOWN PROCEDURE - "AFTER THE SCISSOR WAS OPENED THE PROTECTIVE SHEATH WAS REMOVED AND IN THE PROCESS THE INSULATION TORE." PATIENT STATUS - N/A.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6353713 ↗MalfunctionJan 12, 2017Torn Material; Torn Material
PROCEDURE PERFORMED UNKNOWN - "AFTER THE SCISSOR WAS OPENED THE PROTECTIVE SHEATH WAS REMOVED AND IN THE PROCESS THE INSULATION TORE." TYPE OF INTERVENTION - "USED THE PISTOL GRIP SCISSOR." PATIENT STATUS - "NA".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6246898 ↗MalfunctionJan 12, 2017Torn Material; Torn Material
PROCEDURE PERFORMED UNKNOWN - "AFTER THE SCISSOR WAS OPENED THE PROTECTIVE SHEATH WAS REMOVED AND IN THE PROCESS THE INSULATION TORE." TYPE OF INTERVENTION - "USED THE PISTOL GRIP SCISSOR." PATIENT STATUS - "NA".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6246806 ↗MalfunctionJan 12, 2017Torn Material; Torn Material
PROCEDURE PERFORMED UNKNOWN - "AFTER THE SCISSOR WAS OPENED THE PROTECTIVE SHEATH WAS REMOVED AND IN THE PROCESS THE INSULATION TORE." TYPE OF INTERVENTION - "USED THE PISTOL GRIP SCISSOR." PATIENT STATUS - "NA."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6246667 ↗MalfunctionJan 4, 2017Output Problem; Output Problem
DIAGNOSTIC LAPAROSCOPY - "THE MONOPOLAR ENERGY TRANSMISSION DELIVERED FROM THE SCISSOR TO THE PATIENT WAS INTERMITTENT. THEY REPLACED THE DEVICE AND IT WORKED FINE." PATIENT STATUS - NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6224024 ↗MalfunctionNov 9, 2016Melted; Overheating of Device; Melted; Overheating of Device
ADDITIONAL INFORMATION RECEIVED DECEMBER 7, 2016. "THE CLINICAL DEVELOPMENT TEAM VISITED (B)(6) SURGERY CENTER ON (B)(6) 2016 REGARDING (B)(4) (A CER REFERENCING INSULATION DAMAGE). IN DISCUSSIONS WITH THE SCRUB TECH, SHE COMMENTED THAT, IN HER OPINION, THE ROTATION KNOB HEATING UP (CER 2016-1035) WAS NOT A SERIOUS ISSUE AND THE CONCERNS FROM THE SURGEON WERE MOST LIKELY STEMMING FROM THE INCIDENT RELATED TO THE INSULATION DAMAGE FOR (B)(4)."
Patient problems: Injury; Tissue Damage
MAUDE report 6090034 ↗
Other Laparoscope, Gynecologic (And Accessories) devices
- Olympus WA50042AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)1 recall6,135 reports
- Olympus WA50052AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)2,249 reports
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5Olympus Medical Systems Corp. · LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)1 recall1,279 reports
- Olympus WA50040AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)1 recall954 reports
- Olympus WA50050AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)895 reports
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5Olympus Medical Systems Corp. · Laparoscope, Gynecologic (And Accessories)624 reports
- Epix Laparoscopic Scissors CB030Applied Medical Resources Corporation · Laparoscope, Gynecologic (And Accessories)239 reports
- HD EndoEYE LAPARO-THORACO VIDEOSCOPE OLYMPUS LTF TYPE VHOlympus Medical Systems Corp. · LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)168 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.