Skip to content

PARADIGM® REAL-time MMT-722NAL (UDI 00613994312389)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-722NAL PRDGM V2.2 CL EN PR LK · Class 3

All models of PARADIGM REAL-time (95)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00613994312389), received between Jul 13, 2017 and Mar 11, 2020 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Device Difficult to Program or Calibrate” (2 reports).

14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Difficult to Program or Calibrate2
  • Mechanical Problem2
  • No Apparent Adverse Event2
  • Priming Problem2
  • Battery Problem1
  • Failure of Device to Self-Test1
  • Failure to Cycle1
  • Moisture or Humidity Problem1

Patient problems

  • No Consequences Or Impact To Patient5
  • Hyperglycemia2
  • Elevated Ketones/ Diabetic Ketoacidosis1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2020Mechanical Problem; Battery Problem; Power Problem; Priming Problem; Mechanical Problem; Battery Problem; Power Problem; Priming Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP SQUIRTED OUT INSULIN DURING PRIMING. CUSTOMER ALSO STATED THAT INSULIN PUMP REJECTED NEW BATTERIES, HAD TO REPLACE BATTERIES MORE FREQUENTLY AFTER EVERY 5 TO 6 DAYS, HAD FAILED BATTERY ALARM, LOST TIME AND DATE AFTER BATTERY CHANGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9816378 ↗
  • MalfunctionOct 30, 2019Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER ALSO REPORTED THAT THE INSULIN WAS SQUIRTING DURING MANUAL PRIME PROCESS. THE CUSTOMER WAS UNSURE THAT WHETHER DRIVE SUPPORT CAP WAS PROTRUDED, LOOSE,STICKING OUT OR NOT. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9258473 ↗
  • InjuryApr 11, 2019Moisture or Humidity Problem; No Apparent Adverse Event; Moisture or Humidity Problem; No Apparent Adverse Event

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE LEVEL. CUSTOMER¿S BLOOD GLUCOSE WAS 500 MG/DL AT THE TIME OF INCIDENT. THE CURRENT BLOOD GLUCOSE LEVEL IS 500 MG/DL. CUSTOMER STATES THEY WAS IN HOSPITAL BECAUSE OF NOT GETTING INSULIN. CUSTOMER STATES RESERVOIR HAD LEAK IN PAST SECOND O RING. CUSTOMER STATES THE INSULIN PUMP WAS VIBRATING WITHOUT AN ALARM OR ALERT. THE INSULIN PUMP AND RESERVOIR WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 8503229 ↗
  • MalfunctionMar 28, 2019Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    THE CUSTOMER REPORTED VIA PHONE CALL THAT INSULIN PUMP HAD BUTTON ERROR ALARM. CUSTOMER¿S BLOOD GLUCOSE WAS 85 MG/DL AT THE TIME OF INCIDENT. ADVISED THE PUMP WILL NEED TO BE REPLACED. ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8459831 ↗
  • MalfunctionMar 12, 2019Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    THE CUSTOMER REPORTED VIA PHONE CALL THE INSULIN PUMP HAD A KEYPAD ANOMALY. THE CUSTOMER¿S BLOOD GLUCOSE WAS 131 MG/DL AT THE TIME OF INCIDENT. CUSTOMER STATED THAT THE INSULIN PUMP BUTTONS WERE STICKING AND WOULD RAMP UP. NO SIGNIFICANT EVENTS LEADING TO KEYPAD ANOMALY WERE OBSERVED. KEYPAD ANOMALY TROUBLESHOOTING WAS PERFORMED AND CONFIRMED THAT TIME IS ADVANCING. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF PUMP AND REVERT TO BACK-UP PLAN. THE INSULIN PUMP IS BEING REPLACED AND IS EXPECTED TO RETURN FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8412221 ↗
  • MalfunctionJan 3, 2019Mechanical Problem; Obstruction of Flow; Patient Device Interaction Problem; Priming Problem; Mechanical Problem; Obstruction of Flow; Patient Device Interaction Problem; Priming Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP DAMAGE. THE CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER STATES THAT THE INSULIN PUMP HAD RANDOM NO DELIVERY ALARMS AND HAD TO MANUALLY REWIND TO GET IT TO WORK, BATTERY CAP TOP EDGE HAD BROKEN OFF AND INSULIN SHOOTS OUT EVERY TIME WHEN CUSTOMER PRIMES IT. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8211716 ↗
  • InjuryOct 16, 2018No Apparent Adverse Event; No Apparent Adverse Event

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HYPERGLYCEMIA OF VALUE 415 MG/DL. CUSTOMER ALSO STATED INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER ALSO REPORTED THAT THE INSULIN WAS SQUIRTING DURING MANUAL PRIME PROCESS. CUSTOMER ALSO STATED INSULIN CONTINUES TO DRIP AFTER MOTOR STOPS DURING THE MANUAL PRIME PROCESS AND THEY ARE UNABLE TO EXIT THE PREPARING TO PRIME LOOP. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS NORMAL. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 7972212 ↗
  • MalfunctionJul 13, 2017Failure of Device to Self-Test

    IT WAS REPORTED THAT THE SELF TEST FAILED ON THE INSULIN PUMP. THE INSULIN PUMP WILL RETURN. THE CUSTOMER'S BLOOD SUGAR IS 277 MG/DL.

    Patient problems: Hyperglycemia

    MAUDE report 6712718 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices

All Sensor, glucose, invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page