Skip to content

Olympus WA50050ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Video telescope "ENDOEYE HD II", 5.4 mm, 0°, autoclavable · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus ENDOEYE HD II video telescope appears in 876 adverse event reports, with 3 injuries and no deaths. Many describe poor image quality, fiber breaks, and dents and cracks in its parts, with some mentioning stains and black marks on the connector. Reports in the last year rose to 395, from 259 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
895
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 895 adverse event reports that list this exact device (DI 04042761075008), received between Dec 16, 2020 and Sep 30, 2026 — classified as 3 injuries, 892 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (262 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 14 reportsMay 2024: 32 reportsJun 2024: 42 reportsJul 2024: 24 reportsAug 2024: 31 reportsSep 2024: 41 reportsOct 2024: 31 reportsNov 2024: 13 reportsDec 2024: 14 reportsJan 2025: 21 reportsFeb 2025: 10 reportsMar 2025: 22 reportsApr 2025: 19 reportsMay 2025: 19 reportsJun 2025: 26 reportsJul 2025: 35 reportsAug 2025: 37 reportsSep 2025: 19 reportsOct 2025: 24 reportsNov 2025: 36 reportsDec 2025: 42 reportsJan 2026: 32 reportsFeb 2026: 39 reportsMar 2026: 80 reportsApr 2026: 39 reportsMay 2026: 55 reportsJun 2026: 18 reportsJul 2026: 22 reportsAug 2026: 22 reportsSep 2026: 29 reportsOct 2026: 0 reports20242025202680
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image262
  • Break182
  • Dent in Material139
  • No Display/Image119
  • Crack98
  • Material Twisted/Bent80
  • Scratched Material68
  • Loose or Intermittent Connection49

Patient problems

  • No Clinical Signs, Symptoms or Conditions892
  • Insufficient Information2
  • Burn(s)1

Recent adverse event reports

The 25 most recent of 895 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Break; Contamination; No Display/Image

    The customer reported the Olympus rigid video laparoscope can see through it during an unspecified procedure. During the initial evaluation of the device, it was observed, Fiber breakage, Brown stains under light guide cover glass and absolutely no image. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944571 ↗
  • MalfunctionSep 29, 2026Break; Loose or Intermittent Connection; Poor Quality Image

    The customer reported camera picture was not clear involving an Olympus rigid video laparoscope. During initial evaluation it was observed, distal fiber breakage and loose adapter. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936635 ↗
  • MalfunctionSep 29, 2026Break; Crack; Low Readings; Moisture or Humidity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926868 ↗
  • MalfunctionSep 28, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916068 ↗
  • MalfunctionSep 28, 2026Leak/Splash

    The customer reported a leak during an inspection for use involving an Olympus rigid video laparoscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915848 ↗
  • MalfunctionSep 24, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889774 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    The customer returned the rigid video laparoscope for a separate issue. During the initial evaluation of the device, it was observed, distal fibers and fiber breakage was observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884032 ↗
  • MalfunctionSep 24, 2026Material Discolored

    The customer returned the rigid video laparoscope for a separate issue. During the initial evaluation of the returned device, minor stains and minor brown spots under the light guide cover glass were observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876742 ↗
  • MalfunctionSep 23, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50874148 ↗
  • MalfunctionSep 23, 2026Leak/Splash

    The customer reported that leakage occurred during inspection for use involving an Olympus Rigid Video Laparoscope. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872041 ↗
  • MalfunctionSep 23, 2026Crack; Poor Quality Image

    The customer reported Cracked Lens during inspection for use, Involving an Olympus rigid video Laparoscope. During the initial device evaluation, a cracked image was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863155 ↗
  • MalfunctionSep 23, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862990 ↗
  • MalfunctionSep 22, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50861743 ↗
  • MalfunctionSep 16, 2026Break; Detachment of Device or Device Component

    The customer returned the colonovideoscope, which is an Olympus asset with no reported complaint. During the initial evaluation of the returned device, a chipped light guide bundle and damaged distal end was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50782524 ↗
  • MalfunctionSep 15, 2026No Display/Image; Communication or Transmission Problem; Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50765976 ↗
  • MalfunctionSep 14, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50765793 ↗
  • MalfunctionSep 14, 2026Break; Poor Quality Image

    No additional information received from the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50765743 ↗
  • MalfunctionSep 13, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50747944 ↗
  • MalfunctionSep 12, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50744695 ↗
  • MalfunctionSep 11, 2026No Display/Image; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50740725 ↗
  • MalfunctionSep 11, 2026Break; No Display/Image; Loose or Intermittent Connection; Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50725265 ↗
  • MalfunctionSep 10, 2026Break; Loose or Intermittent Connection; Poor Quality Image; No Device Output

    The customer returned the rigid video laparoscope for a separate issue. During the initial evaluation of the returned device, Fiber breakage, Loose adapter, Out of direction on image and Light Transmission failed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50714188 ↗
  • MalfunctionSep 9, 2026Break; Poor Quality Image; Connection Problem; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50702516 ↗
  • MalfunctionSep 6, 2026Material Twisted/Bent; Optical Discoloration; Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50640744 ↗
  • MalfunctionSep 6, 2026Break; Poor Quality Image; Dent in Material; Material Twisted/Bent; Optical Discoloration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50639821 ↗

Other Laparoscope, Gynecologic (And Accessories) devices

All Laparoscope, Gynecologic (And Accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page