PARADIGM® REAL-time MMT-722NAB (UDI 00613994223715)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-722NAB PRDGM INSULIN BL EN ML · Class 3

All models of PARADIGM REAL-time (95)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00613994223715), received between Sep 18, 2017 and Jul 13, 2020 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Device Displays Incorrect Message2
  • Failure to Cycle2
  • Moisture Damage1
  • No Apparent Adverse Event1
  • No Display/Image1

Patient problems

  • No Consequences Or Impact To Patient4
  • Hyperglycemia2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 13, 2020No Display/Image; No Display/Image

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS NO LONGER TURNING ON. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10262816 ↗
  • MalfunctionApr 8, 2020Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD COMPROMISED BROKEN FORCE SENSOR ALARM DURING PRIMING. CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS LOOSENED. DURING SEATING THE FORCE SENSOR DID NOT DETECT THE RESERVOIR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9937022 ↗
  • InjuryNov 14, 2018Failure to Cycle; Detachment of Device or Device Component; No Apparent Adverse Event; Failure to Cycle; Detachment of Device or Device Component; No Apparent Adverse Event

    CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM AND THEY EXPERIENCED HIGH BLOOD GLUCOSE LEVELS. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 400 MG/DL. CUSTOMER TREATED THEIR HIGH BLOOD GLUCOSE LEVELS VIA MANUAL INJECTION. CUSTOMER ALSO REPORTED THAT THE INSULIN WAS SQUIRTING OUT DURING MANUAL PRIME PROCESS. CUSTOMER WAS UNABLE TO EXIT THE PREPARING TO PRIME LOOP. CUSTOMER ADVISED DURING THE SEATING OF THE FORCE THE SENSOR DID NOT DETECT THE RESERVOIR. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 8068350 ↗
  • MalfunctionMar 14, 2018Device Displays Incorrect Message

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE PATIENT¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7340148 ↗
  • InjuryDec 11, 2017Device Displays Incorrect Message

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY RECEIVED A BUTTON ERROR ON THE INSULIN PUMP. THE CUSTOMER¿S BLOOD GLUCOSE DURING THE INCIDENT WAS 400 MG/DL. THE HIGH BLOOD GLUCOSE WAS TREATED WITH AN INSULIN SHOT. THE CUSTOMER STATED THEY HAD A NEW INSULIN PUMP COMING IN. THEY DECLINED TROUBLESHOOTING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 7104430 ↗
  • MalfunctionSep 18, 2017Moisture Damage

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP WAS EXPOSED TO POOL WATER. THE DISPLAY WENT BLANK IMMEDIATELY AFTER INSULIN PUMP¿S EXPOSURE TO MOISTURE. THE CUSTOMER¿S BLOOD GLUCOSE DURING THE INCIDENT WAS UNKNOWN. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6874171 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices

All Sensor, glucose, invasive devices →