Skip to content

PARADIGM REAL-timeRecalls & reported problems across all models

Medtronic Minimed, Inc. · 95 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one PARADIGM REAL-time model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
2,862
Deaths reported
24
Injuries reported
637
Recalls
1

We found 95 models of PARADIGM REAL-time by Medtronic Minimed, Inc. in the FDA's device registry (GUDID). Together they are named in 2,862 adverse event reports in the FDA's MAUDE database, received between May 5, 2017 and Jun 4, 2026 — classified as 24 deaths, 637 injuries, 2,201 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (603 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

PARADIGM REAL-time recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Feb 13, 2026Open, Classified

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

    Z-1749-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

Most reported PARADIGM REAL-time problems

Device problems

  • Device Displays Incorrect Message603
  • Device Difficult to Program or Calibrate420
  • Mechanical Problem346
  • Failure to Cycle294
  • Patient Device Interaction Problem285
  • Adverse Event Without Identified Device or Use Problem235
  • No Display/Image195
  • Detachment of Device or Device Component194

Patient problems

  • No Consequences Or Impact To Patient1,954
  • Hyperglycemia709
  • Hypoglycemia178
  • Elevated Ketones/ Diabetic Ketoacidosis37
  • Death32
  • No Known Impact Or Consequence To Patient22
  • No Clinical Signs, Symptoms or Conditions10
  • Nausea6

PARADIGM REAL-time models

95 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent PARADIGM REAL-time adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Medtronic Minimed, Inc. products

All Medtronic Minimed, Inc. devices →

Sources and updates