ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5 (REF N3828033)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is an Olympus ENDOEYE FLEX deflectable videoscope used in gynecologic laparoscopy. The 624 reports mostly describe no image or a black screen, flickering, or poor image quality, with some noting image noise and transmission problems. There are 2 injuries and no deaths. No recalls name this device. There were no reports in the last 12 months, down from 46 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
624
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 624 adverse event reports that list this exact device (DI 04953170434648), received between May 4, 2023 and Jul 9, 2025 — classified as 2 injuries, 622 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (384 reports).

1 other recall cover devices cleared under the same FDA submission (K201832); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 33 reportsDec 2023: 42 reportsJan 2024: 49 reportsFeb 2024: 40 reportsMar 2024: 56 reportsApr 2024: 54 reportsMay 2024: 24 reportsJun 2024: 38 reportsJul 2024: 35 reportsAug 2024: 38 reportsSep 2024: 40 reportsOct 2024: 42 reportsNov 2024: 30 reportsDec 2024: 6 reportsJan 2025: 4 reportsFeb 2025: 4 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202656
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image384
  • Communication or Transmission Problem151
  • Poor Quality Image78
  • Erratic or Intermittent Display18
  • Output Problem16
  • Display or Visual Feedback Problem10
  • Noise, Audible10
  • Image Display Error/Artifact9

Patient problems

  • No Clinical Signs, Symptoms or Conditions624

Recent adverse event reports

The 25 most recent of 624 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 9, 2025No Display/Image; Poor Quality Image; No Display/Image; Poor Quality Image

    IT WAS REPORTED THAT DURING SERVICE / MAINTENANCE, THE FLEX DEFLECTABLE VIDEOSCOPE HAD IMAGE NOISE AND NO IMAGE WHEN OPERATING THE ANGLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22459105 ↗
  • MalfunctionMar 7, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THE VIDEOSCOPE WAS NOT DISPLAYING AN IMAGE ON THE MONITOR DURING A LOW ANTERIOR RESECTION PROCEDURE. THE EVENT DID NOT RESULT IN ANY DELAY TO THE PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21544298 ↗
  • MalfunctionFeb 26, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED SCREEN WAS COMPLETELY BLACK AND NO IMAGE WAS DISPLAYED DUE TO DAMAGE TO THE IMAGING SECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21473592 ↗
  • MalfunctionFeb 25, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT DURING A THERAPEUTIC LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, OUTSIDE OF THE SEDATED PATIENT, THE SUBJECT FLEX VIDEO SCOPE DEVICE HAD A GREEN DOT APPEAR ON THE RIGHT SIDE OF THE IMAGE. THERE WAS NO HARM ON INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21463348 ↗
  • MalfunctionFeb 23, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21442079 ↗
  • MalfunctionFeb 10, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT WHEN THE DEVICE WAS ANGULATED, IT CAUSED A BLACKOUT OF THE IMAGE. THE ISSUE OCCURRED DURING AN UNKNOWN THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21350567 ↗
  • MalfunctionJan 17, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT THE SUBJECT DEVICE'S IMAGE WAS FLICKERING. THE ISSUE WAS FOUND DURING SET UP. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21181149 ↗
  • MalfunctionJan 16, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21172678 ↗
  • MalfunctionJan 16, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21169520 ↗
  • MalfunctionJan 14, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE SUBJECTED DEVICE DISPLAYED NO IMAGE. THE ISSUE WAS FOUND DURING SETUP OR INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21149901 ↗
  • MalfunctionDec 29, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT DURING SET UP THE SUBJECT DEVICE ENCOUNTERED SCOPE COMMUNICATION ERROR, E216 DURING SETUP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21032567 ↗
  • MalfunctionDec 11, 2024Poor Quality Image; Communication or Transmission Problem; Poor Quality Image; Communication or Transmission Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20900816 ↗
  • MalfunctionDec 10, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THE FLEX DEFLECTABLE VIDEOSCOPE HAD A COMMUNICATION ERROR. THE ISSUE OCCURRED DURING SETUP/INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20889233 ↗
  • MalfunctionDec 9, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT WHILE USING THE FLEX VIDEO SCOPE AN IMAGE WAS NOT DISPLAYED. THE REPORTED MALFUNCTION OCCURRED DURING SETUP AND INSPECTION FOR USE PRIOR TO AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20879184 ↗
  • MalfunctionDec 6, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT, DURING AN UNKNOWN THERAPEUTIC PROCEDURE, THE SUBJECT DEVICE PRODUCED AN IMAGE WHICH WAS COMPLETELY BLACK, AND THE TARGET COULD NOT BE VISUALIZED. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT OLYMPUS DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20871529 ↗
  • MalfunctionDec 3, 2024No Display/Image; Communication or Transmission Problem; No Display/Image; Communication or Transmission Problem

    IT WAS REPORTED THE FLEX DEFLECTABLE VIDEOSCOPE HAD AN E218 SCOPE ERROR, AS WELL AS AN IMAGE CUTOFF. THE ISSUE OCCURRED DURING INSTALLATION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20835563 ↗
  • MalfunctionNov 27, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT DURING REPROCESSING OF THE SUBJECT DEVICE, NO IMAGE WAS DISPLAYED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20801065 ↗
  • MalfunctionNov 27, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE FLEX VIDEO SCOPE HAD A DISTORTED IMAGE AND CANNOT BE SEEN. THE REPORTED MALFUNCTION OCCURRED SETUP AND INSPECTION FOR USE PRIOR TO AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20793732 ↗
  • MalfunctionNov 25, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED, THE FLEX DEFLECTABLE VIDEOSCOPE'S FLEX TIP WAS DAMAGED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20773723 ↗
  • MalfunctionNov 25, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THE FLEX DEFLECTABLE VIDEOSCOPE HAD POOR IMAGE QUALITY. THE ISSUE OCCURRED DURING SETUP/INSPECTION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20772384 ↗
  • MalfunctionNov 24, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE THE SUBJECT DEVICE HAD POOR IMAGE QUALITY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20767601 ↗
  • MalfunctionNov 24, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE SUBJECT DEVICE DISPLAYED AN E126 ERROR DURING SET UP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20767250 ↗
  • MalfunctionNov 24, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE PRESENTED NOISE WHEN ANGULATION WAS APPLIED. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20767160 ↗
  • MalfunctionNov 24, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THE SUBJECT FLEX VIDEO SCOPE IMAGE DID NOT APPEAR. THE ISSUE OCCURRED WHEN PREPARING THE DEVICE FOR USE. THERE WAS NO HARM OR INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20767155 ↗
  • MalfunctionNov 23, 2024Communication or Transmission Problem; Communication or Transmission Problem

    NO NEW INFORMATION HAS BEEN PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20766456 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201832. One submission often covers many models, so check the product and lot details in each record.

  • Z-0644-2024 · Nov 20, 2023 · Open, Classified
    OLYMPUS ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (5.4 mm diameter) Model: LTF-S190-5 Used with a video system center, light source, documentation equipment, monitor, hand instruments

Other Laparoscope, Gynecologic (And Accessories) devices

All Laparoscope, Gynecologic (And Accessories) devices →